Abstract

Aim

This study compares the precision, accuracy, and user experience of 3D body surface scanning of human subjects using the Artec Leo hand-held scanner and the iPad Pro as 3D scanning devices for capturing cervical and craniofacial data. The investigation includes assessing methods for correcting 'dropped head syndrome' during scanning, to demonstrate the ability of the scanner to be used to reconstruct body surface of patients.

Methods

Eighteen volunteers with no prior history of neck weakness were scanned three times in three different positions, using the two different devices. Surface area, scanning time, and participant comfort scores were evaluated for both devices. Precision and accuracy were assessed using Mean Absolute Deviation (MAD), Mean Absolute Percentage Error (MAPE), and Intra-Class Correlation Coefficients (ICC).

Results

Surface area comparisons revealed no significant differences between devices and positions. Scanning times showed no significant difference between devices or positions. Comfort scores varied across positions. MAD analysis identified chin to chest measurements as having the highest variance, especially in scanning position 3. However, no statistical differences were found. MAPE results confirmed accuracy below 5% error for both devices. ICC scores indicated good reliability for both measurement methods, particularly for chin to chest measurements in positions 1 and 3.

Conclusion

The iPad Pro using the Qlone app demonstrates a viable alternative to the Artec Leo, particularly for capturing head and neck surface area within a clinical setting. The scanning resolution, with an error margin within ±5%, is consistent with clinically accepted standards for orthosis design, where padding and final fit adjustments allow for bespoke devices that accommodate patient comfort. This study highlights the comparative performance of the iPad, as well as suggests two methods which can be used within clinics to correct head drop for scanning.

Details

Title
Comparative evaluation of Artec Leo hand-held scanner and iPad Pro for 3D scanning of cervical and craniofacial data: assessing precision, accuracy, and user experience
Author
Spears, Samuel D. J. 1   VIAFID ORCID Logo  ; Lester, Thomas 1   VIAFID ORCID Logo  ; Torii, Ryo 2   VIAFID ORCID Logo  ; Kalaskar, Deepak M. 1   VIAFID ORCID Logo 

 University College London, Division of Surgery and Interventional Sciences, London, United Kingdom (GRID:grid.83440.3b) (ISNI:0000 0001 2190 1201) 
 University College of London, Department of Mechanical Engineering, London, United Kingdom (GRID:grid.83440.3b) (ISNI:0000000121901201) 
Pages
39
Publication year
2024
Publication date
Dec 2024
Publisher
Springer Nature B.V.
e-ISSN
23656271
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
3132727419
Copyright
© The Author(s) 2024. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.