© 2022 Kripfgans et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Introduction

High frequency ultrasound has shown as a promising imaging modality to evaluate peri-implant tissues. It is not known if the ultrasound imaging settings might influence ultrasound’s ability to differentiate implant structures. The aim of this benchtop study was to evaluate the dependence of ultrasound on imaging angles and modes to measure implant geometry-related parameters.

Methods

A clinical ultrasound scanner (ZS3, Mindray) with an intraoral probe (L30-8) offering combinations of harmonic and compound imaging modes was employed for imaging 16 abutments and 4 implants. The samples were mounted to a micro-positioning system in a water tank, which allowed a range of -30 to 30-degree imaging angles in 5-degree increment between the probe and samples. The abutment angle, implant thread pitch and depth were measured on ultrasound, compared to the reference readings. The errors were computed as a function of the image angles and modes. All samples were replicated 3 times for 3 image modes and 11 image angles, thus resulting in 2,340 images.

Results

The mean errors of ultrasound to estimate 16 abutment angles, compared to the reference values, were between -1.8 to 2.7 degrees. The root mean squared error (RMSE) ranged from 1.5 to 4.6 degrees. Ultrasound significantly overestimated the thread pitch by 26.1 μm to 36.2 μm. The error in thread depth measurements were in a range of -50.5 μm to 39.6 μm, respectively. The RMSE of thread pitch and depth of the tested 4 implants was in a range of 34.7 to 56.9 μm and 51.0 to 101.8 μm, respectively. In most samples, these errors were independent of the image angle and modes.

Conclusions

Within the limitations of this study, high-frequency ultrasound was feasible in imaging abutments and implant fixtures independent of scanning angle within ±30° of normal incidence and for compounding and non-compounding-based imaging modes.

Details

Title
Ultrasound insonation angle and scanning imaging modes for imaging dental implant structures: A benchtop study
Author
Kripfgans, Oliver D; Nikhila Devi Goli; Majzoub, Jad; Rafael Amorim Cavalcanti De Siqueira  VIAFID ORCID Logo  ; Soki, Fabiana; Chan, Hsun-Liang  VIAFID ORCID Logo 
First page
e0270392
Section
Research Article
Publication year
2022
Publication date
Nov 2022
Publisher
Public Library of Science
e-ISSN
19326203
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2741301929
Copyright
© 2022 Kripfgans et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.