Abstract

Background

A simple and sensitive gas chromatographic method was developed and validated for the simultaneous determination of methanol, ethanol, acetone, isopropyl alcohol, dichloromethane, N-hexane, ethyl acetate, tetrahydrofuran, and N,N-diisopropyl ethyl amine in Paclitaxel. A chromatographic separation was done on DB-624 column, 30 m length × 0.53 mm ID, and film thickness 3 μm, using a flame ionization detector (FID) with gradient column oven temperature program. The injection was carried out in split mode, with a split ratio of 5:1. A mixture of N-methyl-2-pyrrolidinone (contains 1% piperazine) and water in the ratio of 80:20 (v/v) was selected as a diluent to obtain good sensitivity along with the recovery.

Results

The developed gas chromatographic method offers symmetric peak shape, good resolution of more than 2.0 between the solvent peaks, and the relative standard deviation for replicate injections of all the solvents were found to be not more than 15.0% with reasonable retention time for all the solvents. The limit of detection for methanol, ethanol, acetone, isopropyl alcohol, dichloromethane, N-hexane, ethyl acetate, tetrahydrofuran, and N,N-diisopropyl ethyl amine was found to be 304.69 ppm, 497.98 ppm, 498.99 ppm, 504.49 ppm, 61.81 ppm, 30.07 ppm, 505 ppm, 73.05 ppm, and 2.09 ppm, respectively. Limit of quantitation of methanol, ethanol, acetone, isopropyl alcohol, dichloromethane, N-hexane, ethyl acetate, tetrahydrofuran, and N,N-diisopropyl ethyl amine was found to be 89.62 ppm, 146.47 ppm, 146.76 ppm, 148.38 ppm, 18.18 ppm, 8.84 ppm, 148.53 ppm, 21.49 ppm, and 0.62 ppm, respectively. Precision was found to be satisfactory. Linear in the range of LOQ to 150% level for all the solvents, and accuracy along with robustness, is performed, and acceptable results were obtained.

Conclusion

The proposed method was demonstrated to be simple, sensitive, specific, linear, precise, accurate, and robust, hence can be used to determine the residual organic solvents in Paclitaxel drug substance and drug product.

Details

Title
Determination of residual solvents in paclitaxel by headspace gas chromatography
Author
Noorbasha, Khaleel 1   VIAFID ORCID Logo  ; Shaik, Abdul Rahaman 2 

 Acharya Nagarjuna University, Department of Pharmaceutical Sciences, Guntur, India (GRID:grid.411114.0) (ISNI:0000 0000 9211 2181) 
 Nirmala College of Pharmacy, Department of Pharmacy, Guntur, India (GRID:grid.411114.0) 
Pages
40
Publication year
2021
Publication date
Dec 2021
Publisher
Springer Nature B.V.
ISSN
23147245
e-ISSN
23147253
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2729530646
Copyright
© The Author(s) 2021. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.