Abstract

Purpose

To design a patient specific quality assurance (PSQA) process for the CyberKnife Synchrony system and quantify its dosimetric accuracy using a motion platform driven by patient tumor traces with rotation.

Methods

The CyberKnife Synchrony system was evaluated using a motion platform (MODUSQA) and a SRS MapCHECK phantom. The platform was programed to move in the superior-inferior (SI) direction based on tumor traces. The detector array housed by the StereoPhan was placed on the platform. Extra rotational angles in pitch (head down, 4.0° ± 0.15° or 1.2° ± 0.1°) were added to the moving phantom to examine robot capability of angle correction during delivery. A total of 15 Synchrony patients were performed SBRT PSQA on the moving phantom. All the results were benchmarked by the PSQA results based on static phantom.

Results

For smaller pitch angles, the mean gamma passing rates were 99.75% ± 0.87%, 98.63% ± 2.05%, and 93.11% ± 5.52%, for 3%/1 mm, 2%/1 mm, and 1%/1 mm, respectively. Large discrepancy in the passing rates was observed for different pitch angles due to limited angle correction by the robot. For larger pitch angles, the corresponding mean passing rates were dropped to 93.00% ± 10.91%, 88.05% ± 14.93%, and 80.38% ± 17.40%. When comparing with the static phantom, no significant statistic difference was observed for smaller pitch angles (p = 0.1 for 3%/1 mm), whereas a larger statistic difference was observed for larger pitch angles (p < 0.02 for all criteria). All the gamma passing rates were improved, if applying shift and rotation correction.

Conclusions

The significance of this work is that it is the first study to benchmark PSQA for the CyberKnife Synchrony system using realistically moving phantoms with rotation. With reasonable delivery time, we found it may be feasible to perform PSQA for Synchrony patients with a realistic breathing pattern.

Details

Title
Feasibility of patient-specific quality assurance (PSQA) for real-time robotic stereotactic body radiotherapy (SBRT) based on tumor motion traces
Author
Xu, Qianyi 1 ; Fan, Jiajin 2 ; Vinogradskiy, Yevgeniy 3 ; Chawla, Ashish K 2 ; Kubicek, Gregory 4 ; Yang, Haihua 5   VIAFID ORCID Logo  ; Huynh, Kiet 2 ; LaCouture, Tamara 3 ; Grimm, Jimm 6 ; Nie, Wei 2 

 Department of Advanced Radiation Oncology and Proton Therapy, Inova Schar Cancer Institute, Fairfax, Virginia, USA; Department of Radiation Oncology, Thomas Jefferson University, Philadelphia, Pennsylvania, USA 
 Department of Advanced Radiation Oncology and Proton Therapy, Inova Schar Cancer Institute, Fairfax, Virginia, USA 
 Department of Radiation Oncology, Thomas Jefferson University, Philadelphia, Pennsylvania, USA 
 Department of Radiation Oncology, University of Miami, Miami, Florida, USA 
 Department of Radiation Oncology, Taizhou Hospital, Taizhou, Zhejiang, China 
 Department of Radiation Oncology, Wellstar Health System, Marietta, Georgia, USA 
Section
RADIATION ONCOLOGY PHYSICS
Publication year
2024
Publication date
Aug 2024
Publisher
John Wiley & Sons, Inc.
e-ISSN
15269914
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
3089531217
Copyright
© 2024. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.