Full Text

Turn on search term navigation

© 2021. This work is published under http://creativecommons.org/licenses/by-nc/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Objective

We aimed to compare the aortic valve area (AVA) calculated using fast high-resolution three-dimensional (3D) magnetic resonance (MR) image acquisition with that of the conventional two-dimensional (2D) cine MR technique.

Materials and Methods

We included 139 consecutive patients (mean age ± standard deviation [SD], 68.5 ± 9.4 years) with aortic valvular stenosis (AS) and 21 asymptomatic controls (52.3 ± 14.2 years). High-resolution T2-prepared 3D steady-state free precession (SSFP) images (2.0 mm slice thickness, 10 contiguous slices) for 3D planimetry (3DP) were acquired with a single breath hold during mid-systole. 2D SSFP cine MR images (6.0 mm slice thickness) for 2D planimetry (2DP) were also obtained at three aortic valve levels. The calculations for the effective AVA based on the MR images were compared with the transthoracic echocardiographic (TTE) measurements using the continuity equation.

Results

The mean AVA ± SD derived by 3DP, 2DP, and TTE in the AS group were 0.81 ± 0.26 cm2, 0.82 ± 0.34 cm2, and 0.80 ± 0.26 cm2, respectively (p = 0.366). The intra-observer agreement was higher for 3DP than 2DP in one observer: intraclass correlation coefficient (ICC) of 0.95 (95% confidence interval [CI], 0.94–0.97) and 0.87 (95% CI, 0.82–0.91), respectively, for observer 1 and 0.97 (95% CI, 0.96–0.98) and 0.98 (95% CI, 0.97–0.99), respectively, for observer 2. Inter-observer agreement was similar between 3DP and 2DP, with the ICC of 0.92 (95% CI, 0.89–0.94) and 0.91 (95% CI, 0.88–0.93), respectively. 3DP-derived AVA showed a slightly higher agreement with AVA measured by TTE than the 2DP-derived AVA, with the ICC of 0.87 (95% CI, 0.82–0.91) vs. 0.85 (95% CI, 0.79–0.89).

Conclusion

High-resolution 3D MR image acquisition, with single-breath-hold SSFP sequences, gave AVA measurement with low observer variability that correlated highly with those obtained by TTE.

Details

Title
A New Method for Aortic Valve Planimetry with High-Resolution 3-Dimensional MRI and Its Comparison with Conventional Cine MRI and Echocardiography for Assessing the Severity of Aortic Valvular Stenosis
Author
Hae Jin Kim  VIAFID ORCID Logo  ; Choe, Yeon Hyeon  VIAFID ORCID Logo  ; Sung Mok Kim  VIAFID ORCID Logo  ; Kim, Eun Kyung  VIAFID ORCID Logo  ; Lee, Mirae  VIAFID ORCID Logo  ; Sung-Ji, Park  VIAFID ORCID Logo  ; Ahn, Joonghyun  VIAFID ORCID Logo  ; Carriere, Keumhee C  VIAFID ORCID Logo 
Pages
1266-1278
Section
Cardiovascular Imaging
Publication year
2021
Publication date
Aug 2021
Publisher
The Korean Society of Radiology
ISSN
12296929
e-ISSN
20058330
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2728060354
Copyright
© 2021. This work is published under http://creativecommons.org/licenses/by-nc/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.