Abstract

Background

The objective of this study was to investigate the effects of locking plugs and the biomechanical properties of a 3.5 mm 8-hole polyaxial locking plate in a fracture gap model. Our hypothesis was that locking plugs would increase the strength and stiffness of the construct. Twelve 3.5 mm 8-hole plates were used to evaluate two different construct designs (with locking plugs vs. without locking plugs) with validated bone substitutes in a 25 mm bridging osteosynthesis gap model. Each construct was subjected to a single cycle four-point bending load to failure using a servo-hydraulic testing machine. Bending stiffness, bending strength, and bending structural stiffness were calculated and compared using an unpaired Student´s t-test.

Results

The plating construct with locking plugs did not show any significant increase in terms of bending stiffness, bending strength, and bending structural stiffness compared to plating construct without locking plugs in a 25 mm gap fracture model during a single cycle four-point bending.

Conclusions

Under the conditions tested, filling empty plate holes with locking plugs in bridging osteosynthesis does not increase stiffness or strength of the plate-bone construct.

Details

Title
Do locking plugs improve implant strength? Biomechanical comparison of polyaxial locking constructs with and without locking plugs in a fracture gap model
Author
Viitanen, Joni; Quinn, Robert; Matthew Allenrt J. G. Broeckx; Bartkowiak, Tomasz; Haimel, Georg
Pages
1-8
Section
Research
Publication year
2023
Publication date
2023
Publisher
BioMed Central
e-ISSN
17466148
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2852064515
Copyright
© 2023. This work is licensed under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.