© 2023 Granja Domínguez et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Objective

To determine the concordance and statistical precision in gait velocity in people with multiple sclerosis (pwMS), measured with FeetMe® (insoles with pressure and motion sensors) compared with GAITRite® (classic reference system of gait analysis) in the timed 25-Feet Walk test (T25WT).

Methods

This observational, cross-sectional, prospective, single center study was conducted between September-2018 and April-2019 in pwMS aged 18–55 years, with Expanded Disability Status Scale (EDSS) 0–6.5 and relapse free ≥30 days at baseline. Primary endpoint was gait velocity. Secondary endpoints were ambulation time, cadence, and stride length assessment, while the correlation between gait variables and the clinical parameters of MS subjects was assessed as an exploratory endpoint.

Results

A total of 207 MS subjects were enrolled, of whom, 205 were considered in primary analysis. Most subjects were women (66.8%) and had relapsing-remitting MS (RRMS) (82.9%), with overall mean (standard deviation [SD]) age of 41.5 (8.0) year and EDSS 3.1 (2.0). There was a statistically significant (p<0.0001) and strong agreement (intra-class correlation coefficient (ICC) >0.830) in gait velocity, ambulation time and cadence assessment between FeetMe® and GAITRite®.

Conclusions

Agreement between devices was strong (ICC≥0.800). FeetMe® is the first validated wearable medical device that allows gait monitoring in MS subjects, being potentially able to assess disease activity, progression, and treatment response.

Details

Title
Study for the validation of the FeetMe® integrated sensor insole system compared to GAITRite® system to assess gait characteristics in patients with multiple sclerosis
Author
Anabel Granja Domínguez; Raúl Romero Sevilla; Alemán, Aurora  VIAFID ORCID Logo  ; Durán, Carmen  VIAFID ORCID Logo  ; Hochsprung, Anja; Navarro, Guillermo; Páramo, Cristina; Venegas, Ana; Lladonosa, Ana; Guillermo Izquierdo Ayuso  VIAFID ORCID Logo 
First page
e0272596
Section
Research Article
Publication year
2023
Publication date
Feb 2023
Publisher
Public Library of Science
e-ISSN
19326203
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2774826777
Copyright
© 2023 Granja Domínguez et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.