Abstract
The aim of the study was to compare the heart rate, QRS interval, and R wave amplitude across three electrocardiogram models, and assess the ability of each of them to provide electrocardiograms (ECG) for clinical interpretation. The three electrocardiogram models included ECG Seiva Praktik Veterinary, CardioStore ECG and AliveCor Veterinary Heart Monitor. The data were collected from twelve healthy adult captive green iguanas (Iguana iguana) monitored under a manual restraint at a room temperature of 22.6–28.0 °C. The ECGs using the Seiva Praktik and CardioStore ECG veterinary electrocardiography were performed with standard 4 lead ECG recordings. The AliveCor Veterinary Heart Monitor was placed (with the use of gel) directly on the lateral body wall. The mean heart rate was 42 ± 8 beats/min (CardioStore), 50 ± 11 beats/min (Seiva Praktik Veterinary), and 51 ± 9 beats/min (AliveCor Veterinary Heart Monitor). No significant difference in the heart rate was observed. A significant difference (P < 0.05) in the QRS duration was observed between the CardioStore and AliveCor Veterinary Heart Monitor. Significant differences (P < 0.01) in the R wave amplitude were detected between the CardioStore and AliveCor Veterinary Heart Monitor and between the Seiva Praktik Veterinary and AliveCor Veterinary Heart Monitor. The ECGs produced by the Seiva Praktik VVeeterinary and CardioStore machines were interpretable at 100%, while those produced by the AliveCor Veterinary Heart Monitor were interpretable at 66%. Seiva Praktik Veterinary is most appropriately used as an anaesthesia monitoring tool. AliveCor Veterinary Heart Monitor could be used as an additional diagnostic tool, but the results should be ideally confirmed with a standard ECG machine. Seiva Praktik Veterinary is the most appropriate tool for monitoring the ECG within the anaesthesia, while CardioStore might be most appropriately used as an advanced diagnostic tool by virtue of its software assistance. The ECGs obtained with AliveCor Veterinary Heart Monitor should be confirmed using a standard ECG machine.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer