Abstract

Objective

To evaluate the reliability of CT measurements of muscle quantity and quality using variable CT parameters.

Materials and Methods

A phantom, simulating the L2–4 vertebral levels, was used for this study. CT images were repeatedly acquired with modulation of tube voltage, tube current, slice thickness, and the image reconstruction algorithm. Reference standard muscle compartments were obtained from the reference maps of the phantom. Cross-sectional area based on the Hounsfield unit (HU) thresholds of muscle and its components, and the mean density of the reference standard muscle compartment, were used to measure the muscle quantity and quality using different CT protocols. Signal-to-noise ratios (SNRs) were calculated in the images acquired with different settings.

Results

The skeletal muscle area (threshold, −29 to 150 HU) was constant, regardless of the protocol, occupying at least 91.7% of the reference standard muscle compartment. Conversely, normal attenuation muscle area (30–150 HU) was not constant in the different protocols, varying between 59.7% and 81.7% of the reference standard muscle compartment. The mean density was lower than the target density stated by the manufacturer (45 HU) in all cases (range, 39.0–44.9 HU). The SNR decreased with low tube voltage, low tube current, and in sections with thin slices, whereas it increased when the iterative reconstruction algorithm was used.

Conclusion

Measurement of muscle quantity using HU threshold was reliable, regardless of the CT protocol used. Conversely, the measurement of muscle quality using the mean density and narrow HU thresholds were inconsistent and inaccurate across different CT protocols. Therefore, further studies are warranted in future to determine the optimal CT protocols for reliable measurements of muscle quality.

Details

Title
Reliability of Skeletal Muscle Area Measurement on CT with Different Parameters: A Phantom Study
Author
Kim, Dong Wook  VIAFID ORCID Logo  ; Ha, Jiyeon  VIAFID ORCID Logo  ; Ko, Yousun  VIAFID ORCID Logo  ; Kim, Kyung Won  VIAFID ORCID Logo  ; Park, Taeyong  VIAFID ORCID Logo  ; Lee, Jeongjin  VIAFID ORCID Logo  ; Myung-Won You  VIAFID ORCID Logo  ; Kwon-Ha, Yoon  VIAFID ORCID Logo  ; Ji Yong Park  VIAFID ORCID Logo  ; Young, Jin Kee  VIAFID ORCID Logo  ; Hong-Kyu, Kim  VIAFID ORCID Logo 
Pages
624-633
Section
Technology, Experiment, and Physics
Publication year
2021
Publication date
Apr 2021
Publisher
The Korean Society of Radiology
ISSN
12296929
e-ISSN
20058330
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2728055765
Copyright
© 2021. This work is published under http://creativecommons.org/licenses/by-nc/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.