© 2023 Reezigt et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: https://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, reproduction and adaptation in any medium and for any purpose provided that it is properly attributed. For attribution, the original author(s), title, publication source (PeerJ) and either DOI or URL of the article must be cited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Background

Experienced assessors show good intra-rater reproducibility (within-session and between-session agreement and reliability) when using an algometer to determine pressure pain thresholds (PPT). However, it is unknown whether novice assessors perform equally well. This study aimed to determine within and between-session agreement and reliability of PPT measurements performed by novice assessors and explored whether these parameters differed per assessor and algometer type.

Methods

Ten novice assessors measured PPTs over four test locations (tibialis anterior muscle, rectus femoris muscle, extensor carpi radialis brevis muscle and paraspinal muscles C5-C6) in 178 healthy participants, using either a Somedic Type II digital algometer (10 raters; 88 participants) or a Wagner Force Ten FDX 25 digital algometer (nine raters; 90 participants). Prior to the experiment, the novice assessors practiced PPTs for 3 h per algometer. Each assessor measured a different subsample of ~9 participants. For both the individual assessor and for all assessors combined (i.e., the group representing novice assessors), the standard error of measurement (SEM) and coefficient of variation (CV) were calculated to reflect within and between-session agreement. Reliability was assessed using intraclass correlation coefficients (ICC1,1).

Results

Within-session agreement expressed as SEM ranged from 42 to 74 kPa, depending on the test location and device. Between-session agreement, expressed as SEM, ranged from 36 to 76 kPa and the CV ranged from 9–16% per body location. Individual assessors differed from the mean group results, ranging from −55 to +32 kPa or from −9.5 to +6.6 percentage points. Reliability was good to excellent (ICC1,1: 0.87 to 0.95). Results were similar for both types of algometers.

Conclusions

Following 3 h of algometer practice, there were slight differences between assessors, but reproducibility in determining PPTs was overall good.

Details

Title
Novice assessors demonstrate good intra-rater agreement and reliability when determining pressure pain thresholds; a cross-sectional study
Author
Reezigt, Roland R; Slager, Geranda E C; Coppieters, Michel W; Scholten-Peeters, Gwendolyne G M
Publication year
2023
Publication date
Jan 4, 2023
Publisher
PeerJ, Inc.
e-ISSN
21678359
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2760690926
Copyright
© 2023 Reezigt et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: https://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, reproduction and adaptation in any medium and for any purpose provided that it is properly attributed. For attribution, the original author(s), title, publication source (PeerJ) and either DOI or URL of the article must be cited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.