Abstract
The purpose of this study was to investigate and compare the optical performance of five trifocal intraocular lenses (IOLs) following the ISO 11979-2 standards, analysing the impact of tilt and decentration. Five different diffractive trifocal IOLs were evaluated in this experimental study: Acriva Trinova (VSY-Biotechnology) (AT), FineVision HP (PhysIOL) (FVHP), AT LISA tri 839 MP (Zeiss) (ATLT), PanOptix TFNT00 IOL (Alcon) (PO), and Tecnis Synergy (J&J Vision) (TS). In-vitro optical quality analysis of them was performed with the Lambda PMTF system that has an aberration neutral cornea model (Lambda-X Ophthalmics). Measurements were performed on-axis, with 5º of IOL tilt and with 0.5 mm of IOL decentration using 543-nm monochromatic light. Finally, IOL dimensions and diffractive disk profile inspection was performed using the VisIOLA system (Rotlex). On-axis measurements showed a far through-focus MTF > 0.3 at 3 mm aperture, except for TS. FVHP and PO showed better far MTFs for larger apertures (3.75 mm and 4.5 mm) while AT showed good intermediate and near vision for such apertures. With 5º of IOL tilt, the better optical performance at all distances was found with AT for medium-sized pupils (3 mm) and an important reduction of MTF was found for ATLT and PO, especially in the intermediate focus. The induction of 0.5 mm of IOL decentration especially affected the intermediate focus of ATLT and TS and the far focus of FVHP and PO. IOL dimensions and diffractive profile were consistent with those described by the manufacturer. In conclusion, there are differences in the optical performance according to the pupil aperture of the five trifocal IOLs evaluated and this should be considered in clinical practice when selecting the most appropriate implant in each specific case. IOL tilt and decentration can affect significantly in most of the designs evaluated the performance of the IOL at intermediate vision range. It should be noted that measurements were made with an aberration-free cornea, being necessary future studies analysing the impact of different levels of corneal aberrations.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details
1 VSY Biotechnology GMBH, R & D Department, Leinfelden Echterdingen, Germany
2 University of Alicante, Group of Optics and Visual Perception, Department of Optics, Pharmacology and Anatomy, San Vicente del Raspeig, Alicante, Spain (GRID:grid.5268.9) (ISNI:0000 0001 2168 1800); Vithas Medimar International Hospital, Department of Ophthalmology, Alicante, Spain (GRID:grid.5268.9)