Abstract
Introduction
Photoscreeners have been shown to provide excellent measurements of the refractive error. However, whether they could be used for assessing cycloplegic refraction has not been examied. This study aimed to evaluate the agreement between cycloplegic and non-cycloplegic measurements obtained using a photoscreener and stationary autorefractor, respectively.
Methods
This study included all patients undergoing routine ophthalmic examination at the Hygeia Clinic (Poland) from June to July 2022. Each patient underwent non-cycloplegic and cycloplegic refraction assessments using the 2WIN photoscreener (Adaptica SRL, Padova, Italy) and an ARK-1 stationary autorefractor ARK-1 (Nidek Co Ltd., Tokyo, Japan), respectively. Each pair of assessments was conducted in random order, and all values were determined at a vertical distance of 12 mm. The agreement between cycloplegic and non-cycloplegic measurements was assessed using paired t-tests, Bland-Altman and ABCD ellipsoids.
Results
This analysis included 82 patients, of which 52 were female. Their mean age was 34.39 ± 13.13 years. The non-cycloplegic spherical equivalent (SE) did not differ significantly between the 2WIN (− 1.22 ± 2.45) and ARK-1 (− 1.19 ± 2.96) devices (p = 0.580). However, the cycloplegic SE values demonstrated more negative values with the 2WIN device (− 1.13 ± 2.19) than with the ARK-1 device (− 0.75 ± 3.03; p = 0.007). The non-cycloplegic and cycloplegic measurements were strongly correlated between the devices (r = 0.9473 and 0.9411, respectively). However, the correlation between their cycloplegic shifts in SE was low (r = 0.2645). Ellipsoid refraction aligned better non-cycloplegic (ARK-1 = 1.00; 2WIN = 1.74) than with cycloplegic refraction (ARK-1 = 1.43; 2WIN = 1.90).
Conclusion
While the cycloplegic measurements obtained with the 2WIN photoscreener were strongly correlated with those obtained with the ARK-1 stationary autorefractor for most of the analyzed parameters, they should not be considered interchangeable.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer