© 2023 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

There are several methods available to assess energy expenditure, all associated with inherent pros and cons that must be adequately considered for use in specific environments and populations. A requirement of all methods is that they must be valid and reliable in their capability to accurately measure oxygen consumption (VO2) and carbon dioxide production (VCO2). The purpose of this study was to evaluate the reliability and validity of the mobile CO2/O2 Breath and Respiration Analyzer (COBRA) relative to a criterion system (Parvomedics TrueOne 2400®, PARVO) with additional measurements to compare the COBRA to a portable system (Vyaire Medical, Oxycon Mobile®, OXY). Fourteen volunteers with a mean of 24 years old, body weight of 76 kg, and a VO2peak of 3.8 L∙min−1 performed four repeated trials of progressive exercises. Simultaneous steady-state measurements of VO2, VCO2, and minute ventilation (VE) by the COBRA/PARVO and OXY systems were conducted at rest, while walking (23–36% VO2peak), jogging (49–67% VO2peak), and running (60–76% VO2peak). Data collection was randomized by the order of system tested (COBRA/PARVO and OXY) and was standardized to maintain work intensity (rest to run) progression across study trials and days (two trials/day over two days). Systematic bias was examined to assess the accuracy of the COBRA to PARVO and OXY to PARVO across work intensities. Intra- and inter-unit variability were assessed with interclass correlation coefficients (ICC) and a 95% limit of agreement intervals. The COBRA and PARVO produced similar measures for VO2 (Bias ± SD, 0.01 ± 0.13 L·min−1; 95% LoA, (−0.24, 0.27 L·min−1); R2 = 0.982), VCO2 (0.06 ± 0.13 L·min−1; (−0.19, 0.31 L·min−1); R2 = 0.982), VE (2.07 ± 2.76 L·min−1; (−3.35, 7.49 L·min−1); R2 = 0.991) across work intensities. There was a linear bias across both the COBRA and OXY with increased work intensity. The coefficient of variation for the COBRA ranged from 7 to 9% across measures for VO2, VCO2, and VE. COBRA was reliable across measurements for VO2 (ICC = 0.825; 0.951), VCO2 (ICC = 0.785; 0.876), and VE (ICC = 0.857; 0.945) for intra-unit reliability, respectively. The COBRA is an accurate and reliable mobile system for measuring gas exchange at rest and across a range of work intensities.

Details

Title
Using a Contemporary Portable Metabolic Gas Exchange System for Assessing Energy Expenditure: A Validity and Reliability Study
Author
McClung, Holly L 1   VIAFID ORCID Logo  ; Tharion, William J 1 ; Walker, Leila A 1 ; Rome, Maxwell N 2 ; Hoyt, Reed W 3 ; Looney, David P 1 

 Military Performance Division, US Army Research Institute of Environmental Medicine (USARIEM), 10 General Green Ave., Natick, MA 01760, USA 
 Oak Ridge Institute for Science and Education (ORISE), 1299 Bethel Valley Rd., Oak Ridge, TN 37830, USA 
 Military Nutrition Division, US Army Research Institute of Environmental Medicine (USARIEM), 10 General Green Ave., Natick, MA 01760, USA 
First page
2472
Publication year
2023
Publication date
2023
Publisher
MDPI AG
e-ISSN
14248220
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2785236732
Copyright
© 2023 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.