Introduction
Functional lung imaging has proven to offer valuable information for surveillance and early detection of various pulmonary diseases [1–4]. Techniques such as phase-resolved functional lung (PREFUL) MRI [5] avoid potential risks of contrast media, increase comfort and are already implemented clinically as they allow the calculation of ventilation- and perfusion-weighted maps enabling detection and monitoring various pulmonary diseases [6–9].
Recently, a perfusion quantification technique for PREFUL MRI was developed with good agreement to dynamic contrast-enhanced MRI in patients with chronic obstructive pulmonary disease (COPD) and chronic thromboembolic pulmonary hypertension in a single center setting [10]. However, a major challenge of medical imaging in general and especially of functional MRI, is the limited comparability across different sites, protocols and software. For instance, influences could be caused by the necessity of gadolinium contrast (GD) injection when these images serve as a primary endpoint of a study. Comparability might also be limited in multi-site studies with MR scanners using differing field-strengths. Furthermore, specific advantages of the imaging sequences could be exploited, for example higher SNR of a balanced steady-state free precession (bSSFP) sequence or fewer off-resonance artifacts of a spoiled gradient echo (SPGR) sequence at higher field-strengths. This would bring more flexibility across different sites and protocols. These influences on the signal are highly complex, which challenges absolute quantification of functional parameters as they are based on stable reference signals [11, 12].
Therefore, the aim of this study is the evaluation of the influences of previously applied gadolinium-based contrast media, a differing field-strength and a differing imaging sequence for the perfusion quantification of PREFUL MRI.
Materials and methods
Participants
In this study, datasets of 74 participants from four cohorts were evaluated (Fig 1). All datasets were analyzed retrospectively. The study was approved by the local ethics committee ‘Ethikkommission der Medizinischen Hochschule Hannover’, and written informed consent was obtained from all study participants.
[Figure omitted. See PDF.]
Spoiled gradient echo (SPGR), balanced steady-state free precession (bSSFP), chronic obstructive pulmonary disease (COPD), pulmonary arterial hypertension (PAH), cystic fibrosis (CF).
The reproducibility cohort was imaged at baseline and after two weeks. The cohort consisted of 26 patients with COPD (7 females, mean age 64 ± 7 years).
The gadolinium cohort was imaged before and after the administration of gadolinium-based contrast media and consisted of 11 patients with chronic obstructive pulmonary disease (COPD, 5 females, mean age 68 ± 6 years), 2 patients with pulmonal arterial hypertension (PAH, 1 female, mean age 65 ± 4 years) and one female patient with asthma (age 78 years).
The field-strength cohort was imaged at 1.5 Tesla (T) and 3T scanners and consisted of 12 healthy participants and 4 patients with cystic fibrosis (8 females, mean age 26 ± 7 years).
The sequence cohort was imaged with a SPGR and with a bSSFP sequence and consisted of 7 healthy participants and 11 patients with COPD (6 females, mean age 51 ± 18 years).
Inclusion criteria were as follows: COPD patients ≥ 40 years with stable cardiovascular function and a smoking history of ≥ 10 years were included. PAH patients ≥ 18 years with suspected pulmonary hypertension in transthoracic echocardiography were included. Patients with known cystic fibrosis were included. Healthy participants ≥ 18 years of age were included. Participants with contraindications to undergo MRI (e.g., due to pregnancy, claustrophobia or a pacemaker) or with lung infection or participation in other trials within the last 4 weeks were excluded.
Imaging
The participants were scanned in head-first supine position to image a coronal slice at the level of the carina during free breathing. Parallel imaging with generalized autocalibrating partial parallel acquisition (GRAPPA) with acceleration factor 2 was performed. All images were interpolated to the final in-plane resolution of 256 × 256 pixels before reconstruction. The non-uniform intensity profiles of the applied receiver coils were corrected using the sensitivity profiles of the surface and body coil before further analysis. The optimal flip angle (Ernst angle) was used together with minimized time of repetition (TR) and echo time (TE) to maximize the MR-signal and to reduce scan time in each protocol. For further details please see following subsections.
Reproducibility cohort.
The participants were imaged at 1.5T using a SPGR sequence (MAGNETOM Avanto n = 26, Siemens Healthineers, Erlangen, Germany). The follow-up measurement was acquired after two weeks using identical parameters: 50 cm x 50 cm field of view, 128 x 128 matrix size, 15 mm slice thickness, 0.82 ms TE, 3 ms TR, 5° flip angle and 1500 Hz/pixel bandwidth.
Gadolinium cohort.
The participants were imaged at 1.5T using a SPGR sequence (MAGNETOM Aera n = 6 or MAGNETOM Avanto n = 8, Siemens Healthineers, Erlangen, Germany). The 1st measurement was acquired before an MR-angiography with i.v. application of 0.1 mmol/kg gadobutrol. The 2nd measurement started a mean 43 ± 1.9 s after the application. The SPGR sequence parameters were: 50 cm x 50 cm field of view, 128 x 128 matrix size, 15 mm slice thickness, 0.82 ms TE, 3 ms TR, 5° flip angle and 1500 Hz/pixel bandwidth.
Field-strength cohort.
The participants were imaged at 1.5T and 3T scanners using a SPGR sequence (MAGNETOM Avanto and MAGNETOM Skyra, Siemens Healthcare, Erlangen, Germany) within 3 days. The examinations were performed at 1.5T with the following parameters: 50 cm x 50 cm field of view, 128 x 128 matrix size, 15 mm slice thickness, 0.82 ms TE, 3 ms TR, 5° flip angle and 1500 Hz/pixel bandwidth. Differing at 3T, the TE was 0.74, the TR was 1.83 and the flip angle was 4°.
Sequence cohort.
The participants were imaged at 1.5T (MAGNETOM Avanto, Siemens Healthcare, Erlangen, Germany). A SPGR sequence was used for the 1st and a bSSFP sequence was used for the 2nd measurement. The SPGR sequence parameters were: 50 cm x 50 cm field of view, 128 x 128 matrix size, 15 mm slice thickness, 0.82 ms TE, 3 ms TR, 5° flip angle and 1500 Hz/pixel bandwidth. Differing for the bSSFP sequence, the TE was 0.64 ms, the TR was 1.41 ms, the flip angle was 35° and the bandwidth was 2055 Hz/pixel.
Image registration
Linear and diffeomorphic image registration was performed by advanced normalization tools (ANTs) with a group-oriented registration scheme (GOREG) to achieve a uniform respiratory position for each image [13, 14]. The first 20 images of each sequence were included to analyze the signal decay towards the steady-state for quantification [10]. Furthermore, the corresponding series in each cohort were registered to allow for voxel-wise comparison between baseline and follow-up, pre and post gadolinium, 1.5T and 3T and SPGR and bSSFP sequences, respectively. Image-guided filtering was applied to all images of every dataset using an edge-preserving filter [15]. The images were segmented automatically using a convolutional neuronal network including the entire lung [16].
Quantified perfusion mapping
As described previously [5], respiratory signal changes were excluded from the analysis by the application of a high pass filter at 0.75 Hz. The cardiac phases were estimated by piecewise fitting the averaged signal of a region of interest within the aorta to a cosine function. The images were then sorted by their phase and interpolated to an equidistant time grid. The signal difference between the time with the highest and the lowest signal of the time series in the parenchyma ROI was considered as the perfusion signal (Q) caused by flow-related enhancement [5].
Perfusion was quantified (QQuant) by estimating the proton density and considering the median height of the signal decay towards the steady-state [10]. For this, the signals at several time points (t) of the time series were extracted to estimate the proportion of spins being exchanged per heart beat (exchange fraction) and the proportion of blood per voxel (blood fraction), which were combined together with the time interval between two heartbeats (T) and additional scaling factors (S1 Fig):The voxel volume is cancelled out. Q [a.u.] is the voxelwise signal change during the heart cycle. S0par [a.u.] is the voxelwise signal in a mid-respiratory parenchyma ROI extrapolated at time = 0. S1par # [a.u.] is the median signal in a mid-respiratory parenchyma ROI during the steady state. S0ves # [a.u.] is the median signal of vessel voxels extrapolated at time = 0 (99th percentile of voxels inside the lung ROI including vessels). T [s] is the time between two heart beats. Values indicated with # are determined globally for the whole imaging slice. To stay consistent with literature [17], the perfusion is quantified in ml/min/100 ml and therefore multiplied by 100. The factor of ½ accounts for the opposing flow directions of arterial and venous blood.
Statistics
Statistical analysis was performed using MATLAB R2020b (The MathWorks, Natick, MA). Shapiro-Wilk test revealed no normal distribution of all functional parameters. Non-parametric paired Wilcoxon signed-rank test was used to assess significant differences of QQuant, Q, S0par, S0ves, S1par and the heart frequency between the corresponding acquisitions. Differences between parameter medians with P < 0.05 were considered as significant. Bland-Altman plots were created and voxelwise Pearson correlation was calculated to compare the quantified perfusion (QQuant) between both series in each cohort. The relative distribution of the quantified perfusion parameter was evaluated by histogram analyses. To ensure visual comparability, the perfusion values were normalized by dividing each value by the 95th percentile of the values. Skewness and kurtosis were calculated for each histogram after excluding zeros and values over 1 to increase robustness against extreme outliers. Bin-range of the histograms was set to 0.01. Additionally, the value distributions of the normalized quantified perfusion of both corresponding series were evaluated using a two-sample Kolmogorov-Smirnov test.
Results
Fig 2 shows representative quantified perfusion maps of each cohort. First of all, the heart frequency presented no significant change between the corresponding series in any cohort (S2 Table).
[Figure omitted. See PDF.]
In the reproducibility cohort a 52-year-old female with COPD and minor changes after two weeks. In the gadolinium cohort a 64-year-old female with COPD and increased values after GD. In the field-strength cohort an 18-year-old female with CF lower values at 3T. In the sequence cohort a 50-year-old-female patient with COPD and fewer low values.
Reproducibility cohort
No significant change of quantified perfusion was observed within the reproducibility cohort (P = 0.73) (Table 1). The Bland-Altman plot showed a mean difference of 0.21 ml/min/100ml with a standard deviation of 8.20 ml/min/100ml and a mean average of 19.69 ml/min/100ml between both series (S2 Fig). Voxelwise Pearson correlation of QQuant between both series was significant with a correlation coefficient of 0.8 (Fig 3 and Table 2). No change was detected for the value distribution in the normalized histogram analysis comparing skewness and kurtosis (P ≥ 0.41) or in the probability distribution using a two-sample Kolmogorov-Smirnov test (Fig 4 and S1 Table).
[Figure omitted. See PDF.]
Patients are marked in different colors. Extreme values represent central lung vessels which were included in A, C, E, G. Values over 500 ml/min/100ml were excluded in B, D, F, H to evaluate only the distribution within the lung parenchyma. The reproducibility cohort is depicted in A&B, the gadolinium cohort in C&D, the field-strength cohort in E&F and the sequence cohort in G&H.
[Figure omitted. See PDF.]
Similar distribution of QQuant values at baseline and follow-up and before and after gadobutrol injection is seen in A and B. Major differences of the value distribution are visible between 1.5T and 3T with a downshift of the values at 3T in C. Smaller differences between the value distributions of both sequences with relatively fewer small values for the bSSFP are depicted in D.
[Figure omitted. See PDF.]
[Figure omitted. See PDF.]
Gadolinium cohort
The quantified perfusion increased significantly from 29.1 pre GD to 67.5 ml/min/100ml post GD (P < 0.01) (Table 1). While the perfusion amplitude Q and the S0-signals in the parenchyma and the vessel decreased significantly after the administration of gadolinium, the steady-state signal S1par increased significantly from 157.5 to 195.8 (S2 Table). The Bland-Altman plot showed a mean difference of -36.0 ml/min/100ml with a standard deviation of 34.9 ml/min/100ml and a mean average of 46.1 ml/min/100ml between both series (S2 Fig). Voxelwise Pearson correlation of QQuant between both series was significant with a correlation coefficient of 0.7 (Fig 3 and Table 2). No change was detected for the value distribution in the normalized histogram analysis with a decrease of the skewness from 1.8 to 1.3 (P = 0.10) and a decrease of the kurtosis from 6.6 to 4.9 (P = 0.12). Also, the two-sample Kolmogorov-Smirnov test revealed no differing probability distribution of the histogram values of both series (Fig 4 and S1 Table).
Field-strength cohort
No significant change of the quantified perfusion was shown (P = 0.20) (Table 1). While the perfusion amplitude Q and the S0-signal in the parenchyma and the steady-state signal in the parenchyma decreased significantly at 3T, the S0-signal in the vessel increased significantly from 1496.7 to 2265.5 (S2 Table). The Bland-Altman plot showed a mean difference of 3.7 ml/min/100ml with a standard deviation of 85.7 ml/min/100ml and a mean average of 81.9 ml/min/100ml between both series (S2 Fig). Voxelwise Pearson correlation of QQuant was significant with a correlation coefficient of 0.8 (Fig 3 and Table 2). In the normalized histogram analysis, a pronounced difference between both series is visible with a downshift of the perfusion values at 3T (Fig 4). The skewness and kurtosis increased significantly from 1.8 to 2.2 and from 7.0 to 8.3. In accordance, the two-sample Kolmogorov-Smirnov test revealed a significantly differing probability distribution of the histogram values of both series (S1 Table).
Sequence cohort
In the sequence cohort, the quantified perfusion decreased significantly from 35.9 using SPGR to 23.7 ml/min/100ml using bSSFP (P < 0.01) (Table 1). Also, all other parameters decreased significantly using the bSSFP (S2 Table). The Bland-Altman plot showed a mean difference of 24.3 ml/min/100ml with a standard deviation of 29.3 ml/min/100ml and a mean average of 38.2 ml/min/100ml between both series (S2 Fig). Voxelwise Pearson correlation of QQuant was significant with a correlation coefficient of 0.7 (Fig 3 and Table 2). In the normalized histogram analysis, a small difference between both series is visible with an upshift of the perfusion values in the bSSFP sequence (Fig 4). The skewness and kurtosis decreased significantly from 1.8 to 1.4 and from 6.8 to 5.1 and the two-sample Kolmogorov-Smirnov test revealed a significantly differing probability distribution of the histogram values of both series (S1 Table).
Discussion
Our results indicated high regional correlation of the quantified perfusion parameters between all corresponding series. However, injection of gadobutrol, changing the field-strength or the sequence leads to significant alterations of the median values and/or the value distribution. In contrast, the reproducibility cohort with identical settings revealed quite stable median values together with a similar value distribution after two weeks.
The regional correlations of QQuant showed significant results supported by the vast quantity of the used data points. The different slopes and spread of extremely high values in Fig 3, which represent central vessels, might be the result of differing anatomy, central flow velocities and heart function between the patients or could display a systematic error of the method. After excluding values >500 ml/min/100ml, the correlation coefficients were lower (Fig 3). A reason for this could be a more accurate registration of regions with high contrasts such as vessels allowing better voxel-matching resulting in higher correlations. Nevertheless, all cohorts presented markedly deviations of the slope and high intercepts when correlating both corresponding series.
Only the reproducibility and gadolinium cohorts presented the same probability distribution in the two-sample Kolmogorov-Smirnov test and no significant change of the skewness and kurtosis. This is likely due to the use of the identical sequence parameters and field-strength. The application of gadolinium-based contrast agents and the following shortening of the T1-time led to a significant increase of the steady-state signal. Together with the signal decrease of the signals at the beginning of the sequence (t = 0), significantly increased quantified perfusion values were calculated. However, it is important to note, that the S0-values are not measured values, but extrapolated from the signal decay. Therefore, the decrease of the S0-values after GD could be the result of a lower extrapolation as a result of the increasing steady-state signal level (S1par). The same effect is present for S0ves, which is weaker, due to the velocity-dependence of the T1-time as described previously [10].
To achieve comparable values, while assuming a differing scan delay after injecting the contrast media, it would be necessary to use individual scaling factors as a function of time since injection and the individual elimination time, which is mostly dependent on the kidney function [18]. This factor would asymptotically decrease towards 1 during the elimination of the contrast over time. Otherwise, the T1-time would have to be considered in the perfusion model, which would necessitate T1-mapping for each patient. This approach could potentially also reduce variations due to differing T1 from varying hematocrit or oxygenation, although the latter is irrelevant at 1.5T [19].
The field-strength cohort showed no significant difference of the median height of the quantified perfusion parameter. Nevertheless, the interquartile range presented a markedly increase at 3T for all parameters, especially for the vessel signal at the beginning of the sequence (S0ves). Likely, the increasing spread of the values is caused by increasing field inhomogeneities occurring in lung MRI on interfaces such as vessel walls at increasing B0 [20]. The observed signal reduction within the parenchyma (S0par and S1par) at 3T could be the result of increasing susceptibility artifacts and spin-spin interactions shortening the T2*-relaxation time, while the signal in the vessel (S0ves) is increasing due to the gain of the SNR [20–23].
The perfusion amplitude (Q) decreases at the bSSFP sequence. However, the sequence parameters of the bSSFP could be further optimized to increase the overall lung signal and its variation over time [24, 25]. The high sensitivity of the bSSFP sequence to flow effects might lead to fewer zero-values within the perfusion maps causing lower skewness and kurtosis in the histogram analysis. Furthermore, the shorter TR and TE might improve sampling of the cardiac cycle enabling the detection of smaller signal differences. Taken together, the bSSFP sequence is very useful for functional imaging given its fast acquisition with sensitivity to flow effects, its high SNR and high contrast between fluids and other structures. However, its complex signal behavior with a mixed PD and contrast during the transient phase and contrast during steady-state causes a differing value distribution as the SPGR sequence [26].
A limitation of our study are the differing sequence parameters between 1.5T and 3T. A smaller flip angle of 4° was used at 3T to compensate for the increasing T1 of blood [27]. Nevertheless, changing field-strength, sequence parameters and hardware together makes an isolated evaluation of a single variable impossible.
Another limitation of our study is the variability of patient groups in each cohort, which could be affected differently. For instance, the COPD patients in the gadolinium cohort could be affected less due to their reduced blood volume and tissue density. There are however to few patient groups to make a subgroup analysis in this cohort. The field-strength cohort is composed of mostly healthy subjects and presents a relatively small deviation of the median values. This makes the difference in the value distribution even more remarkable since the large reproducibility cohort with 26 COPD patients showed a similar value distribution after two weeks. The same argument can be applied to the sequence cohort, which is composed of COPD patients and healthy participants, but also shows significant differences in median values and value distribution.
Taken together, the gadolinium, field-strength and sequence cohorts presented distortions of the quantified perfusion parameter, which may be challenging to correct. Perfusion quantification is influenced by various technical and physiological circumstances, such as inaccuracy of the estimated receive coil sensitivities, the angle of blood flow towards the imaging plane, image registration, field inhomogeneities or physiological perfusion variability [28]. Considering the similar value distribution, only the gadolinium cohort offers potential for a correction factor to compensate for the value increase, which however would necessitate additional measurements. The relatively stable results of the reproducibility cohort are noteworthy, as COPD patients tend to show more variability than healthy participants. This undermines the effects of gadolinium, field-strength and sequence on the perfusion quantification in the other cohorts. To date, PREFUL evaluations using a SPGR at 1.5T are validated best. Thus, the results show, that future multicenter studies should strictly adhere to an identical imaging protocol to generate comparable results across all sites. Also, parameters, that are derived from QQuant such as the perfusion defect percentage (QDP) would deviate due to the distortion of the value distributions [10].
In conclusion, the results indicate severe alterations and distortions of the quantified perfusion parameters after gadolinium administration and depending on the field-strength and on the applied sequence type.
Supporting information
S1 Fig. Theoretical model for the described perfusion quantification.
The blood fraction is estimated by dividing the signal of the S0par voxel by the S0ves. The exchange fraction is estimated by dividing the amplitude of the perfusion signal Q by the median signal decay in the parenchyma ROI between t = 0 and the steady-state (DeltaS).
https://doi.org/10.1371/journal.pone.0288744.s001
(TIF)
S2 Fig. Bland-Altman plots of QQuant of each cohort.
The average of QQuant of both corresponding series is on the x-axis and the difference is on the y-axis. The red solid line represents the mean bias and the dotted red lines the mean bias ± 1.96 standard deviation. In the reproducibility group (A), 26 patients with COPD in blue at baseline and after 2 weeks. In the gadolinium cohort (B), patients with PAH are marked in red, patients with COPD in blue and the patient with bronchial asthma in green. In the field-strength cohort (C), patients with cystic fibrosis are marked in green and the healthy participants in black. In the sequence cohort (D), patients with COPD are marked in blue and the healthy participants in black.
https://doi.org/10.1371/journal.pone.0288744.s002
(TIF)
S1 Table. Skewness and Kurtosis of the normalized QQuant values ± interquartile range in the histogram analysis.
Significant P-values are marked with *. The distribution of the normalized QQuant values of both series were compared by a two-sample Kolmogorov-Smirnov test.
https://doi.org/10.1371/journal.pone.0288744.s003
(DOCX)
S2 Table. Median values of the variables for perfusion quantification ± interquartile range.
Significant P-values are marked with *.
https://doi.org/10.1371/journal.pone.0288744.s004
(DOCX)
Acknowledgments
The authors would like to express their gratitude to the radiographers from the Department of Radiology for their support with the MR measurements and patient care.
Citation: Glandorf J, Brunzema F, Klimeš F, Behrendt L, Voskrebenzev A, Gutberlet M, et al. (2023) Influence of gadolinium, field-strength and sequence type on quantified perfusion values in phase-resolved functional lung MRI. PLoS ONE 18(8): e0288744. https://doi.org/10.1371/journal.pone.0288744
About the Authors:
Julian Glandorf
Roles: Conceptualization, Formal analysis, Software, Visualization, Writing – original draft
E-mail: [email protected]
Affiliations: Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany, Biomedical Research in Endstage and Obstructive Lung Disease Hannover (BREATH), Member of the German Centre for Lung Research (DZL), Hannover, Germany
ORICD: https://orcid.org/0000-0003-1927-9876
Fynn Brunzema
Roles: Formal analysis, Writing – review & editing
Affiliations: Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany, Biomedical Research in Endstage and Obstructive Lung Disease Hannover (BREATH), Member of the German Centre for Lung Research (DZL), Hannover, Germany
Filip Klimeš
Roles: Software, Writing – review & editing
Affiliations: Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany, Biomedical Research in Endstage and Obstructive Lung Disease Hannover (BREATH), Member of the German Centre for Lung Research (DZL), Hannover, Germany
Lea Behrendt
Roles: Writing – review & editing
Affiliations: Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany, Biomedical Research in Endstage and Obstructive Lung Disease Hannover (BREATH), Member of the German Centre for Lung Research (DZL), Hannover, Germany
Andreas Voskrebenzev
Roles: Software, Writing – review & editing
Affiliations: Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany, Biomedical Research in Endstage and Obstructive Lung Disease Hannover (BREATH), Member of the German Centre for Lung Research (DZL), Hannover, Germany
Marcel Gutberlet
Roles: Software, Writing – review & editing
Affiliations: Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany, Biomedical Research in Endstage and Obstructive Lung Disease Hannover (BREATH), Member of the German Centre for Lung Research (DZL), Hannover, Germany
Marius M. Wernz
Roles: Formal analysis, Writing – review & editing
Affiliations: Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany, Biomedical Research in Endstage and Obstructive Lung Disease Hannover (BREATH), Member of the German Centre for Lung Research (DZL), Hannover, Germany
ORICD: https://orcid.org/0009-0003-0452-8574
Robert Grimm
Roles: Writing – review & editing
Affiliation: MR Application Predevelopment, Siemens Healthcare GmbH, Erlangen, Germany
ORICD: https://orcid.org/0000-0002-1233-8215
Frank Wacker
Roles: Supervision
Affiliations: Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany, Biomedical Research in Endstage and Obstructive Lung Disease Hannover (BREATH), Member of the German Centre for Lung Research (DZL), Hannover, Germany
Jens Vogel-Claussen
Roles: Supervision, Writing – review & editing
Affiliations: Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany, Biomedical Research in Endstage and Obstructive Lung Disease Hannover (BREATH), Member of the German Centre for Lung Research (DZL), Hannover, Germany
1. Edelman RR, Hatabu H, Tadamura E, Li W, Prasad P V. Noninvasive assessment of regional ventilation in the human lung using oxygen–enhanced magnetic resonance imaging. Nat Med. 1996;2: 1236–1239. pmid:8898751
2. Berthezène Y, Vexler V, Clément O, Mühler A, Moseley ME, Brasch RC. Contrast-enhanced MR imaging of the lung: assessments of ventilation and perfusion. Radiology. 1992;183: 667–672. pmid:1584916
3. Roberts DA, Gefter WB, Hirsch JA, Rizi RR, Dougherty L, Lenkinski RE, et al. Pulmonary Perfusion: Respiratory-triggered Three-dimensional MR Imaging with Arterial Spin Tagging—Preliminary Results in Healthy Volunteers. Radiology. 1999;212: 890–895. pmid:10478262
4. Wielpütz MO, von Stackelberg O, Stahl M, Jobst BJ, Eichinger M, Puderbach MU, et al. Multicentre standardisation of chest MRI as radiation-free outcome measure of lung disease in young children with cystic fibrosis. J Cyst Fibros. 2018;17: 518–527. pmid:29805050
5. Voskrebenzev A, Gutberlet M, Klimeš F, Kaireit TF, Schönfeld C, Rotärmel A, et al. Feasibility of quantitative regional ventilation and perfusion mapping with phase‐resolved functional lung (PREFUL) MRI in healthy volunteers and COPD, CTEPH, and CF patients. Magn Reson Med. 2018;79: 2306–2314. pmid:28856715
6. Klimeš F, Voskrebenzev A, Gutberlet M, Kern AL, Behrendt L, Grimm R, et al. 3D phase-resolved functional lung ventilation MR imaging in healthy volunteers and patients with chronic pulmonary disease. Magn Reson Med. 2021;85: 912–925. pmid:32926451
7. Kaireit TF, Kern A, Voskrebenzev A, Pöhler GH, Klimes F, Behrendt L, et al. Flow Volume Loop and Regional Ventilation Assessment Using Phase-Resolved Functional Lung (PREFUL) MRI: Comparison With 129 Xenon Ventilation MRI and Lung Function Testing. J Magn Reson Imaging. 2021;53: 1092–1105. pmid:33247456
8. Voskrebenzev A, Kaireit TF, Klimeš F, Pöhler GH, Behrendt L, Biller H, et al. PREFUL MRI Depicts Dual Bronchodilator Changes in COPD: A Retrospective Analysis of a Randomized Controlled Trial. Radiol Cardiothorac Imaging. 2022;4. pmid:35506142
9. Zanette B, Schrauben EM, Munidasa S, Goolaub DS, Singh A, Coblentz A, et al. Clinical Feasibility of Structural and Functional MRI in Free‐Breathing Neonates and Infants. J Magn Reson Imaging. 2022;55: 1696–1707. pmid:35312203
10. Glandorf J, Klimeš F, Behrendt L, Voskrebenzev A, Kaireit TF, Gutberlet M, et al. Perfusion quantification using voxel-wise proton density and median signal decay in PREFUL MRI. Magn Reson Med. 2021;86: 1482–1493. pmid:33837557
11. Bloem JL, Reijnierse M, Huizinga TWJ, van der Helm-van Mil AHM. MR signal intensity: staying on the bright side in MR image interpretation. RMD Open. 2018;4: e000728. pmid:29955387
12. Gulani V, Seiberlich N. Quantitative MRI: Rationale and Challenges. 2020. pp. xxxvii–li.
13. Avants BB, Tustison NJ, Song G, Cook PA, Klein A, Gee JC. A reproducible evaluation of ANTs similarity metric performance in brain image registration. Neuroimage. 2011;54: 2033–2044. pmid:20851191
14. Voskrebenzev A, Gutberlet M, Kaireit TF, Wacker F, Vogel-Claussen J. Low-pass imaging of dynamic acquisitions (LIDA) with a group-oriented registration (GOREG) for proton MR imaging of lung ventilation. Magn Reson Med. 2017;78: 1496–1505. pmid:27859552
15. He K, Sun J, Tang X. Guided image filtering. IEEE Trans Pattern Anal Mach Intell. 2013;35: 1397–1409. pmid:23599054
16. Simonyan K, Zisserman A. Very Deep Convolutional Networks for Large-Scale Image Recognition. 2014.
17. Kjørstad Å, Corteville DMR, Fischer A, Henzler T, Schmid-Bindert G, Zöllner FG, et al. Quantitative lung perfusion evaluation using fourier decomposition perfusion MRI. Magn Reson Med. 2014;72: 558–562. pmid:24006024
18. Davies J, Siebenhandl-Wolff P, Tranquart F, Jones P, Evans P. Gadolinium: pharmacokinetics and toxicity in humans and laboratory animals following contrast agent administration. Arch Toxicol. 2022;96: 403–429. pmid:34997254
19. Barth M, Moser E. Proton NMR relaxation times of human blood samples at 1.5 T and implications for functional MRI. Cell Mol Biol (Noisy-le-grand). 1997;43: 783–91. Available: http://www.ncbi.nlm.nih.gov/pubmed/9298600 pmid:9298600
20. Soher BJ, Dale BM, Merkle EM. A Review of MR Physics: 3T versus 1.5T. Magn Reson Imaging Clin N Am. 2007;15: 277–290. pmid:17893049
21. Glandorf J, Klimeš F, Voskrebenzev A, Gutberlet M, Behrendt L, Crisosto C, et al. Comparison of phase-resolved functional lung (PREFUL) MRI derived perfusion and ventilation parameters at 1.5T and 3T in healthy volunteers. PLoS One. 2020;15: 1–14. pmid:33378373
22. Gai ND, Malayeri AA, Bluemke DA. Three-dimensional T1 and T2* mapping of human lung parenchyma using interleaved saturation recovery with dual echo ultrashort echo time imaging (ITSR-DUTE). J Magn Reson Imaging. 2017;45: 1097–1104. pmid:27726231
23. Hatabu H, Alsop DC, Listerud J, Bonnet M, Gefter WB. T2* and proton density measurement of normal human lung parenchyma using submillisecond echo time gradient echo magnetic resonance imaging. Eur J Radiol. 1999;29: 245–52. pmid:10399610
24. Bauman G, Pusterla O, Bieri O. Ultra-fast Steady-State Free Precession Pulse Sequence for Fourier Decomposition Pulmonary MRI. Magn Reson Med. 2016;75: 1647–1653. pmid:25965158
25. Bauman G, Pusterla O, Bieri O. Functional lung imaging with transient spoiled gradient echo. Magn Reson Med. 2019;81: 1915–1923. pmid:30346077
26. Scheffler K, Lehnhardt S. Principles and applications of balanced SSFP techniques. Eur Radiol. 2003;13: 2409–2418. pmid:12928954
27. Zhang X, Petersen ET, Ghariq E, De Vis JB, Webb AG, Teeuwisse WM, et al. In vivo blood T1 measurements at 1.5 T, 3 T, and 7 T. Magn Reson Med. 2013;70: 1082–1086. pmid:23172845
28. Behrendt L, Voskrebenzev A, Klimeš F, Gutberlet M, Winther HB, Kaireit TF, et al. Validation of Automated Perfusion‐Weighted Phase‐Resolved Functional Lung (PREFUL)‐MRI in Patients With Pulmonary Diseases. J Magn Reson Imaging. 2019; jmri.27027. pmid:31872556
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2023 Glandorf et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
Purpose
The purpose of this study is to evaluate the influences of gadolinium-based contrast agents, field-strength and different sequences on perfusion quantification in Phase-Resolved Functional Lung (PREFUL) MRI.
Materials and methods
Four cohorts of different subjects were imaged to analyze influences on the quantified perfusion maps: 1) at baseline and after 2 weeks to obtain the reproducibility (26 COPD patients), 2) before and after the administration of gadobutrol (11 COPD, 2 PAH and 1 asthma), 3) at 1.5T and 3T (12 healthy, 4 CF), and 4) with different acquisition sequences spoiled gradient echo (SPGR) and balanced steady-state free precession (bSSFP) (11 COPD, 7 healthy). Wilcoxon-signed rank test, Bland-Altman plots, voxelwise Pearson correlations, normalized histogram analyses with skewness and kurtosis and two-sample Kolmogorov-Smirnov tests were performed. P value ≤ 0.05 was considered statistically significant.
Results
In all cohorts, linear correlations of the perfusion values were significant with correlation coefficients of at least 0.7 considering the entire lung (P<0.01). The reproducibility cohort revealed stable results with a similar distribution. In the gadolinium cohort, the quantified perfusion increased significantly (P<0.01), and no significant change was detected in the histogram analysis. In the field-strength cohort, no significant change of the quantified perfusion was shown, but a significant increase of skewness and kurtosis at 3T (P = 0.01). In the sequence cohort, the quantified perfusion decreased significantly in the bSSFP sequence (P<0.01) together with a significant decrease of skewness and kurtosis (P = 0.02). The field-strength and sequence cohorts had differing probability distribution in the two-sample Kolmogorov-Smirnov tests.
Conclusion
We observed a high susceptibility of perfusion quantification to gadolinium, field-strength or MRI sequence leading to distortion and deviation of the perfusion values. Future multicenter studies should strictly adhere to the identical study protocols to generate comparable results.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer