Introduction
Neuroendocrine tumours (NETs) are rare neoplasms that arise from epithelial cells with neuroendocrine features located primarily in the lungs, pancreas and gastrointestinal tissues [1, 2]. Clinical presentation and prognosis are extremely variable, from slow-growing well-differentiated disease to highly aggressive undifferentiated tumours [3]. The worldwide incidence is estimated to be around 50 cases per million inhabitants and has increased in the past decades, mostly due to better imaging [4–6].
Until the second decade of the 2000s, the reference product for scintigraphy examinations was 111In-DTPA-octreotide, as it was the only product registered worldwide, including Brazil. However, with the introduction of positron emission tomography (PET) scan with 68Ga-DOTA-octapeptides (DOTATATE, DOTATOC and DOTANOC), which provide a superior image pattern compared to those obtained by single photon emission computed tomography (SPECT), a new gold standard was established [7–12].
However, these methods have some drawbacks that affect their availability. In Brazil, 111In-DTPA-octreotide is supplied by a single institution (IPEN – Instituto de Pesquisas Energéticas e Nucleares), which is discontinuing its commercial production. Besides, it has high production costs, needs special cameras and collimators, and has long-term radioactivity that may harm patients and technicians [13]. While 68Ga-DOTA-octapeptides are still under regulatory approval by the Brazilian Health Agency (ANVISA – Agência Nacional de Vigilância Sanitária), the 68Ga labelling process requires a more costly and time-consuming process than the one with 99mTc. Furthermore, in most low- and middle-income countries (LMICs), facilities and specialised professionals able to produce and distribute radiopharmaceuticals are lacking. Finally, SPECT cameras availability is still considerably higher than PET scans worldwide. Therefore, identifying other cheaper and quickly produced radiopharmaceuticals is currently a global unmet need.
In previous studies, 99mTC-EDDA/HYNIC-TOC (99mTc marked octreotide) demonstrated similar efficacy to 111In-DTPA-octreotide (OCTREOSCAN®) in images from tumours that express somatostatin receptors (SSTRs), such as several types of NETs [14–18]. 99mTcEDDA/HYNIC-TOC has a higher affinity for SSTR2 and lower affinity for SSTR3 and SSTR5 [19]. Furthermore, 99mTc is also easily available in nuclear medicine facilities worldwide without special technologies. Thus, we conducted the Latin American Cooperative Oncology Group (LACOG) 0214 study, which aimed to evaluate the performance of scintigraphy using 99mTc-EDDA/HYNIC-TOC compared with 111In-DTPA-octreotide for the diagnosis and staging of patients with NETs.
Materials and methods
Study design and eligibility criteria
We planned a prospective diagnostic accuracy study in patients with biopsy-proven diagnoses of NETs from any location. The inclusion criteria were histological diagnosis of NETs in any stage, provided the patient had at least one visible lesion on computed tomography or magnetic resonance imaging. Patients should also have an indication of a 111In scan and be at least 18 years old. Patients should not have received somatostatin analogues in the month preceding the scan. An independent review board approved this research, and informed consent was obtained from all participants included in the study.
Study procedures
Clinical, demographic and pathological data were collected at baseline. All patients underwent vital signs assessment and basic biochemistry labs before the procedures. Patients then initially underwent the scintigraphy using 99mTc, followed by 111In scan in 2 days. Both scans were performed at the nuclear medicine facilities of the Brain Institute (BraIns). Two independent nuclear medicine physicians who were aware of the study evaluated all images.
The type of gamma camera used for the imaging process was the Forte Gamma Camera, manufactured by Philips. The matrix size was 64 × 6. SPECT/CT fusion imaging was conducted, allowing for enhanced anatomical localisation. The acquired images were analysed using the PEGASUS software. The Supplementary Information describes the protocol information.
Patients could undergo more than one pair of scans in case of re-evaluation after therapy initiation, as clinically indicated. Information was also collected regarding the number, location and intensity of lesions’ uptake.
The 99mTc-EDDA/HYNIC-TOC images were obtained using a gamma camera with low energy collimators, focused on the photopeak of the 99mTc (140.5 keV) with the symmetrical opening of 20% and injected activity of 10 mCi. We performed images of whole-body scans with SPECT 1 and 4 hours after radiopharmaceutical injection, in dorsal decubitus, with a 13 cm/minute velocity. The 111In-DTPA-octreotide images were performed in a gamma camera composed of collimators of medium energy, centred in the two photo peaks of the 111In (173 and 247 keV) with the symmetrical opening of 20% and activity injected of 6 mCi. Whole-body scan images were performed 4 and 24 hours after radiopharmaceutical injection, dorsal decubitus position, and 10 cm/minute velocity. SPECT was performed 24 hours after the injection of the radiopharmaceutical. Optional abdomen images were made after 48 hours of injection if bowel movements had interfered with image quality.
Study endpoints
The primary outcome was the diagnostic accuracy of the 99mTc-EDDA/HYNIC-TOC compared to 111In-DTPA-octreotide for diagnosis or staging of patients with NETs, irrespective of the primary site. Secondary outcomes included the number of NETs lesions visualised with radiopharmaceutical agents and safety.
Reports of acquisition-related medical complications evaluated adverse events and toxicity. Grading of adverse events was performed using the common terminology criteria for adverse events grading system.
Statistical analysis
We estimated that 39 tests would have to be performed to demonstrate sensitivity and specificity of at least 90% with an acceptable error of 10% points for 99mTC compared to 111In, with a significance level of 95% and a 10% loss and refusal rate. Categorical variables were described using count and percentage, and numerical variables were summarised using mean and SD. Binary diagnostic tests’ properties were estimated and are presented with 95% Cis, including sensitivity, specificity, positive LR and negative LR. Positive and negative predictive values were intentionally not calculated once the sample was composed of patients with the disease on conventional imaging. Thus, prevalence and predictive values are not estimable from our sample. For all comparisons of sensitivities, specificities and LRs, global comparisons using the Wald test were performed initially and followed by individual comparisons with multiple comparisons correction using the Holm [20] method. Differences in sensitivities and specificities, LRs, and their 95% Cis were calculated using the Roldán-Nofuentes and Sidaty-Regad methods [21, 22]. Analysis was performed in R, version 4.0.5, using the Compbdt package [23]. This study was registered at clinical trials with the identifier NCT02691078 and is being reported according to the last version of the Standards for Reporting Diagnostic Accuracy Studies (STARD) guidelines [24–26] from the Equator Network.
Results
Between May 2016 and July 2017, 14 scans were performed on 9 patients. Five patients performed both tests once, three patients twice, and one patient thrice. Unfortunately, the trial was stopped early due to under-enrolment. Patients were diagnosed from October 2007 to September 2016 and were examining disease re-evaluation after treatment. All nine patients were white, and five patients (55.6%) were female. The location of each patient’s primary tumour, baseline characteristics and comorbidities are described in Table 1.
Overall, 126 lesions were identified on the 14 scans. Two scans demonstrated an uncountable number of lesions, estimated at least 20. All 14 scans were positive using 99mTc, while 4 out of the 14 scans were false negatives using 111In contingency tables presenting 99mTC and 111In scans’ performance according to image acquisition are presented in Table 2.
Estimated sensitivities, specificities and LRs are presented in Table 3. Overall, 99mTc demonstrated higher sensitivity for identifying lesions in all images (p < 0.001, 95% CI for the difference 9.54% – 27.66%) and in liver-specific images (p < 0.001, 95% for the difference: 4.72% – 20.27%). However, the difference in sensitivities of 99mTc and 111In when only whole-body images were considered was not statistically significant (p = 0.06). In addition, no statistically significant differences in specificity were found.
Regarding comparing LRs, the global Wald test demonstrated a statistically significant difference in LRs of 99mTc and 111In when calculated for all images (p = 0.003) and liver-specific images (p = 0.025). However, after multiple comparison adjustments, only the difference of negative LR in liver-specific images was considered statistically significant (99mTc: 0.002, 95% CI 0.009 – 0.1 versus 111In: 0.19, 95% CI 0.12 – 0.42, p = 0.009).
Adverse events were detected in three patients after 111In scans: pruritus [1], constipation [1], vomiting and diarrhoea [1], all grade 1. Two patients experienced adverse events after 99mTc scans: pruritus [1] and vomiting and diarrhoea [1], all grade 1. No severe adverse events happened during the study.
In Figure 1, we have an example of the same patient evaluated by both imaging methods, showing a more significant number of foci of uptake with 99mTc-EDDA/HYNIC-TOC compared to 111In-DTPA-octreotide.
Discussion
The LACOG 0214 study was the first developed in Latin America on clinical validation of 99mTc-EDDA/HYNIC-TOC scan for diagnosing and staging patients with NET. Our results suggest that 99mTc, a cheaper and more available radiopharmaceutical, has an overall higher sensitivity than 111In, with better negative LR for liver metastasis in patients with NET.
Unfortunately, a lower-than-expected accrual resulted in early study termination and impacted our sample size. Causes for the under-enrolment cannot be ascertained. Consequently, our CI estimates for the differences in important parameters are comprehensive. For example, the 95% CI of the difference in sensitivity favouring 99mTC in all images is 9.54% – 27.66%, and in liver-specific images is 4.72% – 20.27%. However, we believe these imprecisions on the uncertainty are unfortunately inherent to the study of rare diseases such as NETs, which naturally imply smaller sample sizes and broader Cis. In this specific case, this limitation was addressed as conservatively as possible by calculating the 95% Cis with the Wald interval using the Bonett–Laplace adjustment, which was demonstrated to be superior to other methods in the previous studies [27].
Despite those limitations, the demonstration of a higher sensitivity overall, and especially a very low negative LR, of which the superior limit of the 95% CI is 0.1, supports that 99mTc may be an adequate alternative to 111In. Furthermore, it provides a better capacity to rule out somatostatin analogue uptake in the presence of liver images consistent with metastases in patients with NETs. This is even more important considering that 99mTc is associated with other beneficial characteristics, especially its availability.
Among the studies evaluating the 99mTc, a study including 173 patients with NETs who underwent 99mTc-EDDA/HYNIC-TOC scans as part of their clinical management found for pancreatic NETs a sensitivity of 94.6%, a specificity of 73.3%, and an accuracy of 90.1%. Gastrointestinal NETs found a sensitivity of 86.7%, specificity of 71.4%, and accuracy of 80.3% [28]. A study comparing 99mTc-EDDA/HYNIC-TOC with 111In-DTPA-octreotide for diagnosis of SSTR-expressing tumours in 41 patients revealed a higher sensitivity of 99mTc as compared with 111In as an imaging agent for the localisation of SSTR-expressing tumours [15].
Another study investigated 495 NETs patients and found an overall sensitivity of the 99mTc-EDDA/HYNIC-TOC of 80%, specificity of 92%, positive predictive value of 98%, negative predictive value of 47%, and accuracy of 82% [29].
A systematic review analysing studies coupled two by two comparing the peptide-based radiopharmaceuticals for PET and SPECT, 111In-DTPA-Octreotide, 99mTc-EDDA/HYNIC-TOC, 68Ga-DOTATATE/TOC and 64Cu-DOTATATE in the evaluation of NETs described true positivity rates of, respectively, 63.7%, 58.5%, 78.4% and 82.4% [30].
A literature review of 99mTc highlighted its wider availability, low cost and long decay compared to peptides labelled with 68 Ga. It concluded that the 99mTc could be proposed for a dosimetry evaluation of patients undergoing peptide receptor radionuclide therapy and for non-oncologic indications of radiolabelled somatostatin analogue (SSA) [8].
Our data reinforce that the diagnostic utility of the 99mTc-EDDA/HYNIC-TOC gain is very likely limited compared to newer but less available technologies such as 68Ga-DOTATATE PET/CT. Still, clinical validation and diagnostic accuracy of 99mTc and other markers are essential to LMICs, where the production and acquisition of radiopharmaceuticals are challenging [31, 32]. Thus, the possibility of using a more affordable and available radiopharmaceutical directly impacts care in these regions. Furthermore, even the demonstration of equivalent or non-inferior alternatives, considering other countries’ different legal constraints and regulations, impacts nuclear medicine services in LMICs due to their fragile supply chains [33]. For example, in Brazil, except for products with a half-life of fewer than 2 hours, production is a governmental monopoly, and the necessary resources are usually produced in small nuclear reactors that cannot have commercial radiopharmaceuticals. Thus, most Brazilian nuclear medicine services use foreign-produced radiopharmaceuticals and are dependent on external production, which leads to insecurity in the supply of radionuclides, as happened during the 2009 99Mo ‘crisis’ [34, 35].
Nevertheless, despite PET-CT being an increasingly used imaging method, it is also unavailable for many patients outside high-income countries [36]. For those countries which cannot afford such health expenditure, 99mTc use may be the most cost-effective strategy [33]. In the context of a rare disease, this gap may be even more significant since rare diseases such as NETs rarely have specifically designed guidance on healthcare systems and payers. Additionally, rare cancers are subject to natural limitations on clinical trial design and implementation, and this should be acknowledged when evaluating data for clinical use. Thus, a greater degree of uncertainty about therapies or diagnostic tests should also be accommodated for decision-making, as stated by the rare cancers Europe consensus panel [23].
Conclusion
Despite its early termination, the LACOG 0214 trial demonstrated a superior sensitivity of 99mTc-EDDA HYNIC-TOC compared to 111In-DTPA-octreotide in patients with NETs with a visible lesion on conventional imaging. The 99mTc was also superior regarding its negative LR when evaluating hepatic lesions, suggesting that 99mTc may be especially useful in ruling out liver metastases. Considering radiopharmaceutical acquisition is especially challenging for LMIC, the clinical validation of cheaper and more available molecules is a potential method to overcome this barrier in care.
Conflicts of interest statement
Dr Moriguchi-Jeckel and Dr Werutsky report grants from Financiadora de Estudos e Projetos (FINEP) for the conduct of the study. Drs R. Madke, Viana, Nabinger, and Fernandes report employment relationships with the RPH Group. Dr. Werutsky reports personal fees from AstraZeneca, Bayer, Beigene, Daiichi Sankyo, Genentech/Roche, GSK, Lilly, MSD, Novartis, Pfizer, Sanofi, and Seattle Genetics outside the submitted work.
The other authors declare that they have no conflict of interest.
Funding statement
This project obtained financial support from the Financier of Studies and Projects (FINEP), linked to the Ministry of Science, Technology and Innovation of Brazil, within the National Plan for Science and Technology framework. This project was also supported by the RPH Group, which provided the HYNIC-Octreotide used in this trial, and by the Latin American Cooperative Oncology Group (LACOG).
Ethical statement
All procedures performed in studies involving human participants followed the ethical standards of the institutional and national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards.
Author contribution
All authors contributed substantially to the project, as defined by the Committee of Medical Journal Editors, participating in the design of the study, the data interpretation, the article's writing, the critical review of the intellectual content, and the final approval of the version to be published.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© the authors; licensee e cancermedicalscience. 2023. This work is published under https://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
99mTc-EDDA/HYNIC-TOC is an easily available and cheaper radionuclide that could be used for somatostatin-receptor-based imaging of neuroendocrine tumours (NETs). We aimed to evaluate the diagnostic performance of 99mTc-EDDA/HYNIC-TOC compared to111In-DTPA-octreotide in patients (pts) with NETs. We performed a prospective diagnostic study including pts with biopsy-confirmed NET and at least one visible lesion at conventional imaging. Two independent nuclear medicine physicians evaluated pts who underwent 99mTc and 111In scans and images. The primary outcome was comparative diagnostic accuracy of 99mTc and 111In. Secondary outcomes include safety.
Nine pts were included and performed 14 paired scans. Overall, 126 lesions were identified. 99mTc demonstrated superior sensitivity both when all images were analysed (93.7, 95% CI 88.1% – 96.8% versus 74.8%, 95% CI 66.6 – 81.6%, p < 0.001) and when liver-specific images were analysed (97.8%, 95% CI 92.7% – 99.5% versus 85.1%, 95% CI 76.6% – 91.0%, p < 0.001). 99mTc was also associated with a lower negative likelihood ratio (LR) (0.002, 95% CI 0.009 – 0.1 versus 0.19, 95% CI 0.12 – 0.42, p = 0.009) when evaluating hepatic lesions. Adverse events happened in 3 pts after 111In and in 2 pts after 99mTc, all grade 1. The 99mTc demonstrated a higher sensitivity overall and a better negative LR in liver-specific images compared to 111In in pts with NETs. Our findings suggest that 99mTc is an alternative to 111In and is especially useful in ruling out liver metastases. NCT02691078.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer