Full text

Turn on search term navigation

© 2024 mums.ac.ir All rights reserved. This work is published under https://creativecommons.org/licenses/by/3.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Objective(s):

To address the problem of using large volumes of long-lived radionuclides in test phantoms to check calibration accuracy of PET and SPECT systems we have developed a test object which (a) contains less radioactivity, (b) has a low total volume, and (c) is easier to store than currently used phantoms, while still making use of readily-available “standardised” test objects.

Methods:

We have designed a hollow acrylic cylindrical insert compatible with the NEMA/IEC PET Body Image Quality (IQ) phantom used in NU 2 performance testing of PET systems. The insert measures 90 mm internal diameter and 70 mm internal height and so is sufficiently large to not be subject to partial volume effects in PET or SPECT imaging. The volume of the insert is approximately 500 mL. It has been designed as a replacement for the standard long cylindrical “lung insert” in the IQ phantom without needing to remove the fillable hollow spheres of the phantom. The insert been tested with 18F, 68Ga and 124I PET/CT and 99mTc, 131I and 177Lu SPECT/CT on scanners that had previously been calibrated for these radionuclides.

Results:

The scanners were found to produce accurate image reconstructions in the insert with  5% of the true value without any confounding uncertainty from partial volume effects when compared to NEMA NU 2-2018 Phantom measurement.

Conclusions:

The “ARTnet Insert” is simple to use, inexpensive, compatible with current phantoms and is suitable for both PET and SPECT systems. It does not suffer from significant partial volume losses permitting its use even with the poor spatial resolution of high-energy imaging with 131I SPECT. Furthermore, it uses less radioactivity in a smaller volume than would be required to fill the entire phantom as is usually done. Long-term storage is practical while allowing radioactive decay of the insert contents.

Details

Title
A practical method for assessing quantitative scanner accuracy with long-lived radionuclides: The ARTnet insert
Author
Bailey, Dale L; Willowson, Kathy P; Muñoz-Ferrada, Carl
University/institution
U.S. National Institutes of Health/National Library of Medicine
Pages
27-34
Publication year
2024
Publication date
2024
ISSN
23225718
e-ISSN
23225726
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2920357483
Copyright
© 2024 mums.ac.ir All rights reserved. This work is published under https://creativecommons.org/licenses/by/3.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.