Abstract

Purpose

The purpose of this study is to characterize the dosimetric properties of a commercial brass GRID collimator for high energy photon beams including 15 and 10 MV. Then, the difference in dosimetric parameters of GRID beams among different energies and linacs was evaluated.

Method

A water tank scanning system was used to acquire the dosimetric parameters, including the percentage depth dose (PDD), beam profiles, peak to valley dose ratios (PVDRs), and output factors (OFs). The profiles at various depths were measured at 100 cm source to surface distance (SSD), and field sizes of 10 × 10 cm2 and 20 × 20 cm2 on three linacs. The PVDRs and OFs were measured and compared with the treatment planning system (TPS) calculations.

Results

Compared with the open beam data, there were noticeable changes in PDDs of GRID fields across all the energies. The GRID fields demonstrated a maximal of 3 mm shift in dmax (Truebeam STX, 15MV, 10 × 10 cm2). The PVDR decreased as beam energy increases. The difference in PVDRs between Trilogy and Truebeam STx using 6MV and 15MV was 1.5% ± 4.0% and 2.1% ± 4.3%, respectively. However, two Truebeam linacs demonstrated less than 2% difference in PVDRs. The OF of the GRID field was dependent on the energy and field size. The measured PDDs, PVDRs, and OFs agreed with the TPS calculations within 3% difference. The TPS calculations agreed with the measurements when using 1 mm calculation resolution.

Conclusion

The dosimetric characteristics of high-energy GRID fields, especially PVDR, significantly differ from those of low-energy GRID fields. Two Truebeam machines are interchangeable for GRID therapy, while a pronounced difference was observed between Truebeam and Trilogy. A series of empirical equations and reference look-up tables for GRID therapy can be generated to facilitate clinical applications.

Details

Title
Dosimetric characterization for GRID collimator-based spatially fractionated radiation therapy: Dosimetric parameter acquisition and machine interchangeability investigation
Author
Xu, Zhengzheng 1 ; Balik, Salim 1   VIAFID ORCID Logo  ; Woods, Kaley 1   VIAFID ORCID Logo  ; Shen, Zhilei 1 ; Cheng, Chihyao 2 ; Cui, Jing 1 ; Gallogly, Haihong 1 ; Chang, Eric 1 ; Lukas, Lauren 1 ; Lim, Andrew 1 ; Natsuaki, Yutaka 1   VIAFID ORCID Logo  ; Ye, Jason 1 ; Ma, Lijun 1 ; Zhang, Hualin 1 

 Keck School of Medicine, University of Southern California, Los Angeles, California, USA; Department of Radiation Oncology, University of Southern California Norris Comprehensive Cancer Center, Los Angeles, California, USA 
 Department of Radiation Oncology, University of Southern California Norris Comprehensive Cancer Center, Los Angeles, California, USA 
Section
RADIATION ONCOLOGY PHYSICS
Publication year
2024
Publication date
Aug 2024
Publisher
John Wiley & Sons, Inc.
e-ISSN
15269914
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
3089531224
Copyright
© 2024. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.