Abstract
Background
The dynamic regional accuracy of electrical impedance tomography has not yet been validated. We aimed to compare the regional accuracy of electrical impedance tomography with that of four-dimensional computed tomography during dynamic ventilation.
Methods
This single-center, prospective, observational study conducted in a general intensive care unit included adult patients receiving mechanical ventilation from July 2021 to February 2024. The patients were mechanically ventilated passively and underwent electrical impedance tomography and four-dimensional computed tomography on the same day.
Results
Overall, 45 patients were analyzed. The correlation coefficients in regional dynamic ventilation between four-dimensional computed tomography and electrical impedance tomography in each region were 0.963, 0.963, 0.835 (ventral, central, and dorsal, respectively) in the right lung and 0.947, 0.927, 0.823 (ventral, central, and dorsal, respectively) in the left lung. The correlation coefficient was low when the regional ventilation distribution detected by the electrical impedance tomography was < 2%. After excluding nine patients with a regional ventilation distribution of < 2%, the ventral, central, and dorsal correlation coefficients were 0.963, 0.963, and 0.946 in the right lung and 0.942, 0.924, and 0.951, respectively, in the left lung.
Conclusions
Regional ventilation using electrical impedance tomography during dynamic ventilation was highly accurate and consistent with the time phase compared to four-dimensional computed tomography. Given the high correlation between these modalities, they can contribute significantly to further studies on regional ventilation dynamics.
Trial registration number ClinicalTrials.gov (No. UMIN00044386).
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer