Abstract
Background
Experimental knee implant wear testing according to ISO 14243 is a standard procedure, but it inherently possesses limitations for preclinical evaluations due to extended testing periods and costly infrastructure. In an effort to overcome these limitations, we hereby develop and experimentally validate a finite-element (FE)-based algorithm, including a novel cross-shear and contact pressure dependent wear and creep model, and apply it towards understanding the sensitivity of wear outcomes to the applied boundary conditions.
Methods
Specifically, we investigated the application of in vivo data for level walking from the publicly available “Stan” data set, which contains single representative tibiofemoral loads and kinematics derived from in vivo measurements of six subjects, and compared wear outcomes against those obtained using the ISO standard boundary conditions. To provide validation of the numerical models, this comparison was reproduced experimentally on a six-station knee wear simulator over 5 million cycles, testing the same implant Stan’s data was obtained from.
Results
Experimental implementation of Stan’s boundary conditions in displacement control resulted in approximately three times higher wear rates (4.4 vs. 1.6 mm3 per million cycles) and a more anterior wear pattern compared to the ISO standard in force control. While a force-controlled ISO FE model was unable to reproduce the bench test kinematics, and thus wear rate, due to a necessarily simplified representation of the simulator machine, similar but displacement-controlled FE models accurately predicted the laboratory wear tests for both ISO and Stan boundary conditions. The credibility of the in silico wear and creep model was further established per the ASME V&V-40 standard.
Conclusions
The FE wear model is suitable for supporting future patient-specific models and development of novel implant designs. Incorporating the Stan data set alongside ISO boundary conditions emphasized the value of using measured kinematics in displacement control for reliably replicating in vivo joint mechanics in wear simulation. Future work should focus on expanding the range of daily activities simulated and addressing model sensitivity to contact mechanics to further enhance predictive accuracy.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer