Abstract

Purpose

This work describes a single institution experience of commissioning a real‐time target tracking and beam control system, known as comprehensive motion management, for a 1.5 T Elekta MR‐Linac.

Methods

Anatomical tracking and radiation beam control were tested using the MRI4D Quasar motion phantom. Multiple respiratory breathing traces were modeled across a range of realistic regular and irregular breathing patterns ranging between 10 and 18 breaths per minute. Each of the breathing traces was used to characterize the anatomical position monitoring (APM) accuracy, and beam latency, and to quantify the dosimetric impact of both parameters during a respiratory‐gated delivery using EBT3 film dosimetry. Additional commissioning tasks were performed to verify the dosimetric constancy during beam gating and to expand our existing quality assurance program.

Results

It was determined that APM correctly predicted the 3D position of a dynamically moving tracking target to within 1.5 mm for 95% of the imaging frames with no deviation exceeding 2 mm. Among the breathing traces investigated, the mean latency ranged between −21.7 and 7.9 ms with 95% of all observed latencies within 188.3 ms. No discernable differences were observed in the relative profiles or cumulative output for a gated beam relative to an ungated beam with minimal dosimetric impact observed due to system latency. Measured dose profiles for all gated scenarios retained a gamma pass rate of 97% or higher for a 3%/2 mm criteria relative to a theoretical gated dose profile without latency or tracking inaccuracies.

Conclusion

MRI‐guided target tracking and automated beam delivery control were successfully commissioned for the Elekta Unity MR‐Linac. These gating features were shown to be highly accurate with an effectively small beam latency for a range of regular and irregular respiratory breathing traces.

Details

Title
Commissioning of a motion management system for a 1.5T Elekta Unity MR‐Linac: A single institution experience
Author
Smith, Blake R 1   VIAFID ORCID Logo  ; St‐Aubin, Joel 1   VIAFID ORCID Logo  ; Hyer, Daniel E. 1 

 Department of Radiation Oncology, University of Iowa, Iowa City, Iowa, USA 
Section
RADIATION ONCOLOGY PHYSICS
Publication year
2025
Publication date
Apr 1, 2025
Publisher
John Wiley & Sons, Inc.
e-ISSN
15269914
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
3186181710
Copyright
© 2025. This work is published under http://creativecommons.org/licenses/by/4.0/ (the "License"). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.