© 2022 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

The aim of this study is to evaluate the wear volume of interim crowns fabricated using digital light processing 3D printing according to the printing angle. A total of five patients undergoing the placement of a single crown on the mandibular molar were included. Interim crowns were fabricated directly in the oral cavity using the conventional method. A digital light processing 3D printer was then used to fabricate crowns with build angles of 0, 45, and 90 degrees. Therefore, four fabricated interim crowns were randomly delivered to the patients, and each was used for one week. Before and after use, the intaglio surfaces of the interim crowns were scanned using a 3D scanner. The volume changes before and after use were measured, and changes in the height of the occlusal surface were evaluated using the root mean square value. Data normality was verified by statistical analysis, and the wear volume in each group was evaluated using a one-way analysis of variance and Tukey’s honestly significant difference test (α = 0.05). Compared with the RMS values of the conventional method (11.88 ± 2.69 µm) and the 3D-printing method at 0 degrees (12.14 ± 2.38 µm), the RMS values were significantly high at 90 degrees (16.46 ± 2.39 µm) (p < 0.05). Likewise, there was a significant difference in the change in volume between the groups (p = 0.002), with a significantly higher volume change value at 90 degrees (1.74 ± 0.41 mm3) than in the conventional method (0.70 ± 0.15 mm3) (p < 0.05). A printing angle of 90 degrees is not recommended when interim crowns are fabricated using digital light processing 3D printing.

Details

Title
Comparison of Wear of Interim Crowns in Accordance with the Build Angle of Digital Light Processing 3D Printing: A Preliminary In Vivo Study
Author
Lee, Hakjun 1   VIAFID ORCID Logo  ; Son, Keunbada 2   VIAFID ORCID Logo  ; Du-Hyeong, Lee 1   VIAFID ORCID Logo  ; So-Yeun, Kim 1   VIAFID ORCID Logo  ; Kyu-Bok, Lee 1   VIAFID ORCID Logo 

 Department of Prosthodontics, School of Dentistry, Kyungpook National University, Daegu 41940, Korea 
 Advanced Dental Device Development Institute (A3DI), Kyungpook National University, Daegu 41940, Korea 
First page
417
Publication year
2022
Publication date
2022
Publisher
MDPI AG
e-ISSN
23065354
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2716486904
Copyright
© 2022 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.