ABSTRACT
The present paper describes the application of Fast-gas chromatography (GC) to quantify the capsalcinoid content in dried pepper fruit with the aim to calculate the Scoville heat values (SHVs) for routine quality control. Fast-GC was used to analyse various samples of dried pepper fruits with a wide distribution of SHVs. The calibration curves demonstrated a polynomial model in the range of 3.0-195.0 µg mL^sup -1^ for capsaicin and 2.6-105.0 µg mL^sup -1^ for dihydrocapsalcin and nordihydrocapsalcin, with correlation coefficients of 0.996 and 0.997, respectively. The data showed an intraday relative standard deviation (RSD) <5% and an interday relative repeatability standard deviation (RSDr) <6%.
Key words: Capsicum, spice, fast-gas chromatography
INTRODUCTION
Pepper fruits [Capsicum spp.) are among the most commonly consumed spices having significant pungency. They contain capsaicinoids, derivatives of phenylpropanoid compounds, which are largely responsible for the hot taste, pungency, and trade quality. Eleven capsaicinoids have been identified, including amide derivatives of vanillylamlne and carboxylic acids with eight to eleven carbon atoms, some unsaturated (GARCÉS-CLAVER et al, 2006; KOZUKUE et al, 2005). The capsalcinoid content can be inferred by the heat value based on the Scoville organoleptic test, which was developed by the American Spice Trade Association (ASTA). However, the Scoville Heat Value (SHV) is not always reliable and various authors have attempted to replace this method with more accurate instrumental measurements of the capsalcinoid content. When evaluating pungency, capsaicin (trcms-8-methyl-N vanillyl-6-nonenamide) and dihydrocapsaicin (8-methyl-N-vanillylnonanamide) constitute 90% or more of the total capsaicinoids, and they account for the majority of the "hotness" (LASKARIDOU-MONNERVILLE, 1999). Nevertheless, the AOAC Officiai Method 995.03 (AOAC, 1995) recommends calculating the total capsalcinoid content as the sum of capsaicin, dihydrocapsaicin, and nordihydrocapsaicin; then, the Scoville Heat Unit (SHU) for a given compound is determined by multiplying the total content of these capsaicinoids (expressed as µg g1 capsicum) by the pungency value for "pure capsaicin" (15 SHU). The SHV of the product is equal to the sum of the single SHU values. TODD et al (1977) showed the capsalcinoid content could be converted to a more accurate SHU value by multiplying the content of each capsaicinoid by a coefficient: 16. 1 for both capsaicin and dihydrocapsaicin, 9.3 for nordihydrocapsaicin, and lower coefficients for other capsaicinoids. The lowest coefficient was 6.9 for homocapsaicin.
Several varieties of Capsicum (Solanaceae family) have been distinguished and described; five are commonly recognised as domesticated: Capsicum annuum L. var. Annuum, Capsicum baccatum L. var. Pendulum, Capsicum pubescens, Capsicum frutescens L. (Cayenne pepper), and Capsicum chínense (PETER, 2001). Many different techniques are used to extract capsaicinoids from dried plant matrices, such as maceration (KIRSCHBAUM-TITZE et al., 2002), magnetic stirring (CONTRERAS-PADILLA and YAHIA, 1998), enzymatic extraction (SANTAMARÍA et al., 2000), a solvent and shake-flask extraction (Soxhlet) (KOREL et al., 2002; KRAJEWSKA and POWERS, 1987), extraction by supercritical fluids (DAOOD et al, 2002), extraction by pressurized liquids (BARBERO et al, 2006a), and ultrasound assisted or microwave assisted extraction (BARBERO et al., 2006a BARBERO et al., 2008a; KARNKA et al., 2002; NAZARI et al., 2007; WILLIAMS et al., 2004). In general, there is always demand for simple extraction techniques, reduced organic solvent consumption, and low toxicity solvents. In particular, the AOAC Officiai Method 995.03 (AOAC, 1995) suggests extraction with warm ethanol in a reflux condenser, liquid chromatography, and fluorimetrie quantification. COLLINS et al. (1995) also suggested extraction with acetonitrile at reflux and high pressure liquid chromatography (HPLC) for quantification, but BARBERO et al (2006b) studied the extraction of capsaicinoids with various solvents and found acetonitrile is a fairly efficacious solvent, but less so than ethanol and methanol.
In recent years, reverse phase HPLC was the most frequently used method for quantification of capsaicinoids (BARBERO et al., 2008b; CHOI et al., 2006; KAALE et al., 2002; MAILLARD et al., 1997; PERUCKA and OLESZEK, 2000; SARJA et al., 1981; THOMPSON et al., 2005; WEAVER and AWDE, 1986). HPLC followed by mass spectrometry (HPLC/MS) was proposed more recently (GARCÉS-CLAVER et al., 2006; KOZUKUE et al., 2005). However, some authors have suggested the use of gas chromatography (GC) followed by mass spectrometry (MS) (HAWER et al., 1994; KRAJEWSHA and POWERS, 1987; KRAJEWSHA and POWERS, 1988; MANIRAKIZA et al , 1999; THOMAS et al., 1998) and a recent report described coupled GC/MS with solid phase microextraction (SPME) (PEÑA-ALVAREZ et al, 2009).
This paper presents an alternative and simple method to detect and quantify capsaicin, dihydrocapsaicin, and nordihydrocapsalcin with a Fast-GC system, which is useful for routine analysis. This method has not previously been proposed for the instrumental quantification of capsaicinoids. The AOAC Official Method 995.03 was used as the extraction method. GC/MS was adopted to confirm the identity of capsaicinoids in the start-up phase. The Fast-GC application demonstrated good peak resolution and the run time was comparable with that of various conventional and rapid HPLC techniques and lower than that corresponding to other GC conditions.
MATERIALS AND METHODS
Samples
Ten samples of dried ground red pepper from different origins were collected from import operators of the Italian market.
Reagents
Analytical-grade ethyl alcohol (99.8%) was purchased from Fluka Chemlka (Sigma Aldrich-Milan, Italy). Capsaicin (>97%) and dihydrocapsaicin (90%) were purchased from Sigma Aldrich (Milan, Italy). Stock solutions of these capsaicinoids were prepared by weighing and dissolving each compound in ethyl alcohol. These solutions were stored at 4°C and used for preparing diluted working standard solutions in ethyl alcohol.
Instruments and programmes
GC/MS. GC/MS analyses were carried out with a Shimadzu 2010 gas Chromatograph coupled to a Shimadzu 2010 MSD quadrupole mass spectrometer (Shimadzu, Italy). Helium was used as the carrier gas at a flow rate of 1 mL min1. Suitable identification of the analytes was achieved using an Equity(TM)-5, a poly (5% diphenyl - 95% dimethylsiloxane) stationary phase column, 30 m × 0.25 mm i.d., 0.25 µm film thickness (Supelco-Milan, Italy). The oven temperature programme was 60°C (held for 1 min) to 240°C at a rate of 3°C min^sup -1^ (held for 5 min). The injector temperature was 230°C and the split injector mode (1:5) was used. The ion source and transfer line temperatures were 200° and 250°C, respectively. The MS operating conditions were: positive electron ionization mode (EI+) with automatic gain control, 70 eV of electron energy, and 250 mA of emission current. The mass spectrometer was operated in full scan mode from m/z 40 to 340. Capsaicin, dihydrocapsalcin, and nordihydrocapsalcin were identified by comparing the retention times and spectra of each capsalcinoid with those of standard compounds. The spectral data were obtained from the Di. Pro. Ve.-UNIMI homemade library.
Fast-GC. Fast-GC analyses were performed on a Shimadzu system composed of a GC 2010 gas Chromatograph (Shimadzu Italia, Milan, Italy) equipped with a split/ splitless injector, an autosampler AOC-20i, and a FID detector. Data acquisition was performed with GC Solution Software (Shimadzu Italia, Milan, Italy). Analyses were performed with a RTX®-5 (Crossbond® 5% diphenyl- 95% dimethyl polysiloxane) stationary phase column (Restek, Bellefonte, PA, USA). The column dimensions were 10 m × 0.10 mm i.d., 0.10 µm film thickness. The oven temperature programme was 180°C (held for 2 min) increased to 240°C at a rate of 8°C min^sup -1^, then increased to 320°C at a rate of 14°C min^sup -1^ (held for 5 min). The pressure programme (336 kPa) was a constant linear velocity. The velocity of the carrier gas (H^sub 2^) was 78.5 cm sec^sup -1^. The injector temperature was 280°C and the split mode (1:100) was used. The injection volume was 1 µL. The detector parameters (FID) were: T = 340°C; H^sub 2^, 40 mL min^sup -1^; Air, 400 mL min^sup -1^; Makeup (N^sub 2^), 30 mL min^sup -1^; Base period, 8 ms; Filter time constant, 50 ms; Sampling rate, 8 ms.
Extraction of capsalcinoids
Data were collected to determine the usefulness of the instrumental response of Fast-GC for quantitative evaluation of capsaicinoids to calculate the SHVs for routine quality control. To prepare the extracts for Fast-GC evaluation, we used the AOAC Official Method 995.03 (1995). The filtered extract, produced by reflux of 25 g of ground or crushed red pepper with 200 mL of ethanol, was diluted 1:2 in ethyl alcohol prior to injection in Fast-GC.
Calibration and precision data
Once the Fast-GC procedure was optimised, calibration curves were created with standard solutions of the two standards of capsaicinoids commercially available, capsaicin and dihydrocapsalcin. Standard solutions were prepared in ethyl alcohol to cover concentrations up to 195 µg mL^sup -1^for capsaicin (i.e.: 3.0, 6.1, 9.8, 24.4, 48.8, 97.5, 146.2, and 195.0 µg mL^sup -1^) and up to 105 µg mL^sup -1^ for dihydrocapsalcin (i.e.: 2.6, 5.2, 13.1, 26.2, 52.5, 78.8, and 105.0 µg mL^sup -1^). These concentration ranges were previously defined to be useful to calculate the content as mg kg^sup -1^ of considered capsaicinoids in dried matrices studied by the concentration data of the alcoholic extract. The previous ranges cover the possible concentration of the extracts deriving from natural matrices: the factor multiplicative 16, deriving from the extraction method parameters, permits to express the capsalcinoid content in the matrix as µg g^sup -1^ from the measured concentration µg mL^sup -1^ in the extract. Regression equations and correlation coefficients were calculated using Microsoft Office Excel 2007 and are reported in Figure 1. The standard distribution data demonstrated a polynomial model for both capsaicin and dihydrocapsalcin in the range 3.0 - 195.0 mg mL^sup -1^ for capsaicin and 2.6 - 105.0 mg mL^sup -1^ for dihydrocapsalcin in the extracts analysed by Fast-GC. The data were fit with an R^sup 2^ = 0.996 for capsaicin and R^sup 2^ = 0.997 for dihydrocapsalcin and nordihydrocapsaicin. The calibration curve of dihydrocapsaicin was also used for evaluating nordihydrocapsalcin, as conventionally adopted by others (BARBERO et al, 2006b; ZIINO et al, 2009).
The limit of detection (LOD) and limit of quantitation (LOQ) values were respectively: 1.2 and 3.5 µg mL^sup -1^ for capsaicin and 1.0 and 3.0 mg mL^sup -1^ for dihydrocapsalcin and nordihydrocapsaicin. The repeatability and reproducibility were studied in respect to the concentration of each capsalcinoid. The average intraday relative standard deviations (RS D%), calculated on five samples shown in Table 1, were 3.9, 4.2, and 4.3 for capsaicin, dihydrocapsalcin, and nordihydrocapsaicin, respectively. The average interday relative repeatability standard deviations (RSDr%), calculated on the ten samples shown in Table 1, were 5.4, 5.6, and 5.8 for the same three capsaicinoids. These data were determined with the NMKL procedure n.5 (NORDIC COMMITTEE ON FOOD ANALYSIS, 1 997) for the estimation and expression of measurement uncertainty in chemical analysis.
Recovery data were estimated by spiking two samples (n. 4 and n. 5 in Table 1) having very different Scoville Heat Value, adding a calculated concentration of 500 mg kg^sup -1^ for capsaicin and 200 mg kg^sup -1^ for dihydrocapsaicin. The resulting recoveries were 98.6 and 97.7% respectively for capsaicin and dihydrocapsaicin for sample n.4 (mean of two determinations); the corresponding values for sample n.5 result 95,0 and 94,2%.
Statistical data performed by other GC methods cited in bibliography don't show to be improved in respect of the ones reported in this paper.
RESULTS AND DISCUSSION
The Fast-GC method was used for quality control measurements in one company that produces a primary brand in the retail spices market. Ten different samples of Capsicumwere tested with standard conditions and showed a wide range of SHVs (Table 1). FastGC equipment permitted good resolution of the peaks of interest. Figure 2 shows two examples of resolution, A and B refer to samples 6 and 7, respectively, in Table 1 with SHVs of 94,585 and 25,619. Figure 2 shows that the identification and quantification of three capsaicinoids, using previously described standard operating conditions, were not difficult even for samples with very different SHVs. The primary advantage of Fast-GC quantitative determination was the rapidity of chromatographic elution. The time required for Fast-GC was comparable with the time required for other proposed HPLC methods, but shorter than that required by other currently proposed GC or GC/MS methods. These advantages were associated with RSD values below 5% and RSDr values below 6%.
CONCLUSION
We evaluated whether Fast-GC was useful for simultaneous analysis of three capsaicinoids commonly utilized for SHV determinations in pepper fruits. The results derived from this chromatographic system showed similar analysis times and good sensitivity and reproducibility compared to more frequently used HPLC-DAD methods. The Fast-GC technique, coupled to a proper extraction method, is particularly suitable for routine quality control. The total time of analysis depends on the preferred extraction technique but the time of Fast-GC analysis results very low (about 8 min) in comparison with the other corresponding GC chromatographicconditlons.
REFERENCES
AOAC Official Method 995.03. 1995 Capsaicinoids in Capsicums and their extractives - Liquid Chromatographic Method. 43: 13-15.
Barbero G.F., Ijazid A, Palma M. and Barroso CG. 2008a. Ultrasound -assisted extraction of capsaicinoids from peppers. Talanta 75: 1332.
Barbero G.F., Liazid A., Palma M. and Barroso CG. 2008b. Fast determination of capsaicinoids from peppers by high-performance liquid chromatography using a reversed phase monolithic column. Food Chem. 107: 1276.
Barbero CF., Palma M. and Barroso CC 2006a. Pressurized Liquid Extraction of Capsaicinoids from Peppers. J. Agrie. Food Chem. 54:3231
Barbero CF., Palma M. and Barroso CG. 2006b. Determination of capsaicinoids in pepper by microwave-assisted extraction-high-performance liquid chromatography with fluorescence detection. Anal. Chtm. Acta, 578: 227.
Choi S.H., Suh B.S., Kozukue E., Kozukue N., Levin CE. and Friedman M. 2006. Analysis of the Contents of Pungent Compounds in Fresh Korean Red Peppers and in Pepper-Containing Foods. J. Agr. Food Chem. 54:9024.
Collins M.D., Wasmund L.M. and Bosland P.W. 1995. Improved method for quantifying capsaicinoids in Capsicum using high-performance liquid chromatography. HortScience 30: 137.
Contreras-Padilla M. and Yahia E.M. 1998. Changes in capsaicinoids during development, maturation, and senescence of chile peppers and relation with peroxidase activity. J. Agr. Food Chem. 46:2057.
Daood H. C, Illés V. Gnayfeed M.H., Mészáros B., Horváth G. and Biacs P.A. 2002. Extraction of pungent spice paprika by supercritical carbon dioxide and subcriticai propane. J. Supercritic. Fluids 23:43.
Garcés-Claver A., Arnedo-Andreis M.S., Abadía J., Gil-Ortega R. and Alvarez-Fernandez A. 2006. Determination of capsaicin and dihydrocapsaicin in Capsicum fruits by liquid chromatography-electrospray/time-of-flight mass spectrometry. J. Agr. Food Chem. 54: 9303.
Hawer W.S., Ha J., Hwang J. and Nam Y. 1994. Effective separation and quantitative analysis of major heat principles in red pepper by capillary gas chromatography. Food Chem. 49: 99.
Kaale, E., Van Schepdael A., Roets E. and Hoogmartens J. 2002. Determination of capsaicinoids in topical cream by liquid- liquid extraction and liquid chromatography. J. Pharm. Biomed. Anal. 30: 1331.
Karnka R., Rayanakorn M., Watanesk S. and Vaneesorn Y. 2002. Optimization of high-performance liquid chromatographic parameters for the determination of capsaicinoid compounds using the simplex method.
Kirschbaum-Titze P., Hiepler C, Mueller-Seitz E and Petz M. 2002. Pungency in paprika [Capsicum annuum). L. decrease of capsaicinoid content following cellular disruption. J. Agrie. Food Chem. 50: 1260.
Korel F., Bagdatlioglu N., Balaban M.Ô. and Hisil Y. 2002. Ground red peppers: capsaicinoids content, scoville scores, and discrimination by an electronic nose. J. Agrie. Food Chem. 50: 3257.
Kozukue N., Han J. S., Kozukue E., Lee S.J., Kim J.A., Lee K.R., Levin CE. and Friedman M. 2005. Analysis of eight capsaicinoids in peppers and pepper-containing foods by high-performance liquid chromatography and liquid chromatography-mass spectrometry. J. Agr. Food Chem. 53: 9172.
Krajewska A.M. and Powers J.J. 1987. Gas chromatography of methyl derivatives of naturally occurring capsaicinoids. J. Chromatogr. A. 409: 223.
Krajewska A.M. and Powers J.J. 1988. Pentafluorobenzylation of capsaicinoids for gas chromatography with electron-capture detection. J. Chromatogr. A., 457: 279.
Laskaridou-Monnerville A. L999. Determination of capsaicin and dihydrocapsaicin by micellar electrokinetic capillary chromatography and its application tovarious species of Capsicum, Solanaceae. J. Chromatogr. A., 838: 293.
Maillard M.N., Giampaoli P. and Richard H.M.J. L997. Analysis of eleven capsaicinoids by reversed -phase high performance liquid chromatography. Flavour Frag. J. 1 2: 409.
Manirakiza P., Covac A. and Schepens P. 1999. Solid -phase extraction and gas chromatography with mass spectrometric determination of capsaicin and some of its analogues from chili peppers [Capsicum spp.). J. AOAC Int. 82: 1399.
Nazari F., Ebrahimi, S.N., Talebi M., Rassouli A. and Bijanzadeh H.R. 2007. Multivariate optimization of microwave-assisted extraction of capsaicin from Capsicum frutescens L. and quantitative analysis by 1H-NMR. Phytochem. Anal. 18:333.
Nordic Committee on Food Analysis (NMKL), 1997. Estimation and expression of measurement uncertainty in chemical analysis. NMKL Procedure No. 5.
Peña-Alvarez A., Ramirez-Maya E. and Alvarado-Suarez L. A. 2009. Analysis of capsaicin and dihydrocapsaicin in peppers and pepper sauces by solid phase microextractiongas chromatography-mass spectrometry. J. Chromatogr. A 1216: 2843.
Perucka I. and Oleszek W. 2000. Analytical, Nutritional and Clinical Methods Section Extraction and determination of capsaicinoids in fruit of hot pepper Capsicum annuum L. by spectrophotometry and high-performance liquid chromatography. Food Chem. 71: 287-291.
Peter K. V. 2001. Capsicum, Chillies, Paprika, Bird's Eye Chilli. Ch.8. In: "Handbook of Herbs and Spices, Vol. 1". K. V. Peter (Ed.) p. 111. Woodhead Publishing Limited, Boca Raton, FL.
Santamaría R.L, Reyes-Duarte M. D., Bárzana E., Fernando D., Gama FM., Mota M., and López-Munguía A. 2000. Selective enzyme-mediated extraction of capsaicinoids and carotenoids from chili Guajillo Puya [Capsicum annuum L.) using ethanol as solvent. J. Agr. Food Chem. 48: 3063.
Saria A., Lembeck F. and Skofitsch G. 1981. Determination of capsaicin in tissue and separation of capsaicin analogues by high-performance liquid chromatography. J. Chromatogr. A. 208: 41-46.
Thomas B.V, Schreiber A.A. and Weisskopf CP. 1998. Simple Method for Quantitation of Capsaicinoids in Peppers Using Capillary Gas Chromatography. J. Agric. Food Chem. 46: 2655.
Thompson R.Q., Phinney K. W., Sander L.C. and M.J. Welch. 2005. Reversed-phase liquid chromatography and argentation chromatography of the minor capsaicinoids. Anal. Bioanal. Chem. 381: 1432.
Todd P.H. Jr, Bensinger M.G. and Biftu T. L 977. Determination of pungency due to capsicum by gas-liquid chromatography. J. Food Sci. 42: 660.
Weaver K.M. and Awde D. B. 1986. Rapid high-performance liquid chromatographic method for the determination of very low capsaicin levels. J. Chrom. A. 367: 438.
Williams O.J., Viyaya-Raghavan G.S., Orsat V. and Dai J. 2004. Microwave-assisted extraction of capsaicinoids from capsicum fruit. J. Food Biochem. 28:113.
Ziino M., Condurso C, Romeo V., Tripodi G. and Verzera A. 2009. Volatile compounds and capsaicinoid content of fresh hot pepper [Capsicum annum L.) of different Calabrian varieties. J. Sci. Food Agr. 89: 774.
Paper received June 6, 2011 Accepted September 27, 2011
M. BONONI and F. TATEO*
Dipartimento di Produzione Vegetale, Facoltà di Agraria, Università degli Studi di Milano,
Vra Celoria 2, 20133 Milano, Italy
Corresponding author: Tel. +39 02 50316540, Fax +39 02 50316539, email: [email protected]
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Copyright Chiriotti Editori 2012
Abstract
The present paper describes the application of Fast-gas chromatography (GC) to quantify the capsalcinoid content in dried pepper fruit with the aim to calculate the Scoville heat values (SHVs) for routine quality control. Fast-GC was used to analyse various samples of dried pepper fruits with a wide distribution of SHVs. The calibration curves demonstrated a polynomial model in the range of 3.0-195.0 µg mL^sup -1^ for capsaicin and 2.6-105.0 µg mL^sup -1^ for dihydrocapsalcin and nordihydrocapsalcin, with correlation coefficients of 0.996 and 0.997, respectively. The data showed an intraday relative standard deviation (RSD) <5% and an interday relative repeatability standard deviation (RSDr) <6%. [PUBLICATION ABSTRACT]
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer





