© 2024 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

(1) Background: In 90Y-TARE treatments, lung-absorbed doses should be calculated according to the manufacturer’s instructions, using the MIRD-scheme. This scheme is derived from the assumption that 90Y-microspheres deliver the dose in a water-equivalent medium. Since the density of the lungs is quite different from that of the liver, the absorbed dose to the lungs could vary considerably, especially at the liver/lungs interface. The aim of this work is to compare the dosimetric results obtained by two dedicated software packages implementing a water-equivalent dose calculation and a Monte Carlo (MC) simulation, respectively. (2) Methods: An anthropomorphic IEC phantom and a retrospective selection of 24 patients with a diagnosis of HCC were taken into account. In the phantom study, starting from a 90Y-PET/CT acquisition, the liver cavity was manually fixed with a uniform activity concentration on PET series, while the lung compartment was manually expanded on a CT series to simulate a realistic situation in which the liver and lungs are adjacent. These steps were performed by using MIM 90Y SurePlan. Then, a first simulation was carried out with only the liver cavity filled, while a second one was carried out, in which the lung compartment was also manually fixed with a uniform activity concentration corresponding to 10% lung shunt fraction. MIM 90Y SurePlan was used to obtain Voxel S-Value (VSV) approach dose values; instead, Torch was used to obtain MC approach dose values for both the phantom and the patients. (3) Results: In the phantom study, the percentage mean dose differences (∆D%) between VSV and MC in the first and second simulation, respectively were found to be 1.2 and 0.5% (absolute dose variation, ∆D, of 0.7 and 0.3 Gy) for the liver, −56 and 70% (∆D of −0.3 and −16.2 Gy) for the lungs, and −48 and −60% (∆D of −4.3 and −16.5 Gy) for the Liver/Lungs Edge region. The patient study reports similar results with ∆D% between VSV and MC of 7.0%, 4.1% and 6.7% for the whole liver, healthy liver, and tumor, respectively, while the result was −61.2% for the left lung and −61.1% for both the right lung and lungs. (4) Conclusion: Both VSV and MC allowed accurate radiation dose estimation with small differences (<7%) in regions of uniform water-equivalent density (i.e., within the liver). Larger differences between the two methods (>50%) were observed for air-equivalent regions in the phantom simulation and the patient study.

Details

Title
The Role of Lung Density in the Voxel-Based Dosimetry of 90Y-TARE Evaluated with the Voxel S-Value (VSV) Method and Fast Monte Carlo Simulation
Author
Capotosti, Amedeo 1   VIAFID ORCID Logo  ; Moretti, Roberto 2   VIAFID ORCID Logo  ; Vaccaro, Maria 3   VIAFID ORCID Logo  ; De Almeida Ribeiro, Cintia 3 ; Placidi, Lorenzo 2   VIAFID ORCID Logo  ; Nardini, Matteo 2   VIAFID ORCID Logo  ; Meffe, Guenda 2 ; Cusumano, Davide 2   VIAFID ORCID Logo  ; Zagaria, Luca 2 ; De Risi, Marina 2 ; Perotti, Germano 2   VIAFID ORCID Logo  ; Leccisotti, Lucia 2   VIAFID ORCID Logo  ; De Spirito, Marco 4   VIAFID ORCID Logo  ; Iezzi, Roberto 2   VIAFID ORCID Logo  ; Indovina, Luca 2   VIAFID ORCID Logo 

 Dipartimento di Diagnostica per Immagini, Radioterapia Oncologica ed Ematologia, Fondazione Policlinico Universitario A. Gemelli IRCCS, 00168 Rome, Italy; [email protected] (A.C.); [email protected] (R.M.); [email protected] (L.P.); [email protected] (M.N.); [email protected] (G.M.); [email protected] (D.C.); [email protected] (L.Z.); [email protected] (M.D.R.); [email protected] (G.P.); [email protected] (L.L.); [email protected] (M.D.S.); [email protected] (R.I.); [email protected] (L.I.); Dipartimento di Ingegneria dell’Informazione, Università di Padova, 35122 Padova, Italy 
 Dipartimento di Diagnostica per Immagini, Radioterapia Oncologica ed Ematologia, Fondazione Policlinico Universitario A. Gemelli IRCCS, 00168 Rome, Italy; [email protected] (A.C.); [email protected] (R.M.); [email protected] (L.P.); [email protected] (M.N.); [email protected] (G.M.); [email protected] (D.C.); [email protected] (L.Z.); [email protected] (M.D.R.); [email protected] (G.P.); [email protected] (L.L.); [email protected] (M.D.S.); [email protected] (R.I.); [email protected] (L.I.) 
 Dipartimento di Neuroscienze, Università Cattolica del Sacro Cuore, 00168 Rome, Italy; [email protected] 
 Dipartimento di Diagnostica per Immagini, Radioterapia Oncologica ed Ematologia, Fondazione Policlinico Universitario A. Gemelli IRCCS, 00168 Rome, Italy; [email protected] (A.C.); [email protected] (R.M.); [email protected] (L.P.); [email protected] (M.N.); [email protected] (G.M.); [email protected] (D.C.); [email protected] (L.Z.); [email protected] (M.D.R.); [email protected] (G.P.); [email protected] (L.L.); [email protected] (M.D.S.); [email protected] (R.I.); [email protected] (L.I.); Dipartimento di Neuroscienze, Università Cattolica del Sacro Cuore, 00168 Rome, Italy; [email protected] 
First page
1019
Publication year
2024
Publication date
2024
Publisher
MDPI AG
e-ISSN
20763417
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2923929418
Copyright
© 2024 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.