Abstract

Introduction

We performed a retrospective review of a radar-based breast lesion localisation system (Savi Scout; Cianna Medical, Merit Medical Systems, Inc., South Jordan [UT], United States) in a Chinese population.

Methods

Placement success (final target-to-reflector distance <10 mm), retrieval success, margin clearance, and re-excision rates were reviewed in the cases of 23 Chinese patients who underwent guided nonpalpable breast lesion excision from October 2019 to December 2020 using the system in a single institution.

Results

Twenty-three reflectors were placed under sonographic (n=13; 57%) or stereotactic (n=10; 44%) guidance to localise 23 target lesions. There was no delayed migration for the 20 reflectors placed before the day of surgery. Placement success was achieved in 21 (91%). Mean final target-to-reflector distance was 3 mm. Of the 23 lesions, two (9%) required alternative localisation owing to reflector distance ≥10 mm away from the target. Retrieval success was achieved in 22 (96%). Deactivation of a reflector was noted in one case. Of these 23 lesions, three were excised for therapeutic intent, of which one required re-excision due to close margins. There were no procedure-related complications.

Conclusion

This radar-based localisation system is a safe and effective device for guiding the excision of non-palpable breast lesions in a Chinese population. Its advantages, such as the fact that it causes minimal artefacts on magnetic resonance imaging, may render it a superior alternative in selected patients.

Alternate abstract:

引言

我們對一種基於雷達的乳腺病變定位系統(Savi Scout)進行一項華人回顧性研究。

方法

對2019年10月至2020年12月期間在單中心接受該系統引導式不可觸乳腺病灶切除術的23名華人患者有關植入成功率(最終目標與反射器距離<10 mm)、回取成功率、邊緣清除率和再切除率進行回顧。

結果

13個反射器(57%)於超聲引導下放置,10個反射器(43%)於立體定向引導下放置,共定位23個目標病灶。手術當天之前放置的20個反射器沒有延遲移位。21例(91%)放置成功。平均最終目標與反射器的距離為3 mm。由於反射器與距離目標間≥10 mm,2個病變(9%)需要其他方式定位。22例(96%)回取成功。1例反射器失活。23個病灶中,3例因治療目的而切除,其中1例因切緣較近須重新切除。沒有手術相關的併發症。

結論

這雷達系統安全有效,能引導華人乳房不可觸及病灶的切除。其優勢克服其他定位方法的一些限制,例如減少磁共振成像偽影,對特定患者可成為良好替代方案。

Details

Title
Radar Localisation of Non-palpable Breast Lesions in a Chinese Population: a Pilot Study
Author
Woo, S C; Wong, T; Chau, C M; Fung, W Y; Chan, RLS; Yung, AWT; Ma, JKF
First page
192
Publication year
2022
Publication date
Sep 2022
Publisher
Hong Kong Academy of Medicine
ISSN
22236619
e-ISSN
23074620
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2765339917
Copyright
© 2022. This work is published under https://creativecommons.org/licenses/by-nc-nd/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.