This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
1. Introduction
Ultrasound is one of the main imaging techniques used in medicine and is characterized by the advantages of being noninvasive, low-cost, and free of ionizing radiation, which makes this test an undoubtedly useful tool in clinical practice [1]. However, it suffers from some limitations, including the need for substantial experience and training and interobserver variability [2].
Remote ultrasound guided by telerobotic articulated systems (telerobotic ultrasound) is a technology in continuous development for more than two decades that makes it possible to perform ultrasound scans remotely using a device capable of reproducing the movement of the sonographer’s arm [3]. This technology can have a remarkable impact in settings where healthcare is limited, such as in rural areas [4], offshore oil platforms, military stations, or aircraft [5]. Moreover, it has played a relevant role during the COVID-19 pandemic, as it has allowed the performance of ultrasound scans on patients in isolation [6].
The use of these systems has shown efficacy and cost-effectiveness in different anatomical regions and indications, including thyroid, carotid, prenatal, echocardiography, and abdominal ultrasound [7]. However, the introduction of this technology into routine clinical practice may be limited by several disadvantages, including a longer scanning time [8] or worse visualization of structures compared to standard ultrasound [9]. Some studies have reported comparable results for these potential limitations using remote abdominal ultrasonography, but the need for further studies has been emphasized [10].
There is a growing limitation in the availability of radiology specialists dedicated to abdominal ultrasound in second-level hospitals and peripheral health centers due to factors such as the lack of radiologists, the greater interest in specialization in third-level hospitals, or the international variability in the qualification requirements for professionals dedicated to ultrasound (sonographers, radiologists, obstetricians, cardiologists, etc.) [11]. All this has led to a growing interest in assessing the usefulness of telerobotic ultrasound in the context of nonurgent care demand.
From a resource management point of view, demonstrating the utility and cost-effectiveness of telerobotic ultrasound could be a paradigm shift by enabling appropriate waiting list logistics in a centralized manner. There is currently a range of commercial telerobotic ultrasound scanners that can be implemented in clinical practice [12–14]. However, the studies available to date are limited and do not yet offer the necessary guarantees to patients in terms of diagnostic quality, scan times, and doctor–patient communication under the radiological quality criteria typical of our environment.
The aim of this work is to evaluate the usefulness of remote ultrasound of the abdomen using an articulated telerobotic system in a real clinical practice setting through a prospective study.
2. Methodology
2.1. Patient Selection
This prospective study was approved by the Provincial Ethics Committee of Granada (Code 1235-N-20). A consecutive cohort of patients attended at the Hospital de Alta Resolucion de Alcala la Real (secondary hospital), located 58.9 km from the Hospital Universitario Virgen de las Nieves (reference hospital), who were requested to undergo an ultrasound examination of the abdomen and pelvis by the primary care physician between March 1 and April 1, 2022, were selected. Patients were invited to participate in the study on a voluntary basis. The inclusion criteria were as follows:
‐
‐ Suspected abdominal or pelvic pathology.
‐ Locality located >50 km from a tertiary-level hospital.
The exclusion criteria were as follows:
‐ Suspected musculoskeletal pathology.
‐ Request for ultrasound on a preferential or urgent basis.
‐ Patient’s refusal to participate in the study.
2.2. Image Acquisition
The MELODY System (AdEchoTech) telerobotic system was used, which consists of an articulated arm installed in the ultrasound room, next to the table where the patient is positioned [10]. The articulated arm makes it possible to reproduce the remote movements of the hand reliably (360° of precession, 45° of nutation, and 360° of self-rotation) with an accuracy equal to or less than 1° [15].
The ultrasound equipment used was a SonixTablet (BK Ultrasound) equipped with a high-frequency linear probe (5–13 MHz) and a low-frequency convex probe (2–6 MHz). All the scans in the study were performed with the convex probe with a standardized abdominal imaging protocol.
Each ultrasound was performed by an on-site radiology technician and an abdominal radiologist at the referral hospital. The ultrasound protocol followed in the study is described in File S1. Using a high-speed Internet connection system and two touch screens, the technician, the radiologist, and the patient communicated via videoconference during the examination (Figure 1). After a training period of 2 weeks with 30 volunteer patients (not included in the study), two radiologists with experience in abdominal ultrasound (5 and 10 years) and two radiology technicians performed scheduled ultrasound shifts over a 4-week period. Patients underwent standard ultrasound immediately after the telerobotic ultrasound examination by an on-site radiologist with 10 years of experience who was blinded to the telerobotic ultrasound findings.
[figure(s) omitted; refer to PDF]
2.3. Study Variables
The following patient demographic variables were collected: age, sex, and BMI. For each scan, the indication, the quality of visualization of the main organs (excellent, good, acceptable, poor, and lousy), the imaging findings for each organ, and the scan time were collected. Finally, the radiologist, technician, and patient were asked to evaluate their degree of satisfaction with the test on a Likert scale from 0 to 10 points (lowest and highest possible degree of satisfaction). The meaning of the scores applied by the radiologist is shown in Table 1.
Table 1
Interpretation of the satisfaction scale applied to radiologists who performed telerobotic ultrasound.
Score | Meaning |
0 | Examination not performed |
1 | No relevant abdominal anatomical structures can be adequately visualized |
2 | Only one organ can be adequately visualized |
3 | Only two or three organs can be adequately visualized |
4 | Only four or five organs can be adequately visualized |
5 | Only six or seven organs can be adequately visualized |
6 | Eight or nine organs can be adequately visualized |
7 | All organs can be visualized, but it is not possible to optimally assess any potentially relevant region given the clinical suspicion |
8a | All organs can be visualized, but it is not possible to optimally assess some regions of no potential interest given the suspicion |
9a | All organs can be optimally visualized and evaluated, except for some regions due to the patient’s morphotype |
10a | All organs can be optimally visualized and assessed |
aIn these cases, the sonographer would not recommend performing a regulated ultrasound.
In the case of the technician survey, the following items were asked to be scored from 0 (lowest agreement) to 10 (highest agreement):
• “The use of the telerobotic tele-echography equipment has been useful to me as a healthcare professional.”
• “It has been easy to get familiar with using the equipment.”
• “It has been easy to use the equipment.”
• “Communication with the remote healthcare partner has been adequate for me.”
In the case of the patient, the following items were asked to be scored from 0 (lowest agreement) to 10 (highest agreement):
• “The examination has been comfortable.”
• “I would repeat the examination in the future.”
Subsequently, the same variables were evaluated in the standard ultrasound, except for satisfaction. Finally, the findings not visualized in either of the two ultrasound scans were collected for each organ examined.
2.4. Statistical Analysis
A descriptive study of the patient sample was performed. Qualitative variables were expressed as absolute and relative frequencies and quantitative variables as mean and standard deviation. A bivariate analysis was then performed comparing the variables measured in telerobotic and regulated ultrasound. Qualitative variables were analyzed by McNemar’s test and quantitative variables by Student’s
Finally, multivariate analyses were performed for the prediction of the variables “no visualization of a potentially relevant finding,” “not excellent visualization of an anatomical structure,” and “radiologist recommends regulated ultrasound” (the latter was obtained by dichotomizing the radiologist’s satisfaction using values < 8 and ≥ 8 as cut-off points). A significance value of
3. Results
The final sample consisted of a total of 40 patients (60% women) with a mean age of
The mean satisfaction with telerobotic ultrasound was
The visualization of the different anatomical structures on telerobotic and standard ultrasound is shown in Table 2. Almost all organs were visualized with good or excellent quality in the latter case, with some exceptions in the pancreas (
Table 2
Visualization of anatomical structures in standard and telerobotic ultrasound.
Organ | Visualization | ||||
Very poor | Poor | Acceptable | Good | Excellent | |
Standard ultrasound | |||||
Liver | 0 (0) | 0 (0) | 5 (12.5) | 20 (50) | 15 (37.5) |
Gallbladder | 0 (0) | 0 (0) | 0 (0) | 18 (47.4)a | 20 (52.6)a |
Biliary tract | 0 (0) | 1 (2.5) | 12 (30) | 22 (55) | 5 (12.5) |
Right kidney | 0 (0) | 1 (2.5) | 3 (7.5) | 20 (50) | 16 (40) |
Left kidney | 1 (2.5) | 1 (2.5) | 4 (10) | 21 (52.5) | 13 (32.5) |
Pancreas | 2 (5) | 2 (5) | 8 (20) | 23 (57.5) | 5 (12.5) |
Spleen | 0 (0) | 0 (0) | 0 (0) | 12 (30) | 28 (70) |
Aorta | 0 (0) | 0 (0) | 3 (7.5) | 17 (42.5) | 20 (50) |
Urinary bladder | 0 (0) | 0 (0) | 2 (5) | 16 (40) | 22 (55) |
Prostate | 0 (0) | 0 (0) | 3 (7.5)a | 9 (22.5)a | 4 (10)a |
Uterus | 0 (0) | 1 (2.5)a | 6 (15)a | 9 (22.5)a | 5 (12.5)a |
Telerobotic ultrasound | |||||
Liver | 0 (0) | 0 (0) | 0 (0) | 9 (22.5) | 31 (77.5) |
Gallbladder | 0 (0) | 0 (0) | 0 (0) | 1 (2.6)a | 37 (97.4)a |
Biliary tract | 0 (0) | 0 (0) | 1 (2.5) | 13 (32.5) | 26 (65) |
Right kidney | 0 (0) | 0 (0) | 0 (0) | 6 (15) | 34 (85) |
Left kidney | 0 (0) | 0 (0) | 0 (0) | 7 (17.5) | 33 (82.5) |
Pancreas | 0 (0) | 0 (0) | 4 (10) | 14 (35) | 22 (55) |
Spleen | 0 (0) | 0 (0) | 0 (0) | 3 (7.5) | 37 (92.5) |
Aorta | 0 (0) | 0 (0) | 0 (0) | 3 (7.5) | 37 (92.5) |
Urinary bladder | 0 (0) | 0 (0) | 0 (0) | 2 (5) | 38 (95) |
Prostate | 0 (0) | 0 (0) | 0 (0) | 2 (12.5)a | 14 (87.5)a |
Uterus | 0 (0) | 0 (0) | 1 (4.8)a | 5 (23.8)a | 15 (37.5)a |
aPercentage of total valid cases (excluding cases of cholecystectomy, hysterectomy, or sexual incompatibility in the case of reproductive organs).
[figure(s) omitted; refer to PDF]
Table 3 shows the comparison of the visualization categories expressed as means of the overall score assigned per organ between both examinations. The mean overall score for visualization was significantly higher using standard ultrasound in each organ except for the gallbladder, with overall significant differences between groups (
Table 3
Comparison of visualization categories between telerobotic and standard ultrasound by anatomical structures.
Organ | Telerobotic ultrasound | Standard ultrasound | |
Liver | 3.25 (0.67) | 3.78 (0.42) | < 0.001 |
Gallbladder | 3.53 (0.51) | 3.63 (0.54) | 0.254 |
Biliary tract | 2.78 (0.70) | 3.63 (0.54) | < 0.001 |
Right kidney | 3.28 (0.72) | 3.85 (0.36) | < 0.001 |
Left kidney | 3.10 (0.87) | 3.83 (0.39) | < 0.001 |
Pancreas | 2.68 (0.94) | 3.45 (0.68) | < 0.001 |
Spleen | 3.70 (0.46) | 3.93 (0.27) | 0.002 |
Aorta | 3.43 (0.64) | 3.93 (0.27) | < 0.001 |
Urinary bladder | 3.50 (0.60) | 3.95 (0.22) | < 0.001 |
Prostate | 3.06 (0.68) | 3.88 (0.34) | < 0.001 |
Uterus | 2.86 (0.85) | 3.67 (0.58) | < 0.001 |
Overall | 3.19 (0.69) | 3.78 (0.42) | < 0.001 |
Note: Data are expressed as mean (
Table 4 shows the comparison between the “excellent” and “good or excellent” visualization categories between the two groups. For “excellent” visualization of organs, there were significant differences in all organs. However, for the combined category (good or excellent), significant differences were observed favoring standard ultrasound only for the biliary tract (
Table 4
Comparison of good or excellent visualization between telerobotic and standard ultrasound by anatomical structures.
Organ | Excellent visualization | Good or excellent visualization | ||||
Telerobotic ultrasound | Standard ultrasound | Telerobotic ultrasound | Standard ultrasound | |||
Liver | 15 (37.5) | 31 (77.5) | < 0.001 | 35 (87.5) | 40 (100) | NS |
Gallbladder | 20 (52.6)a | 37 (97.4)a | < 0.001 | 38 (100)a | 38 (100)a | NS |
Biliary tract | 5 (12.5) | 26 (65) | < 0.001 | 27 (67.5) | 39 (97.5) | < 0.001 |
Right kidney | 16 (40) | 34 (85) | < 0.001 | 36 (90) | 40 (100) | NS |
Left kidney | 13 (32.5) | 32 (80) | < 0.001 | 34 (85) | 40 (100) | NS |
Pancreas | 5 (12.5) | 22 (55) | < 0.001 | 28 (70) | 36 (90.0) | 0.008 |
Spleen | 28 (70) | 37 (92.5) | 0.004 | 40 (100) | 40 (100) | NS |
Aorta | 20 (50) | 37 (92.5) | < 0.001 | 37 (92.5) | 40 (100) | NS |
Urinary bladder | 22 (55) | 38 (95) | < 0.001 | 38 (93) | 40 (100) | NS |
Prostate | 4 (10)a | 14 (87.5)a | 0.002 | 13 (81.3) | 16 (100) | NS |
Uterus | 5 (12.5)a | 15 (37.5)a | 0.002 | 14 (66.7) | 20 (95.2) | 0.031 |
Abbreviation: NS, not significant.
aPercentage of total valid cases (excluding cases of cholecystectomy, hysterectomy, or sexual incompatibility in the case of reproductive organs).
Table 5 shows the findings identified in telerobotic ultrasound that were not observed in standard ultrasound and vice versa, grouped by organs. The total number of patients with findings identified on telerobotic ultrasound that were not detected on standard ultrasound was nine (22.5%), while the total number of patients with findings identified on standard ultrasound that were not detected on telerobotic ultrasound was 28 (70%). The corresponding figures for potentially relevant findings were one (2.5%) and 16 (40%).
Table 5
Findings observed in standard or telerobotic ultrasound that were not identified in the other modality.
Findings by organs | Findings in telerobotic ultrasound not detected on standard ultrasound | Findings in standard ultrasound not detected on telerobotic ultrasound |
Liver | ||
Simple cyst | 1 (2.5) | 3 (7.5) |
Granuloma | 2 (5) | 1 (2.5) |
Hemangioma | 0 | 1 (2.5) |
Others | — | — |
Total | 3 (7.5) | 5 (12.5) |
Gallbladder | 0 | 0 |
Lithiasis | 0 | 1 (2.5) |
Biliary mud | 0 | 2 (5) |
Polyp | 0 | 1a (2.5) |
Adenomyomatosis | 1 (2.5) | 0 |
Others | — | — |
Total | 1 (2.5) | 4 (10) |
Biliary tract | 0 | 0 |
Ectasia | 0 | 1 (2.5) |
Choledocholithiasis | 0 | 0 |
Others | — | — |
Total | 0 | 1 (2.5) |
Right kidney | 0 | 0 |
Simple cyst | 2 (5) | 4 (10) |
Complex cyst | 0 | 0 |
Nonobstructive lithiasis | 0 | 3 (7.5) |
Obstructive lithiasis | 0 | 0 |
Others | — | — |
Total | 2 (5) | 7 (17.5) |
Left kidney | 0 | 0 |
Simple cyst | 2 (5) | 4 (10) |
Complex cyst | 0 | 0 |
Nonobstructive lithiasis | 1 (2.5) | 2 (5) |
Obstructive lithiasis | 0 | 1 (2.5) |
Others | — | — |
Total | 3 (7.5) | 7 (17.5) |
Pancreas | 0 | 0 |
Cystic lesion | 0 | 1 (2.5) |
Wirsung duct ectasia | 0 | 1 (2.5) |
Others | — | — |
Total | 0 | 2 (5) |
Spleen | 0 | 0 |
Focal lesion | 0 | 0 |
Mild splenomegaly | 0 | 1 (2.5) |
Others | 0 | 0 |
Total | 0 | 1 (2.5) |
Aorta | 0 | 0 |
Dilation | 0 | 0 |
Atheromatosis | 0 | 1 (2.5) |
Others | 0 | 0 |
Total | 0 | 1 (2.5) |
Urinary bladder | 0 | 0 |
Outgrowth lesion | 0 | 0 |
Wall bladder thickening | 0 | 1 (2.5) |
Others | 0 | 0 |
Total | 0 | 1 (2.5) |
Uterus | 0 | 0 |
Leiomyoma | 0 | 0 |
Others | — | — |
Total | 0 | 0 |
Total number of patients with findings identified with respect to the other modality | 9 (22.5) | 28 (70) |
Total number of patients with potentially relevant findings identified with respect to the other modality | 1 (2.5) | 16 (40) |
Note: The findings in boldface are considered potentially relevant.
aPercentage with respect to the total number of patients.
The correlation analysis for the quantitative measurements (length) of the kidneys and spleen in both ultrasound modalities showed a statistically significant correlation between both ultrasound modalities (left kidney,
The results of the multivariate analysis for the dependent variable “absence of identification of a potentially relevant finding in telerobotic ultrasound,” adjusted for age, sex, BMI, and scan time, were not statistically significant (File S3).
Table 6 shows the results of the multivariate analysis using logistic regression for the variable “nonexcellent visualization of anatomical structures.” Telerobotic ultrasound was a risk factor for not excellent visualization in all organs, while increased BMI was a risk factor in the visualization of the liver (adjusted odds ratio (ORa) 1.23; 95% confidence interval (CI), 1.02–1.48), gallbladder (ORa, 1.63; 95% CI, 1.25–2.13), and spleen (ORa, 1.22; 95% CI, 1.00–1.49).
Table 6
Multivariate analysis (logistic regression) for nonexcellent visualization of anatomical structures.
DV: nonexcellent visualization | Telerobotic ultrasound (ref: standard ultrasound) | Age (years) | BMI (kg/m)2 | Male sex (ref: Female sex) | Scanning time (minutes) |
ORa (95% CI) | ORa (95% CI) | ORa (95% CI) | ORa (95% CI) | ORa (95% CI) | |
Liver | 12.13 (3.28–44.89) | 1.00 (0.96–1.05) | 1.23 (1.02–1.48) | 4.89 (1.27–18.74) | 1.02 (0.95–1.10) |
Gallbladder | 29.00 (5.07–166.03) | 1.02 (0.96–1.07) | 1.63 (1.25–2.13) | 0.49 (0.10–2.39) | 1.00 (0.91–1.09) |
Biliary tract | 11.47 (3.68–35.66) | 1.02 (0.98–1.06) | 1.08 (0.91–1.28) | 1.23 (0.35–4.28) | 0.99 (0.92–1.06) |
Right kidney | 9.39 (2.97–29.75) | 1.03 (0.99–1.07) | 1.11 (0.95–1.31) | 0.47 (0.13–1.65) | 1.01 (0.95–1.08) |
Left kidney | 9.98 (3.30–30.16) | 0.96 (0.93–1.00) | 1.11 (0.94–1.31) | 0.62 (0.18–2.15) | 1.01 (0.95–1.08) |
Pancreas | 8.69 (2.65–28.54) | 1.03 (0.99–1.08) | 1.10 (0.92–1.31) | 1.51 (0.42–5.41) | 0.99 (0.92–1.06) |
Spleen | 5.94 (1.36–25.69) | 1.02 (0.98–1.07) | 1.22 (1.00–1.49) | 0.75 (0.17–3.36) | 0.92 (0.84–1.00) |
Aorta | 16.81 (3.77–74.90) | 1.04 (0.99–1.09) | 1.15 (0.96–1.38) | 1.35 (0.35–5.19) | 0.95 (0.88–1.03) |
Urinary bladder | 18.29 (3.56–93.88) | 1.03 (0.99–1.08) | 1.02 (0.86–1.23) | 0.78 (0.19–3.12) | 0.92 (0.85–1.00) |
Prostate | 72.32 (3.10–1685.75) | 1.10 (0.99–1.21) | 1.03 (0.79–1.36) | — | 0.86 (0.72–1.04) |
Uterus | 7.70 (1.73–34.26) | 1.01 (0.96–1.06) | 1.25 (0.98–1.59) | — | 1.04 (0.95–1.14) |
Abbreviations: 95% CI, 95% confidence interval; BMI, body mass index; DV, dependent variable; ORa, adjusted odds ratio; Ref, reference group.
Finally, the results of the multivariate analysis for the need to perform a standard abdominal ultrasound after telerobotic ultrasound, adjusted for age, sex, BMI, and scan time, were not statistically significant (File S4).
4. Discussion
The feasibility of telerobotic ultrasonography has been demonstrated in the evaluation of obstetric ultrasonography [16, 17], with some studies reporting favorable results in thyroid [18], carotid [19], and cardiopulmonary [20] assessment. Its usefulness has also been evaluated specifically in abdominal ultrasonography, but with some limitations that undermine the external validity of the results obtained [3, 10, 21]. In addition, to our knowledge, in previous studies, ultrasound examinations were performed by sonographers but not by abdominal radiology specialists.
Our results show that, despite the longer time required for each ultrasound scan and the lower sensitivity for the overall detection of findings, telerobotic ultrasound can be useful for the assessment of abdominopelvic pathologies in different organs with acceptable results, in line with previous studies [3, 10, 19, 21]. Notably, although the radiologists followed a standardized protocol for the assessment of all abdominal organs, we found that the visualization quality of telerobotic ultrasound varies depending on the organ being assessed, with similar results to standard ultrasound in the gallbladder, aorta, urinary bladder, and spleen and slightly inferior—although acceptable—outcomes in the liver and kidneys. The latter organs were of notable importance in our sample, taking into account that more than 60% of the clinical indications for ultrasound were related to suspected renal lithiasis or hepatic steatosis.
The greatest limitations were found in the assessment of the biliary tract, pancreas, and uterus. These findings are slightly different from those reported by Adams et al., who found that the organs that were worst visualized on telerobotic ultrasound were the pancreas, aorta, spleen, gallbladder, and left kidney [10]. Surprisingly, in the said study, the bile duct and liver were visualized in a comparable manner between the two examinations. This could be explained by methodological differences in the design and patient sample between both studies. Overall, we hypothesize that the poorer visualization of abdominal organs in telerobotic ultrasound was mainly due to intrinsic operator–dependent factors (i.e., the need to synchronize the technician with the radiologist’s instructions) and equipment-related factors (limited pressure application, which cannot be sensed by the radiologist, and inability to perform translational displacements on the patient’s surface without the technician).
The mean duration of telerobotic ultrasound examinations was similar to that described in some previous studies [10], although other studies have reported shorter scanning times [19]. One study by Ren et al. in China using a telerobotic ultrasound system with 5G technology reported a mean duration of 7–11 min and concluded that the telerobotic system employed was useful and feasible for rural population health screenings [22]. Scan times shorter than ours (i.e.,
Interestingly, our multivariate analysis to predict the variable “nonexcellent visualization of anatomical structures” showed that telerobotic ultrasonography was a risk factor for all organs and high BMI was a risk factor in the visualization of liver, gallbladder, and spleen. In our opinion, the importance of BMI has not been explored with sufficient attention in previous studies on telerobotic ultrasound, and it is an important factor to consider, particularly in countries with high obesity rates, as it has been shown to worsen the visualization of different organs and lesions [23]. In fact, one of the most important technical limitations of telerobotic ultrasound derives from the inability to control the degree of pressure and transducer displacement [3, 10]. These limitations may be particularly important in the case of overweight or obese patients, where it is necessary to exert a greater degree of pressure and to use several viewing windows.
Overall, our results highlight the need to ensure the professional competence of radiologists in remote secondary hospitals when implementing telerobotic ultrasound. In our study, all remote ultrasound examinations were performed by abdominal radiologists, which ensured high standards of image interpretation and diagnostic accuracy. However, in routine practice, ensuring consistent professional quality in remote settings may require different strategies, such as specific training and continuous education—including novel educational approaches for resource-limited settings [24]—or the development of specific telemedicine protocols similar to the telestroke guidelines [25]. These measures would help mitigate any potential variability in diagnostic performance due to geographic location or resource availability. Of note, the advantages of ultrasound compared to other imaging modalities such as CT or MRI (e.g., lack of ionizing radiation, low cost, and ability to detect specific conditions such as noncalcified cholelithiasis [26, 27]) enhance the need for improving efforts to preserve the use of this imaging modality in resource-limited setting such as second-level hospitals.
The main limitations of the study include the relatively low sample size, which probably affected the multivariate analyses, and the use of ad hoc questionnaires and scores. In addition, we did not address potential confounders such as patient comorbidities or operator experience. Of note, all radiologists involved in the study had significant expertise and similar experience in abdominal ultrasound. The main strengths are its prospective nature, the blinding, the application of the same ultrasound equipment, and the inclusion of abdominal radiologists as ultrasound performers.
Future prospective studies with larger numbers of patients are needed, including a broader representation of patients’ physical characteristics (e.g., BMI) and clinical indications. Particularly important is the evaluation of the usefulness of telerobotic ultrasound in urgent abdominopelvic pathologies, since it could solve the need for on-site radiologists in rural areas and second-level hospitals distant from referral hospitals. In addition, the cost-efficiency of the equipment needs to be evaluated in a regulated manner. Although some studies have found telerobotic ultrasound to be cost-efficient in rural and remote areas where access to specialized medical care is limited [4], considering the current radiological quality standards, further studies should be performed to demonstrate the economical benefits of this technology in nonrural areas.
5. Conclusions
Telerobotic ultrasound allows an acceptable assessment of most abdominopelvic organs, but there are significant limitations compared to standard ultrasound for their optimal evaluation, especially in the pancreas, biliary tract, and uterus. Although this ultrasound modality can help provide adequate healthcare to patients in locations with limited access to radiology specialists, its longer duration and limited visualization of organs should be overcome to justify its widespread use in other medical environments such as second-level hospitals.
Funding
Funding for the open access charge was provided by the Universidad de Granada/CBUA.
[1] Z. Izadifar, P. Babyn, D. Chapman, "Mechanical and biological effects of ultrasound: a review of present knowledge," Ultrasound in Medicine & Biology, vol. 43 no. 6, pp. 1085-1104, DOI: 10.1016/j.ultrasmedbio.2017.01.023, 2017.
[2] Z. Jiang, S. E. Salcudean, N. Navab, "Robotic ultrasound imaging: state-of-the-art and future perspectives," Medical Image Analysis, vol. 89, article 102878,DOI: 10.1016/j.media.2023.102878, 2023.
[3] P. Arbeille, A. Capri, J. Ayoub, V. Kieffer, M. Georgescu, G. Poisson, "Use of a robotic arm to perform remote abdominal telesonography," AJR. American Journal of Roentgenology, vol. 188 no. 4, pp. W317-W322, DOI: 10.2214/AJR.05.0469, 2007.
[4] S. J. Adams, E. Penz, B. Imeah, B. Burbridge, H. Obaid, P. Babyn, I. Mendez, "Economic evaluation of telerobotic ultrasound technology to remotely provide ultrasound services in rural and remote communities," Journal of Ultrasound in Medicine, vol. 42 no. 1, pp. 109-123, DOI: 10.1002/jum.16070, 2023.
[5] P. Vieyres, G. Poisson, F. Courreges, O. Merigeaux, P. Arbeille, "The TERESA project: from space research to ground tele-echography," Industrial Robot: An International Journal, vol. 30 no. 1, pp. 77-82, DOI: 10.1108/01439910310457742, 2003.
[6] W. Jiang, X. Zhao, T. Gan, Y. Liu, S. Luo, M. Zhu, S. Chen, Y. Jiang, M. Wu, "Application of a tele-ultrasound robot during COVID-19 pandemic: a feasibility study," Journal of Ultrasound in Medicine, vol. 42 no. 3, pp. 595-601, DOI: 10.1002/jum.16041, 2023.
[7] S. J. Adams, B. Burbridge, H. Obaid, G. Stoneham, P. Babyn, I. Mendez, "Telerobotic sonography for remote diagnostic imaging: narrative review of current developments and clinical applications," Journal of Ultrasound in Medicine, vol. 40 no. 7, pp. 1287-1306, DOI: 10.1002/jum.15525, 2021.
[8] T. He, Y.-Y. Pu, Y.-Q. Zhang, Z.-B. Qian, L.-H. Guo, L.-P. Sun, C.-K. Zhao, H.-X. Xu, "5G-based telerobotic ultrasound system improves access to breast examination in rural and remote areas: a prospective and two-scenario study," Diagnostics, vol. 13 no. 3,DOI: 10.3390/diagnostics13030362, 2023.
[9] L. Wright, M. Florez, M. Isaksson, G. Lambert, T. Marwick, "Current applications of remote robot-assisted cardiac ultrasound examination: a systematic review of clinical use and technological limitations," Heart, Lung & Circulation, vol. 31, pp. S155-S156, DOI: 10.1016/j.hlc.2022.06.241, 2022.
[10] S. J. Adams, B. E. Burbridge, A. Badea, L. Langford, V. Vergara, R. Bryce, L. Bustamante, I. M. Mendez, P. S. Babyn, "Initial experience using a telerobotic ultrasound system for adult abdominal sonography," Canadian Association of Radiologists Journal, vol. 68 no. 3, pp. 308-314, DOI: 10.1016/j.carj.2016.08.002, 2017.
[11] World Health Organization, Training in diagnostic ultrasound: essentials, principles, and standards, 1998.
[12] M. Tele, "Medirob," . Medirob Website http://www.medirob.com/en/ (accessed June 12, 2024)
[13] U.S. Food and Drug Administration, "510K Pre-market notification," 2017. MELODY, Remote Control System for Ultrasound Probe. https://urldefense.com/v3/__https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=K161354__;!!N11eV2iwtfs!oaSka_d0gqb3dOMAgcTJ92fWu7eBFQEdjCDKBl1dhleYhjwy0u-6BXd0vvXDNWihB5c8e0SIaPW9Pwz0e3s$ Consulted: 12 January 2024
[14] AdEchoTech, "MELODY – Telerobotic ultrasound solution," 2024. https://urldefense.com/v3/__https://www.adechotech.com/products/__;!!N11eV2iwtfs!oaSka_d0gqb3dOMAgcTJ92fWu7eBFQEdjCDKBl1dhleYhjwy0u-6BXd0vvXDNWihB5c8e0SIaPW92gLm-4g$ . Access date: 12 January 2024
[15] "AdEchoTech. Informations sur la sécurité et les performances des dispositifs de la gamme MELODY-Safety and performance information (English version)," . https://urldefense.com/v3/__https://www.adechotech.com/wp-content/uploads/2016/03/Informations-sur-la-s%2A%2AAcurit%2A%2AA-et-les-performances-des-dispositifs-de-la-gamme-MELODY-2.pdf__;w6nDqQ!!N11eV2iwtfs!oaSka_d0gqb3dOMAgcTJ92fWu7eBFQEdjCDKBl1dhleYhjwy0u-6BXd0vvXDNWihB5c8e0SIaPW94AN1oIA$ Access date: 12 January 2024
[16] S. J. Adams, B. E. Burbridge, A. Badea, N. Kanigan, L. Bustamante, P. Babyn, I. Mendez, "A crossover comparison of standard and telerobotic approaches to prenatal sonography," Journal of Ultrasound in Medicine, vol. 37 no. 11, pp. 2603-2612, DOI: 10.1002/jum.14619, 2018.
[17] S. J. Adams, B. Burbridge, L. Chatterson, V. McKinney, P. Babyn, I. Mendez, "Telerobotic ultrasound to provide obstetrical ultrasound services remotely during the COVID-19 pandemic," Journal of Telemedicine and Telecare, vol. 28 no. 8, pp. 568-576, DOI: 10.1177/1357633X20965422, 2022.
[18] Y.-Q. Zhang, H.-H. Yin, T. He, L.-H. Guo, C.-K. Zhao, H.-X. Xu, "Clinical application of a 5G-based telerobotic ultrasound system for thyroid examination on a rural island: a prospective study," Endocrine, vol. 76 no. 3, pp. 620-634, DOI: 10.1007/s12020-022-03011-0, 2022.
[19] M. Georgescu, A. Sacccomandi, B. Baudron, P. L. Arbeille, "Remote sonography in routine clinical practice between two isolated medical centers and the university hospital using a robotic arm: a 1-year study," Telemedicine Journal and E-Health, vol. 22 no. 4, pp. 276-281, DOI: 10.1089/tmj.2015.0100, 2016.
[20] R. Ye, X. Zhou, F. Shao, L. Xiong, J. Hong, H. Huang, W. Tong, J. Wang, S. Chen, A. Cui, C. Peng, Y. Zhao, L. Chen, "Feasibility of a 5G-based robot-assisted remote ultrasound system for cardiopulmonary assessment of patients with coronavirus disease 2019," Chest, vol. 159 no. 1, pp. 270-281, DOI: 10.1016/j.chest.2020.06.068, 2021.
[21] H.-H. Chai, R.-Z. Ye, L.-F. Xiong, Z.-N. Xu, X. Chen, L.-J. Xu, X. Hu, L.-F. Jiang, C.-Z. Peng, "Successful use of a 5G-based robot-assisted remote ultrasound system in a care center for disabled patients in rural China," Frontiers in Public Health, vol. 10, article 915071,DOI: 10.3389/fpubh.2022.915071, 2022.
[22] J.-Y. Ren, Y.-M. Lei, B.-S. Lei, Y.-X. Peng, X.-F. Pan, H.-R. Ye, X.-W. Cui, "The feasibility and satisfaction study of 5G-based robotic teleultrasound diagnostic system in health check-ups," Frontiers in Public Health, vol. 11, article 1149964,DOI: 10.3389/fpubh.2023.1149964, 2023.
[23] L. Kiri, M. Abdolell, A. F. Costa, V. Keough, J. Rowe, R. Butt, S. E. Clarke, "US LI-RADS visualization score: interobserver variability and association with cause of liver disease, sex, and body mass index," Canadian Association of Radiologists Journal, vol. 73 no. 1, pp. 68-74, DOI: 10.1177/08465371211012104, 2022.
[24] B. Terry, D. L. Polan, R. Nambaziira, J. Mugisha, M. Bisanzo, R. Gaspari, "Rapid, remote education for point-of-care ultrasound among non-physician emergency care providers in a resource limited setting," African Journal of Emergency Medicine, vol. 9 no. 3, pp. 140-144, DOI: 10.1016/j.afjem.2019.05.004, 2019.
[25] B. M. Demaerschalk, J. Berg, B. W. Chong, H. Gross, K. Nystrom, O. Adeoye, L. Schwamm, L. Wechsler, S. Whitchurch, "American telemedicine association: telestroke guidelines," Telemedicine Journal and E-Health, vol. 23 no. 5, pp. 376-389, DOI: 10.1089/tmj.2017.0006, 2017.
[26] P. Cianci, E. Restini, "Management of cholelithiasis with choledocholithiasis: endoscopic and surgical approaches," World Journal of Gastroenterology, vol. 27 no. 28, pp. 4536-4554, DOI: 10.3748/wjg.v27.i28.4536, 2021.
[27] C. Fung, R. Spychka, R. Noorelahi, K. Vijay, D. T. Fetzer, "Ultrasound of the gallbladder: not the same bag of tricks," Abdominal Radiology,DOI: 10.1007/s00261-024-04530-2, 2024.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Copyright © 2024 David Luengo Gómez et al. This is an open access article distributed under the Creative Commons Attribution License (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License. https://creativecommons.org/licenses/by/4.0/
Details





1 Abdominal Radiology Unit Department of Radiology Hospital Universitario Virgen de las Nieves 18014 Granada Spain; Advanced Medical Imaging Group (TeCe-22) Instituto de Investigación Biosanitaria de Granada (ibs.GRANADA) 18012 Granada Spain
2 Abdominal Radiology Unit Department of Radiology Hospital Universitario Virgen de las Nieves 18014 Granada Spain
3 Abdominal Radiology Unit Department of Radiology Hospital Universitario Virgen de las Nieves 18014 Granada Spain; Advanced Medical Imaging Group (TeCe-22) Instituto de Investigación Biosanitaria de Granada (ibs.GRANADA) 18012 Granada Spain; Department of Human Anatomy and Embryology School of Medicine University of Granada 18071 Granada Spain