Answer
The first part of the recording (Fig. 1) shows continuous atrial fibrillation (AF) with biventricular pacing. After a single T‑wave is oversensed (1), the device automatically adjusts the sensitivity based on the amplitude of the preceding sense event [1]. This more sensitive setting leads to detection of AF in the ventricular channel, interpreted as ventricular fibrillation (VF) and abortion of ventricular pacing (2). After approximately 9 s, AF oversensing has stopped spontaneously, biventricular pacing is resumed and VF therapy is aborted (3).
Fig. 1 [Images not available. See PDF.]
After atrial fibrillation with biventricular pacing (upper panel), there is a sudden change initiated by ventricular oversensing (1). Subsequently, atrial fibrillation is sensed in the ventricular channel (2); note the synchrony between atrial and ventricular sense markers in the lower channel. There is a spontaneous abortion of atrial oversensing (3) and restoration of biventricular pacing
It is important to realise that, in contrast with the RV tip to ring configuration default depicted on the marker channel, this biventricular implantable cardioverter-defibrillator was programmed RV tip to RV coil. AF sensing in the RV lead is proven by the synchronous atrial and ventricular sensing in Fig. 1 (question) and Fig. 1 (answer). Near-field sensing of AF with the ventricular lead is explained by the use of the coil as anodal electrode for pacing and sensing with the proximal part located in the lower right atrium (Fig. 2).
Fig. 2 [Images not available. See PDF.]
Posterior-anterior chest X‑ray, showing the RA lead, a quadripolar LV lead and RV shock lead. Note the position of the proximal segment of the shock lead, which is most likely located in the lower right atrium at the level of the coronary sinus
In this case, the key problems were the position of the RV lead and the coil being used as a proximal sensing electrode. Once the device was reprogrammed to tip to ring, the problem was solved. Patient had a good clinical response to biventricular pacing and is doing well.
Although this is a rare phenomenon, the lesson to be learned is that, when we change the sensing vector to an integrated bipolar setting, we must make sure that the coil is well within the right ventricle.
Conflict of interest
B.M. van Gelder provides training and education for St. Jude Medical Nederland B.V., the Netherlands, and is clinical adviser for LivaNova, France. A.W.G.J. Oomen and F.A.L.E. Bracke declare that they have no competing interests.
References
1. http://www.pacingdefibrillation.com/en/specificities/icd-sensing/medtronic/medtronic-sensing-and-detecting-episodes. Accessed 1-5-2017
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© The Author(s) 2017. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Details
1 Catharina Hospital, Eindhoven, The Netherlands (GRID:grid.413532.2) (ISNI:0000 0004 0398 8384)