1. Introduction
Osseointegration has been defined as a direct and functional connection between a bone and an artificial implant [1]. Implant stability is a prerequisite for achieving osseointegration. Micromovements exceeding 50–100 µm may result in fibrous tissue formation instead of osseous integration [2]. Implant stability during the healing process is a result of primary and secondary stability. Primary stability depends on several factors, including bone density site dimensions, drill speed, and drill feed-rate during osteotomy preparation, surgical technique, and macro-/microscopic implant morphology [3,4,5]. Secondary stability is sequential to new bone formation and remodeling at the bone-implant interface. It is also dependent on site-specific bone quality, surgical osteotomy preparation, and implant design [6,7,8].
Initial primary stability stems from implant surface zones engaging in direct contact with the surrounding bone. This mechanical compression provides primary stability up to a certain point in time. Later, bone in direct contact with the implant will undergo remodeling, a process leading to increasing secondary stability. Void spaces between the implant and bone, which do not contribute to primary stability, are filled with a blood clot following implant placement; later, these clots are replaced gradually by mature bone [9].
Several studies have suggested that implants should be inserted using an increased torque to increase their primary stability, which may allow immediate loading [4,5]. To increase primary stability, it has been suggested that tapered implants should be inserted into osteotomy sites prepared with a final drill that is slightly narrower than the implant diameter [10]. However, high localized compression may lead to ischemia and eventually bone necrosis in close vicinity to the implant [11]. This phenomenon is generally considered to be limited to cortical bone prevailing in the crestal compartment of the alveolar process. High compression caused by an insertion torque higher than 40–45 Ncm has been claimed to compromise the local microcirculation, leading to bone necrosis and osteocyte death, followed by bone resorption [12].
While some studies have reported increased bone-to-implant contact (BIC) with increased insertion torque, others have reported a reduced BIC in implants installed with a high insertion torque. Duyck et al. [13] compared the BIC of implants inserted with a high torque (above 50 Ncm due to underdrilling) to that of implants inserted with a low torque (up to 10 Ncm due to overdrilling) in the rabbit tibia. After two weeks, implants inserted with a high torque demonstrated a significantly higher BIC than implants inserted with a low torque. At four weeks, no difference in the BIC was found between the groups [13]. In a canine model, 4.0-mm-diameter implants inserted into undersized 3.2-mm or 3.5-mm osteotomy sites presented extensive necrotic bone areas at the sites of the three coronal implant threads after one week. After three weeks, these regions showed bone remodeling with a restricted amount of new bone [14]. Implants inserted with a high torque (mean, 110 Ncm) in the sheep mandible, composed of thick (3–4 mm) cortical bone, demonstrated intrabony pockets around the implant neck as early as six weeks after placement [15]. Recently, it was reported that the insertion of implants with an overdrilling protocol for the crestal aspect of the osteotomy in a rabbit tibia model resulted in increased short-term crestal bone-to-implant contact [16]. Crestal bone comprises of cortical bone; hence, the coronal design of an implant can affect primary stability and crestal bone resorption [17].
Modifications to the implant macroscopic design and surface roughness are intended to allow the bone anchorage of dental implants and shorten the time required for osseointegration, if possible.
Flutes, which are located at the apex of implants, are intended to increase the self-tapping ability of the implant tip and facilitate implant insertion due to the cutting feature. The reported advantages of fluted implants include a decrease in the heat generated during insertion, insertion torque, operative duration, and the number of surgical instruments [18,19]. Historically, several types of flute designs have been used at the implant tip, such as edge, bowl, and spiral designs. In tapered implants, bowl flutes were reported as the optimal design, with a lower resistance to initial insertion and higher stability, for final instrumentation [20]. As implants are inserted, stress is concentrated on the coronal aspect. It has been previously reported that excessive stress can lead to bone resorption [21,22]. To distribute stress at the coronal aspect, microthreads were introduced. It is advocated that microthreads promote bone formation, distribute stress in the cancellous bone [23], and increase the maximal stress the cortical bone can sustain [24]. To date, no general agreement has been reached regarding the effectiveness and influence of the implant macro- or microscopic design on the bone-to-implant contact. The aim of this study was to evaluate the influence of four different coronal implant neck designs using two different drilling protocols on the BIC in the minipig calvarium.
2. Materials and Methods
Twelve female Sinclair minipigs, 7.5 months of age and 48–53 kg in weight, were used in the study. The number of animals was determined by considering previous studies, and performing a sample size calculation for a statistical power of 0.8 [25,26,27].
Ninety-six NeO ® (Alpha Biotech, Petah-Tikva, Israel) implants (titanium grade, 23; Ra, 1.6 µm; Ø, 3.75/7.5 mm) with 4 different coronal designs, 24 implants of each type, were inserted in the parietal bone of the calvarium (Figure 1).
Type 1—NeO ® implant with 6 coronal flutes (2.26 mm in length, 1.15 mm in width and 0.35 mm in depth) located 0.25 mm apically from the implant shoulder and 8 shallow microthreads (0.05 mm deep; pitch, 2.4 mm).
Type 2—NeO ® implant with 6 coronal flutes (2.62 mm in length, 1.35 mm in width and 0.55 mm in depth) located 0.25 mm apically from the implant shoulder and 4 deepmicrothreads (0.3 mm deep; pitch, 2.4 mm).
Type 3—NeO ® implant with 0 coronal flutes and 4 deep microthreads (0.3 mm deep).
Type 4—NeO ® implant with 2 coronal flutes (1.4 mm in length, 1.15 mm in width and 0.35 mm in depth) located 1.32 mm apically from the implant shoulder and 8 shallow microthreads (0.05 mm deep; pitch, 2.4 mm).
Implants were inserted following preparation of the osteotomy by a drilling unit with 3–5 NCm steps (W&H, Elcomed, Burmous, Austria) with two different drilling protocols:
Drilling protocol 1: Underdrilling protocol (DP1). The osteotomy was prepared starting with a marking drill (Ø, 1.2 mm) followed by a second drill (Ø, 2.0 mm) and a final drill (Ø, 2.8 mm).
Drilling protocol 2: Standard drilling protocol (DP2) as suggested by the manufacturer. The osteotomy was prepared starting with a marking drill (Ø, 1.2 mm) followed by a second drill (Ø, 2.0 mm), a third drill (Ø, 2.8 mm) and a final drill (Ø, 3.2 mm).
Animals were sacrificed after 1 and 3 months of healing. The study protocol was approved by the ethics committee of Assaf-Harofeh Medical Center, Israel (45/2015). The study design followed the Animal Research Reporting In Vivo experiments (ARRIVE) guidelines [28]
2.1. Surgical Procedure
All animals were premedicated with an intramuscular injection of 20 mg/kg ketamine (Ketalar, Pfizer, Herzliya, Israel) and 2 mg/kg xylazine (Sedaxylan, Eurovet Animal Health, Bladel, The Netherlands). Isoflurane 2% and oxygen were then used to maintain anesthesia. After animals were placed in a prone position, the surgical site was meticulously shaved, followed by scrubbing using povidone-iodine 10% (Teva Medical, Tel-Aviv, Israel) + chlorhexidine 4% (Vitamed, Binyamina, Israel) and draping to ensure sterility. Antibiotics (cefazolin, 1 g) and analgesics (7.5 mg of midazolam and 1000 mg of dipyrone) were administered before surgery. Analgesics were administered for an additional 3 days after surgery. A linear midline skin incision was made to expose the bone surface. Osteotomies were prepared lateral to the midline suture in the parietal bone apical to the supraoccipital protuberance and coronal to the frontoparietal suture. Four osteotomies were prepared according to DP1 (one side of the midline cranial suture, randomly assigned), whereas another 4 osteotomies were prepared according to DP2 (contralateral side of the midline suture) (Figure 2). Osteotomies were prepared with sharp drills and profuse irrigation with cold, sterile saline. The location of each implant was chosen randomly.
The submerged implants were sealed with a cover screw. Flaps were closed in layers with resorbable sutures (Vicryl, Ethicon, Cornelia, GA, USA) followed by suturing and wound closure.
2.2. Histological Processing and Analysis
Following euthanasia, samples were chemically fixed and prepared for the preparation of nondecalcified ground sections, as previously described [29,30]. Briefly, the samples were fixed by continuous submersion in 4% neutral buffered formaldehyde for 1 week at +4°C, dehydrated by continuous submersion in ascending concentrations of alcohol and finally cleared by xylol I and xylol II. Thereafter, the samples were embedded in methyl methacrylate (MMA). Polymerization was carried out for up to four days in sealed glass vials submerged in a water bath to assure constant room temperature without movement. The embedded tissue blocks were aligned according to µ-CT imaging [30] and depressions made on the implant platform shoulder above the flutes of implant types 1, 2, and 4 to cut through the center of two opposing flutes. Three sections were cut at a thickness of 500 µm using a low-speed diamond saw with coolant (Varicut® VC-50, Leco, Munich, Germany). After mounting the sections onto acrylic glass slabs, they were ground and polished to a final thickness of approximately 100 µm (Knuth-Rotor-3, Struers, Rodovre/Copenhagen, Denmark) and stained with toluidine blue/McNeil combined with basic fuchsin [27]. Digital photography was performed using a digital camera (AxioCam MRc; Carl Zeiss, Gottingen, Germany) connected to a microscope (Axio Imager M2; Carl Zeiss, Gottingen, Germany).
The total bone-to-implant contact (t-BIC) and coronal bone-to-implant contact (c-BIC) were histomorphometrically evaluated in the central-most ground section of each implant.
For determination of the bone-to-implant contact (BIC), the following landmarks were identified on each ground section (Figure 3):
The implant shoulder
The apical end of the implant
The coronal end of the flute
The apical end of the flute
The histomorphometric analysis was performed as described by Janner et al. [31] R software v 3.3.3 (R Core Team 2013. Vienna, Austria) was used for statistical analysis of the data. A nonparametric test (Kruskal-Wallis) was used to evaluate differences among the implant types and between the drilling protocols. All p values were corrected for multiple testing by applying Bonferroni-Holm’s method. p < 0.05 was set as the significance level.
3. Results
Healing occurred uneventfully in all animals. The histological sections revealed that all implants were integrated with bone. Furthermore, adaption of the macro- and microscopic design with new bone close to the implant profile (Figure 4) along with blood vessels was observed, with no signs of inflammation; additionally, porous cortical bone was evident. After 1 month of healing, the total bone-to-implant contact (t-BIC) of the four implant types ranged from 85–91% without significant differences among the implant types or between the drilling protocols (Figure 5 and Table 1). After 3 months of healing, the mean t-BIC of implant type 1 decreased from 90% to 75% (DP1) and from 85% to 73% (DP2), while the mean t-BIC of implant types 2–4 slightly increased to 83–84%, 91–92%, and 91–96%, respectively, to implant type, without significant differences between the two drilling protocols (Table 1). At 1 month, the mean coronal BIC of implant type 4 was higher than that of the other implant types. However, there were no statistically significant differences between any of the implant types. After three months of healing, the mean coronal BIC values of implant types 2, 3, and 4 increased. The mean coronal BIC values of implant types 3 (91%) and 4 (96%), both implant types with 2 coronal flutes, were significantly higher than those of implants types 1 (74%) and 2 (82%), both implant types with 6 coronal flutes (Figure 6 and Table 2). Within a fixed drilling protocol, no differences among implant types could be found, possibly due to the low sample size.
4. Discussion
The present study investigated the effect of the drilling protocol and implant neck design on the BIC. Two drilling protocols were evaluated: in DP1, implants were inserted into osteotomies smaller than the implant diameter by 0.95 mm; whereas in DP2, implants were inserted into osteotomies smaller than the implant diameter by 0.55 mm, according to the manufacturer’s instructions. The measured BIC of implants inserted according to DP1 and DP2 was comparable for all of the implant types after both 1 m and 3 m of healing for each drilling protocol and ranged from 75–96% after 3 m without significant differences among the implant types or between the drilling protocols. The reported BIC is in agreement with that in a study conducted by Kwon, who reported that the BIC of implants inserted in minipigs ranged from 71–82% after 1 m of healing [32].
Undersized drilling protocols are commonly used for increasing implant primary stability. Such protocols were initially proposed for use in low-density bone, especially in the posterior maxilla [33,34] to enhance the initial implant contact with the bone, optimize bone density, and consequently improve the insertion torque [35]. Conflicting results have been published regarding the effect of these protocols on the implant BIC. In a rabbit tibia model, the insertion of implants with an overdrilling protocol of the crestal aspect of the osteotomy resulted in increased short-term crestal bone-to-implant contact compared to underdrilling [16]. Recently, it was reported that the BIC decreased with overpreparation of the osteotomy site. The initial BIC of implants inserted in undersized osteotomies was the greatest, while the lowest BIC was found in the oversized group [36]. The differences between the two studies may be attributed to the different implant systems, the different sites of implant insertion, and the different drilling protocols, the latter two of which were tibia vs condyle and 0.1 mm vs 0.25 mm overpreparation, respectively. In the current study, the undersized drilling protocol did not have any significant effect on the BIC. Implants inserted according to DP1 and DP2 were comparable for any implant type at both healing times. This probably stems from the soft parietal bone with porous cortical bone into which the implants were inserted.
Four types of implants with different neck microthread and flute designs were evaluated. To the best of our knowledge it is the first study that evaluated the effect of flutes incorporation in implants neck on BIC and it is the first study evaluating two types of microthreads design in-vivo. Regarding the inclusion of flutes in the implant neck, for both microthread types (implants 2 and 3 and implants 1 and 4), the addition of flutes to the coronal aspect resulted in significantly lower BIC values at 3 m. The BIC of implant type 1, with 6 coronal flutes, decreased from 1 m of healing to 3 m, while the BIC of implant type 4, with 2 coronal flutes, slightly increased. A similar trend was observed with implant types 2 and 3. The coronal BIC values of both implant types with 6 coronal flutes did not differ from each other and were significantly lower than those of implant types 3 and 4. The lower BIC values measured for implant types with 6 flutes may stem from excessive stress located at the coronal aspect of the implant. By introducing flutes into the coronal aspect, the implant surface area that can distribute the stress associated with implant insertion decreases dramatically, and the remaining implant surface that is in direct contact with the osteotomy walls is subjected to excessive stress. It has been previously reported that excessive stress is associated with bone resorption [21,22]. To distribute stress at the coronal aspect, microthreads were introduced [23]. Two types of microthreads were compared: threads 0.05 mm in depth (implant types 1 and 4) and threads 0.3 mm in depth (implant types 2 and 3). The coronal BIC of implant type 3, with no coronal flutings, was on average lower than the coronal BIC of implant type 4 at 1 m and at 3 m. This statistically insignificant trend was discerned although implant type 2 has coronal fluting (for implant type 4, coronal flutes compromise approximately 10% of the coronal surface area). As stated above, flutes at the coronal aspect are suspected to impair the coronal BIC, and since implant type 4 presented a higher BIC despite the presence of flutes at the coronal aspect, it can be speculated that the higher number of microthreads of implant type 4 improved the stress distribution and contributed to the higher BIC of implant type 4 compared to that of implant type 3. This notion is supported by Lee et al. [37], who suggested that microthreads increase the implant surface area. Since the microthreads of both implant types are similar in pitch, decreasing the microthread depth enables the production of a greater number of microthreads in a confined area, leading to a greater surface area. Thus, by increasing the surface area, better stress distribution is achieved, resulting in a higher BIC [38]. This is further supported by a recent finite element analysis concluding that the most effective parameters of stress distribution were the depth and pitch of the microthreads [39].
Limitations
The main limitations of the current study are the vast implant design variations, which included both flutes and microthreads, and the investigation of unloaded implants inserted in calvarial bone, resulting in inconclusive results regarding the superior microthread design. Further studies with a single implant design variation and testing of loaded and unloaded implants should be conducted in variable models including, cad-cam assisted, stem cell, and systemic patients [40,41,42].
5. Conclusions
Within the limitations of the current study, the following may be concluded:
The BIC was not affected by the drilling protocol.
Flutes located at the implant coronal aspect significantly impaired the crestal BIC.
Multiple shallow microthreads were associated with a greater crestal BIC.
Further studies should be conducted to confirm the above findings.
Author Contributions
Conceptualization—O.C. and O.M.; methodology, D.D.B. and O.M.; investigation—O.C. and O.M.; writing —original draft preparation—O.C.; writing review and editing—E.W. and G.S. All authors have read and agreed to the published version of the manuscript.
Funding
Alpha Biotech: Petah-Tikva, Israel.
Institutional Review Board Statement
The study was conducted according to the guidelines of the Declaration of Helsinki and approved by the ethics committee of Assaf-Harofeh Medical Center, Israel (45/2015).
Informed Consent Statement
Not applicable.
Data Availability Statement
The data presented in this study are available on request from the corresponding author.
Conflicts of Interest
The authors declare no conflict of interest.
Abbreviations
BIC | bone-to-implant contact |
t-BIC | total-BIC |
c-BIC | coronal-BIC |
CFs | coronal flutes |
SMs | shallow microthreads |
DMs | deep microthreads |
DP1 | Drilling protocol 1 |
DP2 | Drilling protocol 2 |
MT | Microthread |
T | Thread |
FZ | flute zone |
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Figures and Tables
Figure 1. The four types of NeO ® implants inserted: Type 1—implant with 6 coronal flutes and 8 shallow microthreads. Type 2—implant with 6 coronal flutes and 4 deep microthreads. Type 3—implant with 0 coronal flutes and 4 deep microthreads. Type 4—implant with 2 shorter coronal flutes and 8 shallow microthreads.
Figure 2. Calvarium after preparation of osteotomies with the standard underdrilling protocol as suggested by the manufacturer (DP2) and the extra underdrilling protocol (DP1).
Figure 3. Micrograph of a whole implant. A represents the region of total bone-to-implant contact (t-BIC). B represents the region of the coronal third of cortical bone-to-implant contact (flute-BIC). Magnification, 10×.
Figure 4. Typical micrographs of the coronal aspect of implants after 1 m and 3 m, revealing new bone close to the implant profile along with blood vessels with no signs of inflammation. MT, microthread; T, thread; FZ, flute zone.
Figure 5. (A)—Total bone-to-implant contact one month after (t-BIC-1 m) implants were inserted with the standard underdrilling protocol (DP1) and the extra underdrilling protocol (DP2). (B)—Total bone-to-implant contact three months after (t-BIC-3 m) implants were inserted with the standard underdrilling protocol (DP2) and the extra underdrilling protocol (DP1). Data are presented as the average ± sd.
Figure 6. Coronal aspect bone-to-implant contact (BIC) after one month (1 m) and 3 months (3 m) for the four implant types. Data are presented as the average ± sd. * denotes p < 0.05.
Total bone-to-implant contact (BIC) values of implants types 1–4 inserted with the standard drilling protocol (DP2) and the extra underdrilling protocol (DP1). Values after 1 month and 3 months of healing are presented as the average (%) ± sd.
Implant Type 1 | Implant Type 2 | Implant Type 3 | Implant Type 4 | ||
---|---|---|---|---|---|
1 month | DP1 | 89.47 ± 4.87 | 80.03 ± 7.26 | 88.92 ± 5.69 | 90.58 ± 3.34 |
DP2 | 84.73 ± 7.39 | 79.64 ± 19.16 | 88.90 ± 5.06 | 90.78 ± 6.38 | |
3 months | DP1 | 74.66 ± 12.93 | 83.68 ± 8.59 | 91.02 ± 8.05 | 91.31 ± 4.38 |
DP2 | 72.78 ± 18.28 | 82.93 ± 15.27 | 92.41 ± 4.47 | 96.17 ± 2.26 |
Coronal aspect bone-to-implant contact (BIC) after one month (1 m) and 3 months (3 m) of healing. Values after 1 month and 3 months of healing are presented as the average (%) ± sd.
Implant Type 1 | Implant Type 2 | Implant Type 3 | Implant Type 4 | |
---|---|---|---|---|
1 month | 80.73 ± 13.60 | 79.78 ± 14.99 | 88.30 ± 5.81 | 92.15 ± 6.21 |
3 months | 73.92 ± 21.11 | 82.27 ± 14.79 | 91.04 ± 6.75 | 95.96 ± 4.13 |
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2021 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
Abstract
Background: Stress concentrated at an implant’s neck may affect bone-to-implant contact (BIC). The objective of this study was to evaluate four different implant neck designs using two different drilling protocols on the BIC. Methods: Ninety-six implants were inserted in 12 minipigs calvarium. Implants neck designs evaluated were: type 1–6 coronal flutes (CFs), 8 shallow microthreads (SMs); type 2–6 CFs,4 deep microthreads (DMs); type 3–4 DMs; type 4–2 CFs, 8 SMs. Two groups of forty-eight implants were inserted with a final drill diameter of 2.8 mm (DP1) or 3.2 mm (DP2). Animals were sacrificed after 1 and 3 months, total-BIC (t-BIC) and coronal-BIC (c-BIC) were evaluated by nondecalcified histomorphometry analysis. Results: At 1 month, t-BIC ranged from 85–91% without significant differences between implant types or drilling protocol. Flutes on the coronal aspect impaired the BIC at 3 m. c-BIC of implant types with 6 CFs was similar and significantly lower than that of implant types 3 and 4. c-BIC of implant type 4 with SMs was highest of all implant types after both healing periods. Conclusions: BIC was not affected by the drilling protocol. CFs significantly impaired the -BIC. Multiple SMs were associated with greater c-BIC.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details


1 Department of Periodontology and Dental Implantology, School of Dental Medicine, Tel-Aviv University, Tel-Aviv 6997801, Israel;
2 Laboratory of Oral Histology, School of Dental Medicine, University of Bern, 3012 Bern, Switzerland;