Abstract

This work proposes a signal processing algorithm to analyse the optical signal from a Low Coherence Interferometric (LCI) system. The system uses a Mach-Zehnder (MZ) interferometer to interrogate a Fabry-Perot cavity, working as an optical sensor. This algorithm is based on the correlation and convolution operations, which allows the signal to be reconstructed based on itself, as well as, on the linearization of the signal phase, allowing the non-linearities of the actuator incorporated on the MZ interferometer to be compensated. The results show a noise reduction of 30 dB in the signal acquired. As a result, a reduction of 8.2 dB in the uncertainty of the measurement of the physical measurand is achieved. It is also demonstrated that the phase linearization made it possible to obtain a coefficient of determination (namely, R-squared) higher than 0.999.

Details

Title
Low Coherence Interferometry Measurement: An Algorithm for fast processing with low noise and phase linearisation
Author
Robalinho, Paulo; Rodrigues, A; Novais, S; Lobo Ribeiro, A B; Silva, S; Frazão, O
Section
Topical Meeting (TOM) 10- Applications of Optics and Photonics
Publication year
2024
Publication date
2024
Publisher
EDP Sciences
ISSN
21016275
e-ISSN
2100014X
Source type
Conference Paper
Language of publication
English
ProQuest document ID
3193680763
Copyright
© 2024. This work is licensed under https://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and conditions, you may use this content in accordance with the terms of the License.