1. Introduction
Radio frequency ablation (RFA) is a minimally invasive therapy that causes coagulative necrosis of tissues by increasing their temperature using a high-frequency current [1, 2]. RFA is used as a nonsurgical treatment for patients who cannot undergo surgical treatment. It has few complications, requires a short hospitalization period, and is also used to treat hypovascular and fibrous metastatic cancers with low treatment effects [3, 4]. When a high-frequency current is applied to tissues through electrodes inserted into the body, the ions in the tissues are arranged in the same direction as that of the alternating current and vibrate along the direction of the current. Thus, frictional or resistance heat is generated in the tissues, making the tissue around the electrodes serve as the heat source for RFA [1]. Various types of living tissues have different structures and properties, such as hydration, density, and impedance [5, 6]. Therefore, the size of the tissue ablated by RFA could vary based on tissue properties, output power of RF energy, application time, tissue temperature, and the electrodes used [1, 7–9].
Many studies have been conducted to increase the ablation size while changing the variables affecting RFA. For example, studies have been conducted to control the ablation range while changing the amount of electrical energy applied to living tissues. Generally, the higher the power, the higher the current flowing through the tissues, and the larger the ablation range [9, 10]. However, the increase in the ablation range was limited even with increased power owing to the shape of the electrode that delivered energy [8, 9]. Moreover, when the impedance of the current flow path increased, the ablation range did not increase even with increased power [7, 11]. An experiment to check the ablation range while changing the RF energy application time was also performed. Although there were differences in the tissue properties and power over a duration of approximately 8 min, the ablation range increased with ablation time; however, the changes were minimal after this duration [12–15]. Furthermore, when energy was applied for a long time at a high tissue temperature, the tissues were carbonized as moisture evaporated. Consequently, the impedance increased significantly, making it impossible to deliver energy [16, 17]. To overcome this problem, a cooling method that supplies saline through electrodes or temperature control using temperature sensors attached to electrodes has been reported [16–18]. Studies have also been conducted to expand the ablation range by enlarging the electrode size to an umbrella or balloon shape. The ablation range increased with the electrode and tissue contact areas. However, the shape of the electrodes limited available tissues [8, 19, 20]. Previous studies have focused on electrical energy, time, and electrode shape. However, in most clinical and research studies, the RF energy applied in a high-frequency generator was only a sine wave, and only one fixed frequency was used among the frequencies from 400 kHz to 500 kHz. A study on the ablation range based on changes in the waveform and frequency of electrical energy is necessary because many types and different characteristics of tissues, such as their hydration, density structure and composition. Recently, studies related to waveforms and frequencies have predominantly utilized theoretical approaches, with a focus on finite element analysis and simulation studies [21, 22]. Theoretically, a square wave has higher energy than a sine wave at the same amplitude and frequency. Additionally, at lower frequencies, the energy transmission distance increases. We hypothesized that using a square wave and lower frequencies could increase the ablation range compared to the conventional method. Therefore, this study developed an RFA generator that can control the waveform and frequency. Variations in the ablation range of the tissues were observed according to the waveform, frequency, electrode type, and distance. An in vitro experiment was performed using pork as tissues, with the tissue temperature not exceeding 70°C.
2. Method
2.1 Development of radio frequency generator
Fig 1 shows a block diagram of the RFA generator. The RFA generator comprised power, control, and amplification units.
[Figure omitted. See PDF.]
For the power unit, a flyback topology capable of electrical isolation while converting the AC inlet to DC was used. It was designed to output up to 200 watts with a 100 V DC voltage using UC3845 (STMicroelectronics, Geneve, Switzerland), a current-mode pulse-width modulation (PWM) controller. For the control unit, a microcontroller ATXMEGA128A1 (Microchip Technology, Inc., Chandler, AZ, USA) was used to control the overall system operations and RF signals while monitoring the impedance and temperature in real-time. RF signals that can convert sine or square waves to frequencies in the range of 10 kHz–500 kHz were generated using an AD9833 programmable waveform generator (Analog Devices, Inc., Wilmington, MA). The amplitudes of the RF signals generated using a digital potentiometer AD5291 (Analog Devices, Inc., Wilmington, MA) were controlled. The output parameters (waveform, frequency, and power) can be fine-tuned because all three control variables can be controlled digitally. The generated RF signals were delivered to the amplification unit before being outputted. A high-voltage power-operational amplifier PA107DP (Apex Microtechnology, Tucson, AZ, USA) was used for the amplification unit. The maximum output voltage, maximum output current, and output frequency range of this amplifier were 90 V, 1.5 A, and 2 MHz, respectively. Various waveforms and frequencies can be output even with a simple structure because a monolithic operational amplifier was used. An output of up to 70 watts can be achieved using this method. Maintenance of the set output power was regulated in real-time using a sense resistor and voltage divider comprising resistors.
An analog front-end ADS1292 (Texas Instruments, Inc., Dallas, TX, USA) was used for measuring the biopotential. The impedance changes and contact between the tissue and electrode were measured before and after applying RF energy. The temperature was monitored in real-time using an instrumentation amplifier AD620 (Analog Devices, Inc., Wilmington, MA) with four possible temperature settings (60, 70, 80, and 90°C). The selected temperature was maintained by setting the RF energy output on and off. The five control parameters of the RFA generator (output power, waveform, frequency, temperature, and time) can be set using a user interface comprising a character liquid crystal display (LCD), buttons, and a rotary encoder.
2.2 Experimental protocol
Fig 2 shows the experimental setup. Changes in the tissue ablation range due to the output waveform and frequency were investigated using the developed RFA generator. Furthermore, changes in the ablation range according to the control parameters were observed while changing the electrode type and distance between the electrodes.
[Figure omitted. See PDF.]
Refrigerated pork hind legs supplied by a butcher shop were used in the in vitro experiment. To maintain a constant distance (d) between the electrodes and insertion depth between experiments, 40 mm-sliced tissues were cut in half and divided into two A sandwich-shaped experimental sample was set up by positioning bipolar electrodes at the center of one piece of tissue and covering it with the other piece. The stent electrode was used in a deployed state. A plastic plate larger than the sample was placed on top of the sample to maintain a constant contact area between the electrode and tissue. It was then adjusted to the same impedance using a balanced weight. For the stent electrode, the electrode distance was set to approximately 90.3 Ω ± 3.6 Ω at 10 mm, 101.5 Ω ± 2.9 Ω at 15 mm, and 112.5 Ω ± 6.8 Ω at 20 mm. The resistance of the needle electrode was set to approximately 62.3 Ω ± 2.5 Ω at 10 mm, 72.5 Ω ± 6.2 Ω at 15 mm, and approximately 81.4 Ω ± 3.8 Ω at 20 mm. Furthermore, the temperature sensor was placed in the center between all of the two types of electrodes, and samples within an initial temperature of approximately 15°C ± 3°C were maintained for the experiments. All experiments were performed by applying RF energy for 90 s, with the average temperature not exceeding 70°C.
Changes in the ablation range caused by the output waveform were tested using sine and square waves. The output voltage and frequency were fixed at 45 V and 500 kHz, respectively. The electrode distance (d), based on the distance between the electrode centers, was varied to 10, 15, or 20 mm during the experiment. A bipolar stent electrode (S&G Biotech, Ltd., Yongin, South Korea) [10] was used for the prototype. An Octopus RF electrode (STARmed, Ltd., Goyang, South Korea) was used for the bipolar needle electrode. An experiment on the change in ablation range according to the output range was performed for both waveforms at 10, 100, and 500 kHz. The output voltage was fixed at ±45 V, and the largest distance of 20 mm was applied for both types of electrodes.
2.3 Data analysis
Twelve experiments were performed for the output waveform and frequency. A total of 72 experiments were performed using three samples for each experimental condition. The surface parallel to the electrodes of the samples is called the ablation area, the surface perpendicular to the electrodes is called the ablation depth, and the sum of the area and depth is defined as the range. The ablation area and depth of the samples were quantified using ImageJ software (National Institutes of Health, Bethesda, MD, USA), and the average was calculated. Statistical analysis was performed to detect the differences among the ablation range groups. ANOVA and t-test were performed using SPSS software (SPSS 29.0.2, IBM, Chicago, IL). Data were expressed as a mean ± standard deviation (M±SD). P<0.05 was considered statistically significant.
3. Results
3.1 Testing the output of the radio frequency generator
Fig 3 shows the actual view and operational results of the RFA generator. A load resistance of 60 Ω was applied under maximum output conditions. The voltage and current of the RF waveform were measured using a TBS200B oscilloscope (Tektronix, Inc., Beaverton, OR), high-voltage differential probe P5210A (Tektronix, Inc.), and current amplifier TCPA300 (Tektronix, Inc.). The effective value was used for the power and calculated using P = Vrms×Irms. Up to 70 watts was outputted with frequencies of 10 kHz to 500 kHz for both waveforms.
[Figure omitted. See PDF.]
(a) Actual appearance of the generator and operation test environment, (b) Maximum output of the sine and square waves for frequencies of 10 kHz and 500 kHz.
3.2 Tissue ablation according to the waveform
Fig 4 shows the changes in the waveform, ablation area, and depth under each experimental condition.
[Figure omitted. See PDF.]
The large photograph at the center shows the ablation area. The small photograph at bottom right shows the ablation depth. The stent electrode was positioned as shown in the first photograph.
In the experiment using the stent electrode, a sine wave was ablated to the electrode center at 10 mm and 15 mm. The temperature was maintained after the set maximum temperature of 70°C was reached. By contrast, the surroundings of the electrode were ablated at 20 mm, but not the electrode center. The maximum temperature at the center was approximately 44°C. The square wave was ablated to the electrode center at every distance. The ablation area increased with the distance, but the ablation depth decreased. The temperature at the center increased to 70°C and remained stable. In the experiment using the needle electrode, a sine wave was ablated to the center at 10 mm and 15 mm. At 20 mm, only the surroundings of the electrode were ablated, and the maximum temperature at the center was approximately 40°C. The square wave was ablated to the center under all conditions. The ablation area and depth increased with the distance. The center temperature also increased to 70°C.
Table 1 lists the mean and standard deviation ablation ranges (M±SD) calculated for the two waveforms according to the electrode type and distance. Samples with only the surroundings of the electrode ablated were excluded from the analysis. In experiments using the stent electrode, at an electrode distance of 10 mm, the square wave produced a significantly larger ablation area than the sine wave, with the sine wave producing an area of 91.3±8.4 mm2 and the square wave producing an area of 165.5±10.2 mm2 (p<0.001). At 15 mm, the sine wave produced an area of 222.5±14.1 mm2, while the square wave produced an area of 353.8±19.2 mm2 (p<0.001). In experiments using the needle electrode, at an electrode distance of 10 mm, the square wave produced an ablation area of 396.7±10.4 mm2, while the sine wave produced an area of 372.7±12.4 mm2. At 15 mm, the square wave produced an area of 517.7±20.4 mm2, while the sine wave produced an area of 485.0±17.9 mm2. In two conditions, the square wave produced a slightly larger area, but it were not statistically significant (p>0.05). The ablation depth showed significant differences under all four conditions (p<0.05). At an electrode distance of 20 mm, which was excluded from the analysis, the sine wave did not form a continuous ablation area, whereas the square wave formed a sufficient ablation area. The needle electrode exhibited a larger ablation area and depth than the stent electrode under all experimental conditions.
[Figure omitted. See PDF.]
Fig 5 shows the electrode type and temperature profile by distance based on the waveform changes. The temperature rise time varied according to the RF waveform, electrode type, and electrode distance. The square wave reached the set temperature faster than the sine wave. The stent electrode exhibited a faster temperature increase than the needle electrode. In addition, the shorter the electrode distance, the higher the temperature.
[Figure omitted. See PDF.]
(a) Temperature rise graph of the experiment using the stent electrode, (b) Temperature rise graph of the experiment using the needle electrode.
3.3 Tissue ablation according to the frequency
Fig 6 shows the mean and standard deviation ablation area and depth (M±SD) with varying frequencies under each experimental condition.
[Figure omitted. See PDF.]
The large photograph at the center shows the ablation area. The small photograph at bottom right shows the ablation depth. The stent electrode was positioned as shown in the first photograph.
In the experiment using the stent electrode, the sine and square waves exhibited a larger ablation area and depth and a faster temperature rise as the frequency decreased from 500 kHz to 100 Hz and 10 Hz. The center temperature of the square wave increased to 70°C, and every sample was ablated. A sine wave was ablated only around the electrode at a 500 kHz frequency. In the experiment using the needle electrode, a sine wave was ablated only in the surroundings of the electrode at all three frequencies. However, a square wave was ablated at the center of the electrode. Both waveforms showed an increased ablation area and depth as the frequency decreased.
Table 2 shows the ablation ranges calculated using the three frequencies and waveforms and according to electrode type. Samples that were not ablated to the tissue center were excluded from the analysis. In experiments using the stent electrode, the sine wave ablated a larger area at 100 kHz and 10 kHz compared to 500 kHz (p<0.01), but there was no statistically significant difference between 100 kHz and 10 kHz (p = 0.52). For the square wave, a larger ablation area was formed as the frequency decreased (p<0.01). In experiments using the needle electrode, the sine wave only ablated the area around the electrode and was thus excluded from the analysis. However, it was observed that the ablated thickness increased as the frequency decreased. For the square wave, the ablation area increased as the frequency decreased (p<0.05). In the frequency change experiment, the needle electrode exhibited a larger ablation area and depth than the stent electrode under all experimental conditions.
[Figure omitted. See PDF.]
Fig 7 shows the temperature profiles by electrode type and waveform based on the frequency changes. Similar to the waveform change experiment, as the frequency decreased, the temperature increased faster as the frequency decreased with the square wave than with the sine wave and with the stent electrode than with the needle electrode.
[Figure omitted. See PDF.]
(a) Temperature rise graph of the experiment using the stent electrode, (b) Temperature rise graph of the experiment using the needle electrode.
4. Discussion
This study reports the effects of the waveform and frequency of RF energy on the tissue ablation range. In this study, we developed an RFA generator that can provide sine and square waves and allow frequency control between 10 Hz and 500 kHz. The experiments with varying waveform, frequency, electrode type, and distance between electrodes revealed that the ablation area and depth were the largest for the square wave and low frequency. The needle electrode formed a larger ablation range than the stent electrode because it has a wider contact area with the tissue.
The developed RFA generator has the following advantages: first, the RF energies of the various waveforms were generated. In addition to the sine and square waves used in this study, various other waveforms, amplitudes, and frequency modulations are possible. Thus, the findings of this study can be applied to research on increasing ablation efficiency for various types of tissues; second, the output impedance was small. The impedance of the tissue, which was the load of the RFA generator, varied according to type [5, 6]. When the generator output impedance was high, the current decreased at the same power, and the ablation range decreased [11]. Thus, low-power impedance can effectively deliver energy to various tissues; third, RF energy can be output at various frequencies. Most RFA generators use transformers to facilitate output control and galvanic isolation [23–25]. The tissue impedance also changes with the frequency [5]. When a transformer is used, the output frequency is fixed, and the ablation result differs, even when the same energy is applied. The variability in the ablation range can be reduced by matching the RF energy of the frequency with the impedance measurement.
In the experiment on waveform changes, the ablation range increased more in the square wave compared to that in the sine wave. The same peak voltage of ±45 V was applied to both waveforms; however, the flow current was higher for the square wave than for the sine wave. This was because when the effective current for the square wave was 1, the effective current for the sine wave was approximately 0.7 [26]. The mean effective currents of the square and sine waves used in the experiment were 0.92 A and 0.64 A, respectively, corresponding to a difference of approximately 0.696 times. The tissue temperature reached 70°C faster in the square wave than in the sine wave owing to the high current. The impedance decreased when the tissue temperature increased [17, 27]. The high current flowed easily owing to the decreased impedance, resulting in the ablation of the center even with increased distance between the electrodes [7, 11]. In the experiment on frequency changes, the ablation range increased with low-frequency RF energy compared to that with high-frequency RF energy. In skin-depth studies of biological tissues, the energy transfer depth increased as the frequency decreased [28, 29]. In this study, the mean effective current increased from 0.86 A to 1.06 A and 1.12 A as the frequency decreased from 500 kHz to 100 kHz and 10 Hz. High current enables a rapid temperature rise rate and a large ablation range. These results suggest that variations in waveform and frequency enable the control of the rate of temperature increase, thereby allowing for the adjustment of the ablation range.
The ablation range differed according to the electrode type. Under the same conditions, the needle electrode showed a larger ablation range than the stent electrode. This result was attributed to the difference in the contact area between the two types of electrodes and the tissues. The larger the contact area, the smaller the impedance, and the faster the current flow. The mean impedances of the tissue measured using each electrode were 102 Ω and 72 Ω for the stent and needle electrodes, respectively. More current flowed into the needle electrode owing to the lower impedance.
This study has several limitations. It was conducted exclusively on the muscle tissues of the pork hind legs, chosen for the accessibility of sample acquisition and the clarity in identifying the ablation range. However, it is necessary to study the variations in ablation range based on waveform and frequency changes in tissues such as the liver, lungs, and bones where RFA is clinically applied. Moreover, these experiments were performed only in an in vitro environment. The tissue activities of the cells were suspended in this study; therefore, additional ex vivo and in vivo experiments are necessary. Research on the side effects of tissue caused by waveform and frequency is also essential. For example, the use of low frequencies in cardiac ablation is limited because it can induce arrhythmias [30, 31]. Such complication can significantly impact the patient’s safety. Therefore, it is crucial to thoroughly research and understand the potential side effects that may occur under various frequency conditions. This understanding will help provide more safer and better effective treatment methods for patients in the context of using RFA treatment.
5. Conclusion
In this study, we hypothesized that using a square wave, which has higher energy than a sine wave, and a lower frequency, which can transmit energy further, would increase the ablation range. To verify this, we developed a radiofrequency (RFA) generator capable of adjusting the waveform and frequency and observed changes in the ablation range according to the various electrical characteristics of RF energy. Traditional RF ablation using sine waves required higher voltage to achieve a wide ablation range. However, it was confirmed that by adjusting the frequency and waveform, a wide ablation range could be obtained with control of the temperature rise rate at the same voltage. This suggests that using various waveform and frequency changes can provide new methods for expanding and controlling the size and depth of the ablation range.
Acknowledgments
We thank Director Young Jae Lee, CEO SeChul Lee and Dr. Sang Soo Lee for their help with the organization of this project.
References
1. 1. Ni Y., Mulier S., Miao Y., Michel L., Marchal G. A review of the general aspects of radiofrequency ablation, Abdominal Imaging. 2005 Aug; 30:381–400. pmid:15776302
* View Article
* PubMed/NCBI
* Google Scholar
2. 2. Ahrar K., Wallace M.J., Matin S.F. Percutaneous radiofrequency ablation: minimally invasive therapy for renal tumors, Expert Review of Anticancer Therapy. 2006; 6: 1735–1744. pmid:17181487
* View Article
* PubMed/NCBI
* Google Scholar
3. 3. Hong K., Georgiades C. Radiofrequency ablation: mechanism of action and devices, Journal of Vascular and Interventional Radiology. 2010 Aug; 21(8): Supplement S179–S186. pmid:20656227
* View Article
* PubMed/NCBI
* Google Scholar
4. 4. Solbiati L., Ierace T., Goldberg S.N., Sironi S., Livraghi T., Fiocca R., et al. Percutaneous US-guided radio-frequency tissue ablation of liver metastases: treatment and follow-up in 16 patients, Radiology. 1997; 202:195–203. https://doi.org/10.1148/radiology.202.1.8988211, http://www.ncbi.nlm.nih.gov/pubmed/8988211 pmid:8988211
* View Article
* PubMed/NCBI
* Google Scholar
5. 5. Hartsgrove G., Kraszewski A., SUROWIEC , et al. Simulated biological materials for electromagnetic radiation absorption studies. Bioelectromagnetics: Journal of the Bioelectromagnetics Society, The Society for Physical Regulation in Biology and Medicine, the European Bioelectromagnetics Association. 1987, 8(1):29–36. pmid:3579997
* View Article
* PubMed/NCBI
* Google Scholar
6. 6. Sharp J., Bouazza-Marouf K., Noronha D., Gaur A. Tissue type determination by impedance measurement: A bipolar and monopolar comparison, Saudi Journal of Anaesthesia. 2017; 11:15–20. pmid:28217047
* View Article
* PubMed/NCBI
* Google Scholar
7. 7. Bourier F., Ramirez F.D., Martin C.A., Vlachos K., Frontera A., Takigawa M., et al. Impedance, power, and current in radiofrequency ablation: insights from technical, ex vivo, and clinical studies, Journal of Cardiovascular Electrophysiology. 2020; 31:2836–2845. pmid:32757434
* View Article
* PubMed/NCBI
* Google Scholar
8. 8. Zhang B., Moser M.A., Zhang E.M., Luo Y., Zhang W. Numerical analysis of the relationship between the area of target tissue necrosis and the size of target tissue in liver tumours with pulsed radiofrequency ablation, International Journal of Hyperthermia. 2015; 31:715–725. pmid:26360111
* View Article
* PubMed/NCBI
* Google Scholar
9. 9. Itoi T., Isayama H., Sofuni A., Itokawa F., Tamura M., Watanabe Y., et al. Evaluation of effects of a novel endoscopically applied radiofrequency ablation biliary catheter using an ex‐vivo pig liver, Journal of Hepato–Biliary–Pancreatic Sciences. 2012; 19: 543–547. pmid:22038500
* View Article
* PubMed/NCBI
* Google Scholar
10. 10. Won D.-S., Park Y., An J., Ryu D.S., Kang J.M., Kim J.W., et al. Stent-based electrode for radiofrequency ablation in the rat esophagus: a preliminary study, Scientific Reports. 2022; 12:19135. pmid:36352051
* View Article
* PubMed/NCBI
* Google Scholar
11. 11. Bhaskaran A., Barry M.A., Pouliopoulos J., Nalliah C., Qian P., Chik W., et al. Circuit impedance could be a crucial factor influencing radiofrequency ablation efficacy and safety: a myocardial phantom study of the problem and its correction, Journal of Cardiovascular Electrophysiology. 2016; 27:351–357. pmid:26648095
* View Article
* PubMed/NCBI
* Google Scholar
12. 12. Patterson E.J., Scudamore C.H., Owen D.A., Nagy A.G., Buczkowski A.K., Radiofrequency ablation of porcine liver in vivo: effects of blood flow and treatment time on lesion size, ANNALS of SURGERY. 1998; 227:559–565. pmid:9563546
* View Article
* PubMed/NCBI
* Google Scholar
13. 13. Zorbas G., Samaras T., Simulation of radiofrequency ablation in real human anatomy, International Journal of Hyperthermia. 2014; 30:570–578. pmid:25366922
* View Article
* PubMed/NCBI
* Google Scholar
14. 14. Goldberg S.N., Gazelle G.S., Dawson S.L., Rittman W.J., Mueller P.R., Rosenthal D.I., Tissue ablation with radiofrequency: effect of probe size, gauge, duration, and temperature on lesion volume, Academic Radiology. 1995; 2:399–404. pmid:9419582
* View Article
* PubMed/NCBI
* Google Scholar
15. 15. Lorentzen T., Christensen N.E., Nolsłe C.P., Torp-Pedersen S.T., Radiofrequency tissue ablation with a cooled needle in vitro: ultrasonography, dose response, and lesion temperature, Academic Radiology. 1997; 4:292–297. pmid:9110027
* View Article
* PubMed/NCBI
* Google Scholar
16. 16. J. Alba, A. Gonzalez-Suarez, M. Trujillo, E. Berjano, Theoretical and experimental study on RF tumor ablation with internally cooled electrodes: when does the roll-off occur?, 2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society, IEEE Publications. 2011; pp. 314–317. https://doi.org/10.1109/IEMBS.2011.6090082
17. 17. da Fonseca R.D., Santos P.R., Monteiro M.S., Fernandes L.A., Campos A.H., Borges D.L., et al. Parametric evaluation of impedance curve in radiofrequency ablation: A quantitative description of the asymmetry and dynamic variation of impedance in bovine ex vivo model, PLOS ONE. 2021; 16:e0245145. pmid:33449951
* View Article
* PubMed/NCBI
* Google Scholar
18. 18. Petersen H.H., Chen X., Pietersen A., Svendsen J.H., Haunsø S., Tissue temperatures and lesion size during irrigated tip catheter radiofrequency ablation: an in vitro comparison of temperature‐controlled irrigated tip ablation, power‐controlled irrigated tip ablation, and standard temperature‐controlled ablation, Pacing and Clinical Electrophysiology. 2000; 23:8–17. pmid:10666748
* View Article
* PubMed/NCBI
* Google Scholar
19. 19. Andrade J.G., Dubuc M., Guerra P.G., Macle L., Mondésert B., Rivard L., et al. The biophysics and biomechanics of cryoballoon ablation, Pacing and Clinical Electrophysiology. 2012; 35:1162–1168. pmid:22671922
* View Article
* PubMed/NCBI
* Google Scholar
20. 20. Meijerink M.R., van den Tol P., van Tilborg A.A., van Waesberghe J.H., Meijer S., van Kuijk C., Radiofrequency ablation of large size liver tumours using novel plan-parallel expandable bipolar electrodes: initial clinical experience, European Journal of Radiology. 2011; 77:167–171. pmid:19616911
* View Article
* PubMed/NCBI
* Google Scholar
21. 21. Haemmerich D, Schutt DJ. RF ablation at low frequencies for targeted tumor heating: in vitro and computational modeling results. IEEE Transactions on Biomedical Engineering. 2011 Feb; 58(2):404–10. pmid:20934940
* View Article
* PubMed/NCBI
* Google Scholar
22. 22. Iasiello M, Andreozzi A, Bianco N, Vafai K. Effects of Pulsed Radiofrequency Source on Cardiac Ablation. Bioengineering. 2023 Feb 8; 10(2):227. pmid:36829721
* View Article
* PubMed/NCBI
* Google Scholar
23. 23. Nociarova J., Novak M., Polak J., Hrudka J., Porubsky S., Koc M., et al. Development of radiofrequency ablation generator and balloon-based catheter for microendoluminal thin-layer ablation therapy using the rat duodenum as a model of low-impedance tissue, Journal of Healthcare Engineering. 2021 Sep; 9986874. https://doi.org/10.1155/2021/9986874
* View Article
* Google Scholar
24. 24. J. Moore, K. Nilsson, B. Wood, S. Xu, Z.T. Ho Tse, Liver radiofrequency ablation using wirelessly powered catheter and generator, 2018 IEEE PELS Workshop on Emerging Technologies: Wireless Power Transfer (Wow), IEEE Publications. 2018; pp. 1–5. https://doi.org/10.1109/WoW.2018.8450904
25. 25. Simon J., Probst U., Klar P.J., Development of a radio-frequency generator for RF ion thrusters, AEROSPACE TECHNOLOGY JAPAN. 2016; 14:Pb_33–Pb_39. https://doi.org/10.2322/tastj.14.Pb_33
* View Article
* Google Scholar
26. 26. Bird J. Electrical Circuit Theory and Technology, Routledge, 2017; pp. 224–228.
27. 27. Dadd J.S., Ryan T.P., Platt R., Tissue impedance as a function of temperature and time, Biomedical Sciences Instrumentation. 1996; 32:205–214. http://europepmc.org/abstract/MED/8672670 pmid:8672670
* View Article
* PubMed/NCBI
* Google Scholar
28. 28. Fan X., Zhou Q., Liu Z., Xie F., New derivation method and simulation of skin effect in biological tissue, Bio-Medical Materials and Engineering. 2015; 26 (s1):S429–S437. pmid:26406033
* View Article
* PubMed/NCBI
* Google Scholar
29. 29. V. Sydorets, A. Dubko, O. Bondarenko, R. Kosenko, Influence of skin effect on current flow through electrodes of electro-surgical instruments and biological tissue, 2016 15th Biennial Baltic Electronics Conference (BEC), IEEE Publications. 2016; pp. 211–214. https://doi.org/10.1109/BEC.2016.7743766
30. 30. IKEDA , TAKANORI , KENJI NINOMIYA, YOSHIHISA ENJOJI, RYOJI ABE, et al. Optimal frequency of radiofrequency catheter ablation in vitro and in vivo for treating ventricular tachycardias: A study using various frequencies. Japanese circulation journal. 1993; 57(9):921–927. pmid:8371485
* View Article
* PubMed/NCBI
* Google Scholar
31. 31. Orczykowski Michał, Derejko P., Urbanek P., Bodalski R., et al. Ventricular fibrillation induced by radiofrequency energy delivery for premature ventricular contractions arising from the right ventricular outflow tract: is implantable cardioverter-defibrillator indicated?, Polish Archives Internal Medicine. 2018 Jan; 128(3):166–170. pmid:29600967
* View Article
* PubMed/NCBI
* Google Scholar
Citation: An J, Won D-S, Park Y, Park J-H, Park K-H, Lee J-H, et al. (2024) Effects of changes in the waveform and frequency of radio frequency energy on tissue ablation range. PLoS ONE 19(9): e0308691. https://doi.org/10.1371/journal.pone.0308691
About the Authors:
Jinsu An
Roles: Conceptualization, Formal analysis, Software, Validation, Writing – original draft, Writing – review & editing
Affiliation: Department of Biomedical Engineering, School of ICT Convergence Engineering, College of Science & Technology, Konkuk University, Chungju-si, Chungcheongbuk-do, Republic of Korea
Dong-Sung Won
Roles: Conceptualization, Data curation, Investigation, Methodology
Affiliation: Biomedical Engineering Research Center, Asan Institute for Life Sciences, Asan Medical Center, Songpa-gu, Seoul, Republic of Korea
Yubeen Park
Roles: Conceptualization, Data curation, Investigation, Methodology
Affiliations: Biomedical Engineering Research Center, Asan Institute for Life Sciences, Asan Medical Center, Songpa-gu, Seoul, Republic of Korea, Department of Gastroenterology, Asan Medical Center, University of Ulsan College of Medicine, Songpa-gu, Seoul, Republic of Korea
Jung-Hoon Park
Roles: Conceptualization, Data curation, Investigation, Methodology
Affiliations: Biomedical Engineering Research Center, Asan Institute for Life Sciences, Asan Medical Center, Songpa-gu, Seoul, Republic of Korea, Department of Convergence Medicine, Asan Medical Center, University of Ulsan College of Medicine, Songpa-gu, Seoul, Republic of Korea
Ki-Hyeon Park
Roles: Conceptualization, Investigation
Affiliation: Department of Mechatronics Engineering, School of ICT Convergence Engineering, College of Science & Technology, Konkuk University, Chungju-si, Chungcheongbuk-do, Republic of Korea
Ji-Ho Lee
Roles: Conceptualization, Investigation
Affiliation: Department of Mechatronics Engineering, School of ICT Convergence Engineering, College of Science & Technology, Konkuk University, Chungju-si, Chungcheongbuk-do, Republic of Korea
Hyung-Sik Kim
Roles: Conceptualization, Data curation, Methodology, Project administration, Supervision, Writing – review & editing
E-mail: [email protected]
Affiliation: Department of Mechatronics Engineering, School of ICT Convergence Engineering, College of Science & Technology, Konkuk University, Chungju-si, Chungcheongbuk-do, Republic of Korea
ORICD: https://orcid.org/0000-0002-7566-7665
[/RAW_REF_TEXT]
[/RAW_REF_TEXT]
[/RAW_REF_TEXT]
[/RAW_REF_TEXT]
1. Ni Y., Mulier S., Miao Y., Michel L., Marchal G. A review of the general aspects of radiofrequency ablation, Abdominal Imaging. 2005 Aug; 30:381–400. pmid:15776302
2. Ahrar K., Wallace M.J., Matin S.F. Percutaneous radiofrequency ablation: minimally invasive therapy for renal tumors, Expert Review of Anticancer Therapy. 2006; 6: 1735–1744. pmid:17181487
3. Hong K., Georgiades C. Radiofrequency ablation: mechanism of action and devices, Journal of Vascular and Interventional Radiology. 2010 Aug; 21(8): Supplement S179–S186. pmid:20656227
4. Solbiati L., Ierace T., Goldberg S.N., Sironi S., Livraghi T., Fiocca R., et al. Percutaneous US-guided radio-frequency tissue ablation of liver metastases: treatment and follow-up in 16 patients, Radiology. 1997; 202:195–203. https://doi.org/10.1148/radiology.202.1.8988211, http://www.ncbi.nlm.nih.gov/pubmed/8988211 pmid:8988211
5. Hartsgrove G., Kraszewski A., SUROWIEC , et al. Simulated biological materials for electromagnetic radiation absorption studies. Bioelectromagnetics: Journal of the Bioelectromagnetics Society, The Society for Physical Regulation in Biology and Medicine, the European Bioelectromagnetics Association. 1987, 8(1):29–36. pmid:3579997
6. Sharp J., Bouazza-Marouf K., Noronha D., Gaur A. Tissue type determination by impedance measurement: A bipolar and monopolar comparison, Saudi Journal of Anaesthesia. 2017; 11:15–20. pmid:28217047
7. Bourier F., Ramirez F.D., Martin C.A., Vlachos K., Frontera A., Takigawa M., et al. Impedance, power, and current in radiofrequency ablation: insights from technical, ex vivo, and clinical studies, Journal of Cardiovascular Electrophysiology. 2020; 31:2836–2845. pmid:32757434
8. Zhang B., Moser M.A., Zhang E.M., Luo Y., Zhang W. Numerical analysis of the relationship between the area of target tissue necrosis and the size of target tissue in liver tumours with pulsed radiofrequency ablation, International Journal of Hyperthermia. 2015; 31:715–725. pmid:26360111
9. Itoi T., Isayama H., Sofuni A., Itokawa F., Tamura M., Watanabe Y., et al. Evaluation of effects of a novel endoscopically applied radiofrequency ablation biliary catheter using an ex‐vivo pig liver, Journal of Hepato–Biliary–Pancreatic Sciences. 2012; 19: 543–547. pmid:22038500
10. Won D.-S., Park Y., An J., Ryu D.S., Kang J.M., Kim J.W., et al. Stent-based electrode for radiofrequency ablation in the rat esophagus: a preliminary study, Scientific Reports. 2022; 12:19135. pmid:36352051
11. Bhaskaran A., Barry M.A., Pouliopoulos J., Nalliah C., Qian P., Chik W., et al. Circuit impedance could be a crucial factor influencing radiofrequency ablation efficacy and safety: a myocardial phantom study of the problem and its correction, Journal of Cardiovascular Electrophysiology. 2016; 27:351–357. pmid:26648095
12. Patterson E.J., Scudamore C.H., Owen D.A., Nagy A.G., Buczkowski A.K., Radiofrequency ablation of porcine liver in vivo: effects of blood flow and treatment time on lesion size, ANNALS of SURGERY. 1998; 227:559–565. pmid:9563546
13. Zorbas G., Samaras T., Simulation of radiofrequency ablation in real human anatomy, International Journal of Hyperthermia. 2014; 30:570–578. pmid:25366922
14. Goldberg S.N., Gazelle G.S., Dawson S.L., Rittman W.J., Mueller P.R., Rosenthal D.I., Tissue ablation with radiofrequency: effect of probe size, gauge, duration, and temperature on lesion volume, Academic Radiology. 1995; 2:399–404. pmid:9419582
15. Lorentzen T., Christensen N.E., Nolsłe C.P., Torp-Pedersen S.T., Radiofrequency tissue ablation with a cooled needle in vitro: ultrasonography, dose response, and lesion temperature, Academic Radiology. 1997; 4:292–297. pmid:9110027
16. J. Alba, A. Gonzalez-Suarez, M. Trujillo, E. Berjano, Theoretical and experimental study on RF tumor ablation with internally cooled electrodes: when does the roll-off occur?, 2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society, IEEE Publications. 2011; pp. 314–317. https://doi.org/10.1109/IEMBS.2011.6090082
17. da Fonseca R.D., Santos P.R., Monteiro M.S., Fernandes L.A., Campos A.H., Borges D.L., et al. Parametric evaluation of impedance curve in radiofrequency ablation: A quantitative description of the asymmetry and dynamic variation of impedance in bovine ex vivo model, PLOS ONE. 2021; 16:e0245145. pmid:33449951
18. Petersen H.H., Chen X., Pietersen A., Svendsen J.H., Haunsø S., Tissue temperatures and lesion size during irrigated tip catheter radiofrequency ablation: an in vitro comparison of temperature‐controlled irrigated tip ablation, power‐controlled irrigated tip ablation, and standard temperature‐controlled ablation, Pacing and Clinical Electrophysiology. 2000; 23:8–17. pmid:10666748
19. Andrade J.G., Dubuc M., Guerra P.G., Macle L., Mondésert B., Rivard L., et al. The biophysics and biomechanics of cryoballoon ablation, Pacing and Clinical Electrophysiology. 2012; 35:1162–1168. pmid:22671922
20. Meijerink M.R., van den Tol P., van Tilborg A.A., van Waesberghe J.H., Meijer S., van Kuijk C., Radiofrequency ablation of large size liver tumours using novel plan-parallel expandable bipolar electrodes: initial clinical experience, European Journal of Radiology. 2011; 77:167–171. pmid:19616911
21. Haemmerich D, Schutt DJ. RF ablation at low frequencies for targeted tumor heating: in vitro and computational modeling results. IEEE Transactions on Biomedical Engineering. 2011 Feb; 58(2):404–10. pmid:20934940
22. Iasiello M, Andreozzi A, Bianco N, Vafai K. Effects of Pulsed Radiofrequency Source on Cardiac Ablation. Bioengineering. 2023 Feb 8; 10(2):227. pmid:36829721
23. Nociarova J., Novak M., Polak J., Hrudka J., Porubsky S., Koc M., et al. Development of radiofrequency ablation generator and balloon-based catheter for microendoluminal thin-layer ablation therapy using the rat duodenum as a model of low-impedance tissue, Journal of Healthcare Engineering. 2021 Sep; 9986874. https://doi.org/10.1155/2021/9986874
24. J. Moore, K. Nilsson, B. Wood, S. Xu, Z.T. Ho Tse, Liver radiofrequency ablation using wirelessly powered catheter and generator, 2018 IEEE PELS Workshop on Emerging Technologies: Wireless Power Transfer (Wow), IEEE Publications. 2018; pp. 1–5. https://doi.org/10.1109/WoW.2018.8450904
25. Simon J., Probst U., Klar P.J., Development of a radio-frequency generator for RF ion thrusters, AEROSPACE TECHNOLOGY JAPAN. 2016; 14:Pb_33–Pb_39. https://doi.org/10.2322/tastj.14.Pb_33
26. Bird J. Electrical Circuit Theory and Technology, Routledge, 2017; pp. 224–228.
27. Dadd J.S., Ryan T.P., Platt R., Tissue impedance as a function of temperature and time, Biomedical Sciences Instrumentation. 1996; 32:205–214. http://europepmc.org/abstract/MED/8672670 pmid:8672670
28. Fan X., Zhou Q., Liu Z., Xie F., New derivation method and simulation of skin effect in biological tissue, Bio-Medical Materials and Engineering. 2015; 26 (s1):S429–S437. pmid:26406033
29. V. Sydorets, A. Dubko, O. Bondarenko, R. Kosenko, Influence of skin effect on current flow through electrodes of electro-surgical instruments and biological tissue, 2016 15th Biennial Baltic Electronics Conference (BEC), IEEE Publications. 2016; pp. 211–214. https://doi.org/10.1109/BEC.2016.7743766
30. IKEDA , TAKANORI , KENJI NINOMIYA, YOSHIHISA ENJOJI, RYOJI ABE, et al. Optimal frequency of radiofrequency catheter ablation in vitro and in vivo for treating ventricular tachycardias: A study using various frequencies. Japanese circulation journal. 1993; 57(9):921–927. pmid:8371485
31. Orczykowski Michał, Derejko P., Urbanek P., Bodalski R., et al. Ventricular fibrillation induced by radiofrequency energy delivery for premature ventricular contractions arising from the right ventricular outflow tract: is implantable cardioverter-defibrillator indicated?, Polish Archives Internal Medicine. 2018 Jan; 128(3):166–170. pmid:29600967
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2024 An et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
This study reports the effects of changes in the waveform and frequency of radio frequency (RF) energy on the tissue ablation range. We developed a 70-watt RFA generator that provides sine and square waves and allows frequency control between 10 Hz and 500 kHz. The changes in the ablation range according to the waveform and frequency were observed using the developed generator. In the waveform variation test, the distance between the electrodes and the electrode type were changed for both waveforms with the frequency set to 500 kHz. In the frequency variation test, the waveform and electrode type were changed with the frequency set to 10, 100, and 500 kHz, while the distance between the electrodes was set to 20 mm. A fixed 45 voltage was applied using the bipolar method. RF energy was applied for 90 s in vitro. The temperature was regulated to not exceed 70°C. The ablation range was calculated using ImageJ software. The analysis results showed that the ablation range was larger with the square wave than with the sine wave and at 10 kHz than at 500 kHz. The developed generator can advance research on ablation area and depth in RF ablation.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer