INTRODUCTION
Total body irradiation (TBI) is indeed a critical component in the preparative regimen for bone marrow transplantation.1–3 Its primary goals are twofold: to eradicate malignant cells that may be resistant to chemotherapy and to immunosuppress the patient sufficiently to allow for the engraftment of donor stem cells. Clinically, TBI can be delivered in two typical regimens: single-fraction TBI (a total of 3–4 Gy delivered in one fraction) or hyper-fraction TBI (10–12 Gy delivered in 5 or 6 fractions, BID, 6 h apart, over 2 or 3 consecutive days). Regardless of the treatment regimen, it involves exposing the entire body to mega-voltage photon beams, and all organs-at-risk (OAR) receive a uniform dose. This uniform exposure could lead to significant side effects and complications in radiosensitive ones, particularly in the lungs, causing pulmonary toxicity (PT), which can severely impact respiratory function and, in severe cases, lead to mortality.
In our center, we prepared to deliver hyper-fraction TBI treatment (8–10 Gy in 4 or 5 fractions, twice daily, 6 h apart, on 2 or 3 consecutive days) using a Linear Accelerator (LINAC). A reduced total dose of 8–10 Gy for TBI is administrated in conjunction with chemotherapy, taking into account the patient's poor performance status, age, and cumulative chemotherapy exposures.4 By conducting this treatment, our major concern with this treatment is the incidence of PT following TBI. To mitigate it, it is crucial to reduce both the total dose and instantaneous dose rates to the lungs simultaneously.5
Lung shielding is important in TBI treatment. In 1996, Gore's study first revealed that higher radiation doses to the lungs could damage long-term pulmonary functions.6 Additionally, Sampath's study7 in 2004 reported that 50% lung-dose shielding, which means mean lung dose (MLD) ≤6 Gy here, could reduce the incidence of IP from 11% (no shielding) to 2.3% (lung shielding) in hyper-fractionated TBI (12 Gy in 6 fractions, BID). Recent studies8 also reported a higher risk of lung toxicity with MLD > 8 Gy after hyper-fractionated TBI. Based on the previous study we aim to constraint the MLD ≤7 Gy with multi-leaf collimators (MLC).
Meanwhile, the ACR-ASTRO guideline9 also suggests using relatively low dose rates to mitigate PT in patients undergoing hyper-fractionated TBI. Several studies over the last few decades have proved the dose-rate effect on PT10–14 following hyper-fractionated TBI treatment. Carruthers's study in 200410 found that patients who received hyper-fractionated TBI (12 Gy in 6 fractions, BID, over 3 days) using a dose rate of 15 cGy/min had a higher rate of PT than those using a dose rate of 7.5 cGy/min. In 2019, Gao's study11 reported that patients who received hyper-fractionated TBI (13.2 Gy in eight fractions, BID, over 4 days) using dose rates ≤15 cGy/min had a lower incidence of PT and improved overall survival as compared to those receiving dose rates ≥15 cGy/mi.
Even though, it remains challenging to determine the optimal dose rate to reduce PT. Gogna's study in 1992 concluded that dose rates ≤8.9 cGy/min would not influence PT for hyper-fractionated TBI,15 while more recent studies1,5,10,11 suggested that hyper-fractionated TBI using dose rates ≤15 cGy/min could decrease the occurrence of PT. Also, Oya's study16 reported no significant difference in PT between hyper-fractionated TBI (12 Gy in 6 fractions, BID, consecutive 3 days) using dose rates of 8 and 19 cGy/min. These findings suggest a safe threshold for dose rates to lungs during hyper-fractionated TBI treatment ranges from 8 to 19 cGy/min. Although lowering the instantaneous dose to the lungs is safer, it also extends the treatment time, increasing the treatment risks. Therefore, a balance between treatment safety and delivery efficiency is necessary. Considering that TBI treatments have been conducted in a large open field at extended source-to-distance (SSD) safely over the past few decades, as reported in AAPM Report No. 17, with low instantaneous dose rates ≤approximately 10 cGy/min, we propose to use instantaneous dose rates ≤10 cGy/min as the threshold for a relatively safe hyper-fractionated TBI in our study.
In this study, we aim to develop a hyper-fractionated TBI technique that delivers a total dose of 8–10 Gy in 4–5fractions, twice daily, 6 hours apart, with a low MLD ≤ 7 Gy and a low instantaneous dose rate ≤10 cGy/min to reduce the incidence of PT and improve treatment efficiency simultaneously.
MATERIALS AND METHODS
Preparation of LINAC's dose rate
Transmission block
Regarding setups, the low instantaneous dose rate (≤10 cGy/min) is typically achieved by using large open fields at an extended SSD of around 400 cm, employing low machine meter-set dose rates, or sometimes a combination of both. Patients were generally irradiated in anterior-to-posterior/posterior-to-anterior (AP/PA) lying positions17–19 or AP/PA standing positions,20 with lungs shielding using a customized block.21 However, in the absence of dedicated TBI-specific meter-set dose rates in our LINAC (Trilogy, Varian Medical Systems, Palo Alto, CA) and limited room size in our center, we employed a transmission block and positioned patients at extended SSDs of 180–200 cm to achieve low instantaneous dose rates ≤10 cGy/min.
As calculated using lead attenuation coefficients, we manufactured a 2-cm thick, 28 cm × 28 cm wide lead block to achieve lower instantaneous dose rates ≤10 cGy/min in the LINAC. To ensure the safety and proper repositioning of the transmission block, we implemented several measures, including machine interlocks, pre-treatment checklists, and in-vivo dose monitoring. For machine interlocks, the transmission block was screwed on the TBI-dedicated block tray with TBI-specific customer coding, which is added to each field during TBI planning in the TPS. The block's secure attachment to the tray addresses repositioning and falling concerns. Before each fraction, the TBI-dedicated block tray is mounted in the slot and identified by the LINAC, as illustrated in Figure 1. Before beam on, therapists must confirm this TBI-specific customer coding on the treatment console. The LINAC will interlock the beam immediately if the block tray is not correctly positioned or if the wrong block is used. Additionally, an in-room timeout checklist guides block placement and patient setup before beam-on, ensuring treatment safety under the supervision of attending therapists, physicists, and oncologists.
[IMAGE OMITTED. SEE PDF]
The block-filtered dose rates of 6 MV x-rays were measured using a calibrated ion chamber (PTW, Freiburg, German) at patients’ midplane (15.0/12.5/10.0/7.5 depth) at treatment SSDs of 185, 190, 195, and 200 cm in a solid water phantom, using LINAC's minimum meter-set dose rate of 100 MU/min. Similarly, its beam quality index (TPR20/TPR10) was measured in water with a 10 × 10 cm2 field size at 100 cm source-to-axis distance (SAD).
A custom-designed sliding couch system
A custom-designed sliding couch system (Patent No. ZL202123085880.1), elevated 15 cm over the floor, was manufactured to support patients during the treatment (Figure 1). The sliding couch system is constructed from wood to minimize backscattering as low as possible. Its lateral position is verified with external lasers on the walls, while its longitudinal position is aligned with the optical distance indicator (ODI) projection. The estimated uncertainty of the positioned sliding couch is approximately 1 mm in the lateral direction and 2 mm in the longitudinal direction, owing to the laser width of about 1 mm laterally and the 2 mm width of the ODI projection at 180 cm SSD longitudinally. Once the sliding couch position is determined, therapists lock the wheels at the bottom to prevent any movement during treatment. A 1-cm-thick polymethyl-methacrylate (PMMA) plate was designed to bring the dmax of the block-filtered beam closer to the skin surface and should be placed as close as possible to the patients.
Preparation of TBI-specific dose calculation model
Model configuration
While the transmission block (as discussed in Section 2.1) invalidated the pre-configured dose calculation model in TPS (Eclipse, Anisotropic Analytic Algorithm, Varian Medical Systems, Palo Alto, CA) with 6 MV beam data acquired at 100-cm SSD, we re-configured a TBI-specific dose calculation model using 6 MV beams acquired at the TPS-permitted maximum 140-cm SSD and commissioned it at 180-cm SSD.
The TBI-specific dose calculation model was configured with measured percentage depth dose (PDD), beam profiles (PRF), diagonal profiles of maximum square field (DPR), and output factors at an SSD of 140 cm. The model was calibrated with the absolute point dose measured at a 180-cm SSD to align with the TBI treatment scenario as closely as possible, following the recommendations of the AAPM TG-106 report22 and TPS manufacturer guidelines.23 The PDDs, PRFs, and DPRs were scanned using two cylindrical ion chambers (CC13, IBA Dosimetry, Schwazenbruck, German) using a 3D water tank (Blue Phantom 2, IBA Dosimetry, Schwazenbruck, Germany). Due to the beam divergence effect, the 40 × 40 cm2 field size at 100-cm SAD would augment to 56 × 56 cm2 field size at 140-cm SSD, which exceeds the scanning ranges of the water tank (478 mm × 478 mm × 410 mm). Therefore, we scanned half of the PRFs and DPRs by shifting the water tank ∼25 cm off the beam central axis and then reconstructed complete PRFs and DPRs symmetrically. When acquiring DPRs, we rotated the collimator to 45° and used the PRFs setting to scan diagonal profiles instead of the DPR setting. The output factors and absolute point dose were measured with 10 × 10 cm2 field size and 140-cm SSD using a calibrated ion chamber (PTW, Freiburg, German), adhering to the AAPM TG-106 report recommendations.22 The MLC transmission factor and dosimetric leaf gap (DLG) were measured using “Ada” and “Chair” fields at 10 cm depth and 180 cm SSD, following the TPS vendor recommendations.23
Model commissioning
To verify the accuracy of the TBI-specific dose calculation model, we conducted an end-to-end (E2E) test at 180 cm SSD using a Farmer-type ion chamber (CC13, IBA Dosimetry, Schwazenbruck, German) with a nominal 30 × 30 cm2 field size (actual size at 100 cm SAD, and so far below) in both a solid water phantom and heterogeneous thorax phantom (CIRS anthropomorphic thorax phantom, SunNuclear, Florida, USA). Phantoms were scanned individually using a large-bore CT simulator (GE Discovery 590 RT, GE Healthcare, Milwaukee, WI, USA) with a 2.5 mm slice thickness. The image dataset was then imported into TPS (Eclipse v.15.1, Varian Medical System, Palo Alto, CA, USA) for subsequent end-to-end dose calculation.
We selected six depth points and ten off-axis points for E2E verification. Point doses at depths of 1.5, 5.0, 10.0, 15.0, 20.0, and 25.0 cm were measured using a calibrated ion chamber (PTW, Freiburg, German) and compared against TPS-calculated doses at corresponding points. Similarly, ten off-axis point doses on the collimator X-axis, at an interval of 10 cm (from central axis up to 100 cm off-central axis distance, simulating a 200 cm tall patient), were measured at 5 cm deep and compared against TPS dose calculation. Due to the solid water phantom's size exceeding the measurement range, we moved the phantom laterally and rotated the gantry angle 15° and 30°, respectively, for off-axis dose measurements. However, measurements at 0° gantry angle were performed directly in the CIRS thorax phantom, considering perpendicular delivery for the thorax region during treatment.
Regarding transmission factor and DLG, we used a 2-D array detector (MatriXX, IBA, Schwazenbruck, Germany) for planar dose collection. We imported two dynamic MLC test fields into TPS: the dynamic “chair” and the AIDA test fields.24 The AIDA pattern contains five rectangle fields shaped by MLC, while the “Chair” test25 comprises three main parts for evaluating transmission and DLG. All measurements were conducted in an extended SSD setup. A 2D gamma analysis (2%, 2 mm) with a 10% threshold was performed in the global mode.
in-vivo dose measurement devices and methods
For in-vivo dose monitoring, we employed a wireless SunNuclear IVD 2TM System with two ISORADTM and ten QEDTM diode detectors (SunNuclear, Florida, USA), as illustrated in Figure 2(a). These detectors were calibrated against a 0.6 cm3 calibrated ion chamber (PTW, Freiburg, German) at the dose-maximum depth of 180 cm SSD in a 30 × 30 cm2 field size, as shown in Figure 2(b). The QED diode detectors are flat-designed with active dimensions of 0.8 × 0.8 mm2 for easy placement on the patients’ surface, recommended for measuring direct incident beams. ISORAD diode detectors have a cylindrical active volume of 1.4 mm diameter for angular-independent measurements, suitable for oblique incident beams.
[IMAGE OMITTED. SEE PDF]
We marked six points (head, neck, lungs, umbilicus, hip, and knee) in the patient's skin during CT simulation, as illustrated in Figure 2(c), and placed diodes (SunNuclear, Florida, USA) on the marked points for real-time dose monitoring during TBI treatment. Concurrently, a calibrated ion chamber (PTW, Freiburg, German) was placed on the umbilicus for absolute dose reference. Therapists would immediately stop the treatment and contact responsible physicists and oncologists to address significant variations (>5% tolerance) between the real-time readings and TPS-calculated dose. Physicists would adapt treatment plans for the remaining fractions based on the in-vivo readings if necessary.
TBI implementation
Prescription
The prescription dose for TBI is 8–10 Gy, delivered in 4–5 fractions over 2–3 consecutive days, twice daily. The planning target volume (PTV) encompasses the entire body with a 5 mm contraction from the skin and is typically divided into four sub-parts: head, thorax, abdomen, and legs. The plan criteria require that 90% of the prescription dose covers 95% of the PTV, and the MLD should remain below 7 Gy.
CT simulation
During the CT simulation, patients were immobilized with vacuum bags in both supine and prone positions. Patients lay flatly in each position with their arms along the body. Before scanning, repositioning lines and 12 monitoring points were marked with radiation-visible markers. The upper and lower body scans were conducted separately and then reconstructed into a whole-body CT using in-house software,26 due to the CT scanner's range limitation of approximately 160 cm.
TBI planning
All clinical plans were designed at three couch positions using a TBI-specific dose calculation model in TPS (Eclipse, Varian Medical Systems, Palo Alto, CA). Instead of the original treatment couch, a custom-designed sliding couch system was employed due to its limited travel range during treatment. The couch angle was set to 90 degrees during planning to mimic the patient's lying position under the gantry, and a TBI-specific table tolerance was created to accommodate this. Large open fields were used for head and leg irradiations at two different couch positions to improve treatment efficiency. For the thorax and abdomen regions, large fields with MLC shielding were used for lung protection at another couch position. The static MLC shielding was based on the Beam's Eye View (BEV) projection. All field widths were set to 40 cm laterally to account for possible patient intra-fractional movement. Field junctions of the adjacent fields were shifted to smear hot/cold regions. Upon completion of the planning, oncologists evaluated the plans based on dose-volume histogram (DVH) for lung-sparing and dose lines for target coverage.
In the aspect of planning strategies, the main challenges with the open-field TBI technique are achieving dose homogeneity at field junctions and patient edges due to significant variances in patient thickness. The hot/cold areas in the field junctions could be smeared by shifting the junctions of each field. Optimizing MLC sequencing to shield hot areas at patient extremities might help, though it can affect plan robustness. Adding an appropriate bolus to compensate for changes in the patients’ thickness during treatment can also eliminate hot/cold areas. Additionally, a decent movement tolerance is necessary when setting up field widths to ensure adequate dose delivery. Typically, a ±1 cm setup uncertainty is recommended for planning robustness.
Clinical workflow and quality controls for TBI treatment
The clinical workflow concludes with a multidisciplinary team discussion, baseline assessment of the patient and prescription, patient immobilization and CT scanning, treatment planning in TPS, initial plan check and dosimetry check, treatment preparation, in-room timeout checklist, beam delivery, in-vivo monitoring, and follow-ups after treatment.
To assure the safety of the TBI treatment, we implement several measures such as machine interlocks, pretreatment checklists, and in-vivo dose monitoring. After planning, a qualified medical physicist reviews all TBI plans according to the TBI-specific initial plan chart check (Table 1), which includes checks for dose rate, field size, block positions, and TBI-specific dose calculation model. Physicists also perform a patient-specific quality assurance (PSQA) by measuring the absolute dose of delivered beams using a calibrated ion chamber (PTW, Freiburg, German) in a solid water phantom before approving the treatment plans.
TABLE 1 TBI-specific initial planning chart checks.
No. | Check items | Status |
1 | Patient identification and prescription check | |
2 | Integrity of the planning CT | |
3 | Planning SSD check | |
4 | Isocenter placement check (usually at umbilicus) | |
5 | Patient treatment position checks (FFS, FFP) for each sub-plan | |
6 | Check if field's maximum gantry rotation angle ≤ 30 | |
7 | Check if field sizes are enough tolerance for patient movements | |
8 | Check if using TBI-specific dose calculation model | |
9 | Dose rate check (for lungs and abdomen, dose rates ≤100 MU/min) | |
10 | Check multi-leaf collimator aperture is shaped for lungs in both AP and PA positions, and sparing MLD ≤7 Gy | |
11 | Add block to each treatment field | |
12 | Check if PMMA board needed for planning | |
13 | If PMMA needed, check density override of the PMMA board | |
14 | If PMMA needed, state the height of the PMMA board |
Special attention is given to the safety of the transmission block to prevent risks of falling and misplacement. In our institution, the transmission block is securely screwed onto a TBI-dedicated block tray which is mounted on the slots, preventing it from falling or inter-fractional misplacement. During treatment planning, this transmission block is added to each field in TPS with custom coding. Additionally, the maximum gantry rotation angle is limited to 10° for TBI planning to ensure the transmission block's stability. A qualified physicist performs TBI-specific initial planning chart checks to ensure these measures are correctly implemented. On treatment day, the transmission block is mounted on the slot using TBI-specific custom coding, and patients are repositioned using external lasers on the wall and ceiling. Two therapists and attending oncologists conduct a TBI-specific in-room timeout checklist (Table 2), which includes daily dose checks with the block, 40-min disinfection by ultraviolet light, setup of the sliding couch and vacuum bags, patient repositioning with wall and ceiling lasers, placement of in vivo monitoring devices, and SSD checks. On the treatment console, therapists must verify the block placement again before the beam is turned on; otherwise, the beam will be automatically interlocked. Once everything is confirmed, therapists start the treatments and monitor the real-time readings. If more than a 5% dose variation is detected, therapists immediately stop the beam and contact attending physicists and oncologists to address the issues.
TABLE 2 TBI-specific in-room timeout checklist.
No. | Roles | Check items | Status |
1 | Nurses | Room disinfection by ultraviolet light for 40 min | |
2 | Therapists | Planning checks and LINAC preparation | |
3 | Therapists/Physicists | Position TBI custom-designed sliding couch with external lasers | |
4 | Therapists/Physicists | Check the usage and height of the PMMA board | |
5 | Therapists/Physicists | Check placement of transmission block and machine interlocks | |
6 | Therapists/Physicists | Reposition the patient | |
7 | Therapists/Physicists | SSD checks | |
8 | Physicists | Placement and connection of in-vivo dose measurement devices | |
9 | Therapists | Checked transmission block interlocks on the console | |
10 | Therapists | Beam on |
After TBI treatments, the toxicity is evaluated during follow-ups according to toxicity criteria from the Radiation Therapy Oncology Group and the European Organization for Research and Treatment of Cancer (RTOG/EORTC).28
RESULTS
Instantaneous dose rates after the transmission block
As SSD varies from 185 to 200 cm, the instantaneous dose rates were around 7 cGy/min at patients’ mid-plane (15.0/12.5/10.0/7.5 cm), as tabulated in Table 3. The beam quality index (TPR20/TPR10) of the block-filtered 6 MV x-rays is 0.695, compared to 0.670 for the un-blocked 6 MV beam at 100 cm SAD with 10 × 10 cm2 field size. The quantifies the beam hardening effect induced by the transmission block. As the beam hardens, the build-up region might expand, necessitating consideration of the absorbed dose in the shallow area, which implies potential usage of the PMMA board.
TABLE 3 The instantaneous dose rates after transmission block with variable source-to-chamber distances (cm).
Source-to-chamber distance (cm) | 185.0 | 190.0 | 195.0 | 200.0 |
Depth (cm) | 15.00 | 12.50 | 10.00 | 7.50 |
Measured point dose at patients’ midplane (cGy) | 7.08 | 6.75 | 7.02 | 7.30 |
Commissioning of TBI-specific dose calculation model
TPS modeling
The absolute dose of block-filtered 6 MV x-rays at 5 cm depth with a 10 × 10 cm2 field size is 763.744 MU for 1 Gy at 180 cm SSD (as compared to 100 MU for 1 Gy without transmission lead at 100 cm SSD). Compared to the previous model, the new output factors for various field sizes increased, indicating enhanced scattering caused by the transmission block.
Figures 3 and 4 present the measured PDDs and PRFs. The PDDs for 6 MV x-rays with transmission lead consistently increased after the depth of maximum dose (dmax) compared to 6 MV x-rays without transmission lead. When comparing measured PDDs with transmission lead against TPS-calculated PDDs with the transmission lead, the variance was primarily observed in the build-up regions before dmax. For PRFs, there was a noticeable difference between measurements with transmission lead and TPS-calculated doses with transmission lead with for a 3 cm × 3 cm field size. However, there were few differences in data with other field sizes.
[IMAGE OMITTED. SEE PDF]
[IMAGE OMITTED. SEE PDF]
Our TBI setting demonstrated a transmission factor of 0.0114 and a DLG value of 3.17 mm (compared to 0.0135 and 1.80 mm for standard treatment distances). Compared to the standard TPS model at 100 cm SSD for 6 MV X-rays, the transmission factor decreased by approximately 15.6%, and DLG increased by 76.1%. These changes in DLG, induced by the transmission block, are challenging to quantify due to the involvement of multiple factors such as the beam hardening effect and the extended measurement distance.
E2E verification
AAPM MPPG27 recommends a 2% tolerance for the average difference between measured and TPS-calculated doses in the low-gradient target region for typical SSDs. In this study, the discrepancy between the TPS-calculated dose and measurements using a calibrated ion chamber (PTW, Freiburg, German) ranges from 0.22% to 1.47% for depth dose and off-axis dose in a solid water phantom, as shown in Tables 4 and 5. Although the solid water phantom validation was conducted at the extended SSDs, our results still fall within AAPM MPPG recommended tolerance.
TABLE 4 Depth dose validation results on solid water phantom.
Depth (cm) | 1.5 | 5.0 | 10.0 | 15.0 | 20.0 | 25.0 |
Dphantom (cGy) | 10.47 | 9.62 | 8.145 | 6.785 | 5.59 | 4.54 |
DTPS (cGy) | 10.4 | 9.5 | 8.1 | 6.8 | 5.6 | 4.6 |
Dose difference (%) | 0.67% | 1.25% | 0.55% | −0.22% | −0.18% | −1.32% |
TABLE 5 Off-axis dose validation results on solid water phantom.
Gantry angle (°) | 0 | 15 | 30 | ||||||||
Off-axis distance (cm) | 0 | 10 | 20 | 30 | 40 | 50 | 60 | 70 | 80 | 90 | 100 |
Dphantom (cGy) | 6.72 | 6.78 | 6.67 | 6.35 | 6.26 | 5.77 | 5.34 | 5.07 | 4.91 | 4.41 | 3.35 |
DTPS (cGy) | 6.70 | 6.70 | 6.58 | 6.40 | 6.20 | 5.7 | 5.40 | 5.15 | 4.85 | 4.40 | 3.40 |
Dose difference (%) | 0.30 | 1.19 | 1.37 | 0.78 | 0.97 | 1.23 | 1.11 | 1.55 | 1.24 | 0.23 | 1.47 |
For E2E verification using a CIRS phantom, AAPM MPPG recommends a 5% relative dose difference for IMRT commissioning with the photon beams.27 In this study, our results (Table 6) demonstrated a 1.2%–4.85% difference in dose to lungs, spinal cord, and normal tissues, indicating that the TBI-specific dose calculation model demonstrated acceptable accuracy when used for calculating TBI plans with a transmission block at extended SSDs of 180–200 cm.
TABLE 6 End-to-end (E2E) testing in anthropomorphic CIRS Thorax phantom.
Calculated and measured condition | Gantry angle = 0°, MU = 200, SSD = 180 cm, Field size = 30 × 30 cm2 | |||||
Number of points | 1 | 2 | 3 | 4 | 5 | 6 |
Materials | Tissue | Tissue | Tissue | Tissue | Spinal cord | Lungs |
Off-axis distance to the central axis (cm) | 0 | 0 | 0 | 4 | 0 | 8 |
Depth of points (cm) | 7.0 | 3.0 | 10.0 | 7.0 | 15.0 | 10.0 |
DMeasured (cGy) | 8.83 | 10.51 | 7.64 | 9.06 | 5.83 | 9.51 |
DTPS (cGy) | 8.60 | 10.00 | 7.90 | 8.70 | 5.90 | 9.10 |
Dose difference (%) | 2.60 | 4.85 | 3.40 | 3.97 | 1.20 | 4.31 |
Plan parameters
The dose distribution of the TBI planning is illustrated in Figure 5. The PTV coverage outside the lung regions is V90%PD (PTV) ≈ 95%, V115%PD (PTV) ≤ 4.0%. For lung protection, the MLD ≈ 6.8 Gy, D95% ≈ 6.0 Gy, and Dmax ≈ 8.0 Gy (Table 7).
[IMAGE OMITTED. SEE PDF]
TABLE 7 Dose distribution of open-field TBI technique.
Techniques | Open-field TBI technique |
Planning time | ∼ 60 min |
Calculation algorithm | TBI-specific dose calculation model |
PTV without lungs | V90%PD ≈ 95%, V115%PD ≤ 5.0% |
Lungs | MLD ≈ 6.8 Gy, D95%(lungs) ≈ 6.0 Gy, Dmax(lungs) ≈ 8.0 Gy |
Hot areas | Field junctions and body edges |
Cold areas | Field junctions and tips of the toes |
Delivery time | ∼ 50 mins |
Cold areas were identified in two regions: one in the upper thoracic regions as a trade-off for lung protection, which still ensured about 5.8–6.0 Gy dose coverage, and the other at the tips of the toes, which could be improved using a 5 mm bolus. In contrast, hot areas are mainly concentrated around the edges of the limbs due to the apparent thickness decrease compared to the surrounding tissues but still maintained V115% ≤ 5%. Besides, the final dose distribution indicated that an additional PMMA board is not required.
Regarding planning and delivery efficiency, a TBI plan with the TBI-specific dose calculation model took about 1 h to complete, and beam delivery took about 50 min.
In-vivo dose monitoring and follow-ups
Thirty patients received hyper-fractionated TBI (8–10 Gy in 4–5 fraction, BID, over three consecutive days) utilizing the open-field TBI technique. All fractions were monitored in real-time using six diodes and one calibrated ion chamber (PTW, Freiburg, German) in prone and supine positions, respectively. As compared to the TPS-calculated dose in treatment planning, the real-time readings demonstrated within 5% mean relative difference from (MRD) from head to patella, as stated in Table 8, meeting AAPM requirements for clinical treatment.27 After treatment, we followed these patients regularly to evaluate their lung toxicity for 6–12 months and found that no patients experienced grade 2 or higher radiation pneumonitis, based on RTOG/EORTC criteria.28
TABLE 8 in-vivo dose monitoring results using diodes and calibrated ion chamber.
PA positions | Head | Neck | Left lung | Umbilicus | Hip | Patella | Ion chamber |
Dose MRD (%) | −1.49 ± 2.41 | 0.10 ± 1.83 | −0.66 ± 2.31 | −0.45 ± 3.27 | −1.39 ± 2.71 | 2.10 ± 2.15 | −1.55 ± 2.11 |
AP Position | Head | Neck | Left Lung | Umbilicus | Hip | Patella | Ion chamber |
Dose MRD (%) | −1.49 ± 1.84 | 0.21 ± 2.85 | −2.59 ± 2.76 | −1.59 ± 2.64 | −1.38 ± 3.03 | 0.40 ± 3.55 | −3.42 ± 1.51 |
DISCUSSION
This is the first study to configure a TBI-specific dose calculation model using beam data acquired at SSD of 140 cm and apply it to TBI treatment at SSDs ranging from 180 to 200 cm. In contrast, some studies29 applied the standard dose calculation model (configured at 100 cm SSD) to TBI treatment at extended SSDs after some commissioning procedures. However, that does not work in our cases. However, the goal of reducing PT after TBI treatment promoted us to employ a lead transmission block to reduce dose rates down to ≤10 cGy/min at the patients’ midplane at 180−200 cm SSDs, leading to alterations in beam properties. Consequently, we reconfigured a TBI-specific dose calculation model for 6-MV x-rays at a TPS-permitted maximum SSD of 140 cm, balancing the TBI treatment scenarios with the requirements of the TPS. LINCAs with lower meter-set dose rates may not necessitate an additional transmission block, thus obviating the need for subsequent commissioning procedures.
Configuration of a TBI-specific model at an extended distance is challenging since the field size there has been beyond the measurement range of the water tank. We have to segment the beam scanning and reconstruct them to complete PRFs and DPRs. Also, the methodologies to determine the DLG values at a 180 cm SSD could be further investigated even though the current DLG setting demonstrated acceptable performance with our TBI technique based on current clinical results, possibly because we did not employ inverse-modulated optimization. Nevertheless, the dose difference is within 5% in the E2E verification, meeting the AAPM recommendations.27 Also, some studies directly employed a 600 MU/min instantaneous dose rate to irradiate patients at around 100 cm SSD30–34 but there are no follow-up results for us to refer to. We still cannot take the risk to level up the instantaneous dose rates at the beginning but it could be taken into consideration in the future. So far, we have monitored 30 patients for PT over 6 months. No patients experienced grade 2 or higher radiation pneumonitis, accounting for RTOG/EORTC toxicity criteria,28 which results from the combination of both a low instantaneous dose rate (≤10 cGy/min) and a low total dose to the lungs (MLD ≤ 7 Gy).
There are some limitations in our study. First, the prescription of the hyper-fractionated TBI (8–10 Gy) in our center is relatively lower as compared to others (typically 10–12 Gy).9 It might be another reason contributing to the low incidence of PT due to fractionation changes. Also, lung shielding usually causes unexpected dose heterogeneity in the thoracic regions, which might increase the risk of transplantation failure. A sequent boost dose35,36 should be considered in case of marrow residue. For instance, Hansen35 incorporated VMAT or IMRT boosts in four out of seven fractions to thoracic regions at standard SSDs combined with extended-SSD TBI treatment as a strategy to address this issue.
CONCLUSIONS
A low instantaneous dose rate TBI technique can be implemented in the clinic.
AUTHOR CONTRIBUTIONS
Zhengxin Gao contributed to all aspects of the research and manuscript preparation. Qiuyi Xu and Qing Gong were responsible for the configuration and validation of TBI-specific TPS models, and the implementation of TBI. Dr. Fengjiao Zhang and Dr. Qiaoying Hu focused on the clinical aspects of TBI treatment. Yaling Hong conducted in-vivo dose monitoring. Dr. Qi Yu carried out the follow-up analysis of post-TBI patients. Qing Gong and Dr. Shen Fu supervised the research, with Qing Gong overseeing the physics aspects and Dr. Shen Fu overseeing the medical aspects.
ACKNOWLEDGMENTS
Thanks to Pei-feng Wong (medical physicist) from MD Anderson Cancer Center for his suggestions on the TBI project and his review on data acquisition.
CONFLICT OF INTEREST STATEMENT
This study was supported by grants from the research project of Shanghai Huangpu District Health System (No. HLZ202218); the Project of State Key Laboratory of Radiation Medicine and Protection, Soochow University (No. GZK1202131); the Guangzhou High-tech, Key, and Characteristic Clinical Technology of the Guangzhou Municipal Health Commission (grant number: 2023C-TS64); the National Natural Science Foundation of China (81773225).
Wong JYC, Filippi AR, Dabaja BS, Yahalom J, Specht L. Total body irradiation: guidelines from the international lymphoma radiation oncology group (ILROG). Int J Radiat Oncol Biol Phys. 2018;101(3):521‐529.
Paix A, Antoni D, Waissi W, et al. Total body irradiation in allogeneic bone marrow transplantation conditioning regimens: a review. Crit Rev Oncol Hematol. 2018;123:138‐148.
Stephens SJ, Thomas S, Rizzieri DA, et al. Myeloablative conditioning with total body irradiation for AML: Balancing survival and pulmonary toxicity. Adv Radiat Oncol. 2016;1(4):272‐280.
Dyk JV, Galvin JM, Glasgow GP, et al. AAPM Report No. 17: The physical aspects of total and half body photon irradiation. American Association of Physicists in Medicine; 1986: 6.
Gao RW, Weisdorf DJ, DeFor TE, Ehler E, Dusenbery KE. Influence of total body irradiation dose rate on idiopathic pneumonia syndrome in acute leukemia patients undergoing allogeneic hematopoietic cell transplantation. Int J Radiat Oncol Biol Phys. 2019;103(1):180‐189.
Gore EM, Lawton CA, Ash RC, Lipchik RJ. Pulmonary function changes in long‐term survivors of bone marrow transplantation. Int J Radiat Oncol Biol Phys. 1996;36(1):67‐75.
Sampath S, Schultheiss TE, Wong J. Dose response and factors related to interstitial pneumonitis after bone marrow transplant. Int J Radiat Oncol Biol Phys. 2005;63(3):876‐884.
Esiashvili N, Xiaomin L, Hunger S, et al. Association of higher lung dose received during total body irradiation for allogeneic hematopoietic stem cell transplantation in children with acute lymphoblastic leukemia with inferior progression‐free and overall survival: a report from the Children's Oncology Group. J Clin Oncol. 2015;33: [eLocator: 10030].
Reilly M, Dandapani SV, Kumar KA, et al. ACR‐ARS practice parameter for the performance of total body irradiation. Am J Clin Oncol. 2023;46(5):185‐192.
Carruthers SA, Wallington MM. Total body irradiation and pneumonitis risk: a review of outcomes. Br J Cancer. 2004;90(11):2080‐2084.
Kim DY, Kim IH, Yoon SS, Kang HJ, Shin HY, Kang HC. Effect of dose rate on pulmonary toxicity in patients with hematolymphoid malignancies undergoing total body irradiation. Radiat Oncol. 2018;13(1):180.
Kim TH, Rybka WB, Lehnert S, Podgorsak EB, Freeman CR. Interstitial pneumonitis following total body irradiation for bone marrow transplantation using two different dose rates. Int J Radiat Oncol Biol Phys. 1985;11(7):1285‐1291.
Ozsahin M, Belkacémi Y, Pène F, et al. Interstitial pneumonitis following autologous bone marrow transplantation conditioned with cyclophosphamide and total‐body irradiation. Int J Radiat Oncol Biol Phys. 1996;34(1):71‐77.
Weiner RS, Bortin MM, Gale RP, et al. Interstitial pneumonitis after bone marrow transplantation. Assessment of risk factors. Ann Intern Med. 1986;104(2):168‐175.
Gogna NK, Morgan G, Downs K, Atkinson K, Biggs J. Lung dose rate and interstitial pneumonitis in total body irradiation for bone marrow transplantation. Australas Radiol. 1992;36(4):317‐320.
Oya N, Sasai K, Tachiiri S, et al. Influence of radiation dose rate and lung dose on interstitial pneumonitis after fractionated total body irradiation: acute parotitis may predict interstitial pneumonitis. Int J Hematol. 2006;83(1):86‐91.
Leung PM, Rider WD, Webb HP, et al. Cobalt‐60 therapy unit for large field irradiation. Int J Radiation Oncol Biol Phys. 1981;7(6):705‐712.
Peters VG, Herer AS. Modification of a standard cobalt‐60 unit for total body irradiation at 150 cm SSD. Int J Radiation Oncol Biol Phys. 1984;10(6):927‐932.
Miller RJ, Langdon EA, Tesler AS. Total body irradiation utilizing a single 60Co source. Int J Radiat Oncol Biol Phys. 1976;1(5‐6):549‐552.
Glasgow GP, Mill WB. Cobalt‐60 total body irradiation dosimetry at 220 cm source‐axis distance. Int J Radiat Oncol Biol Phys. 1980;6(6):773‐777.
Serban M, Seuntjens J, Roussin E, et al. Patient‐specific compensation for Co‐60 TBI treatments based on Monte Carlo design: a feasibility study. Phys Med. 2016;32(1):67‐75.
Das IJ, Cheng C‐W, Watts RJ, et al. Accelerator beam data commissioning equipment and procedures: report of the TG‐106 of the Therapy Physics Committee of the AAPM. Med Phys. 2008;35:4186‐4215.
Varian Medical System Inc. Eclipse Algorithms Reference Guide. Palo Alto, CA: Varian Medical Systems, Inc; 2010.
Calvo‐Ortega JF, Greer PB, Hermida‐López M, et al. Validation of virtual water phantom software for pre‐treatment verification of single‐isocenter multiple‐target stereotactic radiosurgery. J Appl Clin Med Phys. 2021;22(6):241‐252.
Van Esch A, Bohsung J, Sorvari P, et al. Acceptance tests and quality control (QC) procedures for the clinical implementation of intensity‐modulated radiotherapy (IMRT) using inverse planning and the sliding window technique: experience from five radiotherapy departments. Radiother Oncol. 2002;65(1):53‐70.
Gao ZX, Xu QY, Gong F, et al. Total‐body CT simulation and image reconstruction method for total‐body irradiation. Chin J Med Phys. 2023;30(3):286‐290.
Smilowitz JB, Das IJ, Feygelman V, et al. AAPM Medical Physics Practice Guideline 5.a.: commissioning and QA of treatment planning dose calculations‐megavoltage photon and electron beams. J Appl Clin Med Phys. 2015;16(5):14‐34.
Cox JD, Stetz J, Pajak TF. Toxicity criteria of the Radiation Therapy Oncology Group (RTOG) and the European Organization for Research and Treatment of Cancer (EORTC). Int J Radiat Oncol Biol Phys. 1995; 31(5):1341‐1346.
Pierce G, Balogh A, Frederick R, et al. Extended SSD VMAT treatment for total body irradiation. J Appl Clin Med Phys. 2019;20(1):200‐211.
Springer A, Hammer J, Winkler E, et al. Total body irradiation with volumetric modulated arc therapy: dosimetric data and first clinical experience. Radiat Oncol. 2016;11:46.
Symons K, Morrison C, Parry J, et al. Volumetric modulated arc therapy for total body irradiation: a feasibility study using Pinnacle3 treatment planning system and Elekta AgilityTM Linac. J Appl Clin Med Phys. 2018;19(2):103‐110.
Teruel JR, Taneja S, Galavis EP, et al. Automatic treatment planning for VMAT‐based total body irradiation using Eclipse scripting. J Appl Clin Med Phys. 2021;22(3):119‐130.
Guo BQ, Chrian S, Erin M, et al. Image‐guided volumetric‐modulated arc therapy of total body irradiation: an efficient workflow from simulation to delivery. J Appl Clin Med Phys. 2021;22(10):169‐177.
Jiang XQ, Su BF, Chen CX, et al. First clinical application and dosimetric verification of total body irradiation with volumetric modulated arc therapy. Chin J Radiat Oncol. 2021;30(8):828‐834.
Hansen AT, Rose HK, Yates ES, et al. Two compound techniques for total body irradiation. Tech Innovations Patient Support Radiat Oncol. 2021;21:1‐7.
Tyson TE, Podgorsak MB, Singh AK, et al. Improved dose homogeneity using electronic compensation technique for total body irradiation. J Appl Clin Med Phys. 2018;19(3):159‐167.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2024. This work is published under http://creativecommons.org/licenses/by/4.0/ (the "License"). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
Purpose
This paper describes the implementation of an instantaneous low‐dose‐rate total body irradiation (TBI) technique using block‐filtered 6 MV X‐rays with a linear accelerator (LINAC) to reduce pulmonary toxicity.
Methods
In the absence of dedicated TBI‐specific meter‐set dose rates in LINAC and sufficient treatment room size, a 2‐cm‐thick transmission block was used together with a 200‐cm source‐to‐surface distance (SSD) to reduce the instantaneous dose rates of 6 MV x‐rays down to 10 cGy/min, thus alteration to the beam properties. A TBI‐specific dose calculation model was built with data acquired at the treatment planning system (TPS)‐permitted maximum 140‐cm SSD and was validated in phantoms at a 180‐cm SSD. As for planning strategies, we adopted large anterior‐to‐posterior/posterior‐to‐anterior (AP/PA) open fields with multi‐leaf collimator shielding for lungs to achieve target coverage, lung protection, and efficient dose delivery. A custom‐designed sliding couch (Patent No. ZL202123085880.1) was manufactured to support patients during treatment. Measures to control the quality and safety of TBI treatment include machine interlocks, pretreatment checklists, and in‐vivo dose monitoring.
Results
The instantaneous dose rate of block‐filtered 6MV X‐ray was reduced to approximately 7.0 cGy/min at 12.5–7.5 cm depth with a 185–200 cm SSD. The dose calculated by TPS differs from the measurements by 0.15%–1.55% in the homogeneous phantom and 1.2%–4.85% in the CIRS thorax phantom. The open‐field TBI technique achieved V90% (PTV) ≈ 96.8% and MLD = 6.6 Gy with 1‐h planning and 50‐min beam delivery in a single fraction. From February 2021 to July 2023, 30 patients received TBI treatments in our center, and in‐vivo monitoring results differed from TPS calculations by −1.49%–2.10%. After 6–12 months of follow‐ups, all the patients treated in our center showed no pulmonary toxicities of grade 2 or higher.
Conclusion
A low instantaneous dose rate TBI technique can be implemented in the clinic.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer