Abstract
Background
Lymphocytic leukemia (LL) is a primary malignant tumor of hematopoietic tissue, which seriously affects the health of children and the elderly. The study aims to establish a new detection method for screening acute/chronic LL using time-resolved fluorescence immunoassay (TRFIA) via quantitative detection of S100 calcium binding protein A8 (S100A8) and leucine-rich alpha-2-glycoprotein 1 (LRG1) in serum.
Methods
Here a sandwich TRFIA was optimized and established: Anti-S100A8/LRG1 caputre antibodies immobilized on 96-well plates captured S100A8/LRG1, and then banded together with the anti-S100A8/LRG1 detection antibodies labeled with Europium(III) (Eu3+)/samarium(III) (Sm3+) chelates. Finally time resolved fluorometry measured the fluorescence intensity.
Results
The sensitivity of S100A8 was 1.15 ng/mL(LogY = 3.4027 + 0.4091 × LogX, R2 = 0.9828, P < 0.001, dynamic range: 2.1–10,000 ng/mL), and 3.2 ng/mL for LRG1 (LogY = 3.3009 + 0.4082 × LogX, R2 = 0.9748, P < 0.001, dynamic range: 4.0–10,000 ng/mL). The intra-assay and inter-assay CVs were low, ranging from 5.75% to 8.23% for S100A8 and 5.30% to 9.45% for LRG1 with high specificity and affinity in serum samples. Bland–Altman plots indicated TRFIA and ELISA kits have good agreement in clinical serum samples. Additionally, the cutoff values for S100A8 and LRG1 were 1849.18 ng/mL and 588.08 ng/mL, respectively.
Conclusion
The present TRFIA method could be used for the quantitative detection of S100A8 and LRG1 in serum, and it has high sensitivity, accuracy and specificity. Clinically, this TRFIA method could be suitable for screening of LL via the quantitative detection of S100A8 and LRG1.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer