1. Introduction
Today, acquiring an intraoral scanner (IOS) for the dental office can be a major step forward for both the quality of the treatments and the overall impression for the patient also serving as an ideal marketing tool [1]. However, the latest IOSs can be a big investment for the dental office and not all dentists may be able to afford it. Thus, the following question arises: are the IOSs released in the previous years a good and reliable alternative to the latest scanners emerging on the market?
The clinical applications of intraoral scanners (IOSs) have increased significantly providing dentists with relevant information that enhance the entire treatment plan and procedure. Minimally invasive dentistry with inlays, onlays and veneers, as well as short span prosthetic restorations, full arch rehabilitation and implant supported restorations can be successfully obtained with the aid of these systems [2,3,4,5].
One of the main advantages of IOSs is the possibility to completely replace the conventional impression, reducing the discomfort for the patient, the working procedure and the overall costs of the impression materials. The immediate evaluation of the quality of the digital impression is another beneficial aspect that enhances the impression procedure [6,7,8,9]. These systems also have a number of limitations including the difficulty in detection of deep preparation margins, the presence of blood and saliva prevents data acquisition of the scanned surface, and also high purchasing and maintenance costs [10].
A key aspect for any long lasting prosthetic restoration is the accuracy of the impression which generates a working model (digital or conventional) that eventually dictates the marginal adaptation and fit of the final restoration. Two variables, “trueness” and “precision”, describe the accuracy of a measurement method (ISO 5725). “Trueness” refers to the closeness of agreement between the arithmetic mean of a large number of test results and the true or accepted reference value. “Precision” refers to the closeness of agreement between test results [11,12,13].
The analysis of the trueness involves a reference virtual model that is generally obtained with an industrial or a desktop laboratory scanner (with accuracy < 5 µm). This high-quality virtual model is then overlapped in a 3D inspection software with the virtual model obtained with the intraoral scanner which shows the differences and similitudes between the meshes [14,15,16].
Primescan (CEREC v.5, Dentsply-Sirona Dental Systems, Bensheim, Germany, 2019) is the latest intraoral scanner from CEREC and uses high-resolution sensors and shortwave light with optical high-frequency contrast analysis for dynamic deep scan (20 mm) [17].
i700 (Medit Link v2.4.4, Medit, Seoul, Korea, 2021) is the latest intraoral scanner from Medit, improving most of the features presented by Medit i500. Data acquisition is based on the principle of 3D in Motion Video Technology/3D Full Color Streaming Capture [18].
CEREC Omnicam (v4.6 Dentsply-Sirona Dental Systems, Bensheim, Germany, 2012), the predecessor of Primescan, is based on the principle of confocal microscopy and active triangulation with a white LED light source [17].
The Planmeca PlanScan (Romexis v4.6.2, powered by E4D Technologies, LLC, Richardson, TX, USA, 2014) works based on the principle of optical coherence tomography and confocal microscopy. It is incorporated with a blue light real-time laser video streaming technology [17].
Medit T500 (Seoul, Korea) is a desktop dental scanner with an accuracy of <7 µm, that works based on the principle of phase-shifting optical triangulation, using a blue LED light source [19].
Additional information about the intraoral scanners is provided in Table 1.
The aim of this study was to evaluate the differences in accuracy between four intraoral scanners in the scenario of a short span fixed partial denture and to determine if the improved accuracy could have a major impact on the final outcome of the digital impression procedure.
The null hypothesis was that the latest IOSs will generate significantly higher accuracy values when compared with IOSs released more than five years ago (to the date of the present study).
2. Materials and Methods
Four intraoral scanners: Cerec PrimeScan, Medit i700, Cerec Omnicam and Planmeca PlanScan and one laboratory scanner (Medit T500) were used in this study.
We considered IOSs released more than five years ago as being part of the old generation of IOSs. The two old generation scanners selected for this study were Planmeca PlanScan and Cerec Omnicam. The new generation includes recently launched IOSs that are considered by the manufacturer to be the most advanced and developed in technology and software, also being their current flagship product. Medit i700 and Cerec Primescan were selected as the new generation scanners to be tested in this study.
A typodont (AG-3; Frasaco, Tettnang, Germany) was selected for the study. Full all ceramic crown preparations with deep chamfer, 1 mm of axial reduction and 1.5 mm of occlusal reduction were manually conducted on the left mandibular second premolar and on the left mandibular second molar of the typodont. An edentulous area was present in place of the left mandibular first molar, obtained after removing the tooth from the typodont and filling the gap with polyvinyl siloxane (Elite HD+, Zhermack, Badia Polesine, Italy).
Ten consecutive scans were performed by a single operator with each intraoral scanner in order to obtain the 3D meshes to be compared with the reference data. The scanning path was identical for all the IOS’s starting on the occlusal surface of the distal abutment, coming to the buccal of both abutments and the edentulous area, then, the occlusal surface of mesial abutment and edentulous area and finishing with the lingual surfaces. Calibration was performed according to the manufacturer’s instructions. All the scans were performed in the same day, around the same time, the light intensity being verified to be constant during all of the scans with the help of a lux-meter (GM1010, Benetech, Shenzhen, China). During the scanning procedure, the typodont was mounted inside the phantom (Phantom head PK-2 TSE; Frasaco) and the position of the assembly was not moved or altered in any way between the scans. Consequently, 40 STL files were obtained: 10 meshes for PrimeScan, 10 meshes for i700, 10 meshes for Omnicam and 10 meshes for PlanScan.
The desktop dental scanner Medit T500 was used to obtain the reference scan in standard tessellation language (STL). The calibration of the desktop scanner was performed according to the manufacturer’s instructions in order to obtain a highly accurate 3D model (the reference model). The scanned data were processed with the help of Geomagic Control X (Version:16.0.2.16496, 3D Systems, Wilsonville, OR, USA), a complete metrology grade, quality control software used for analyzing and comparing 3D meshes. The reference data were uploaded into the software and the area of interest including the two prepared teeth and the edentulous area was isolated from the rest of the mesh. The isolating procedure was done manually inside the software, respecting the margins of the preparations and the boundary of the polyvinyl siloxane that simulated the edentulous area. For consistency, the area of interest remained the same for all the following comparisons, not being necessary to retrace the margins of the area of interest as the reference data remain the same and only the measured data are being swapped. The intraoral scanner mesh was uploaded into the software as measured data to be compared with the reference data. Only the isolated area of interest was analyzed between the meshes, in order to focus on the scanning accuracy of the area for the fixed partial denture, excluding the remaining area of the typodont.
The “initial alignment” function was executed to superimpose the intraoral scanner mesh over the reference mesh followed by the “best fit alignment” that concentrates on the alignment of the isolated area of interest in order to secure a precise overlapping of the two meshes. In order to analyze the deviation between the reference and measured data, the “3D Compare” function was used by projecting all paired points onto the reference data. A color-coded map was rendered displaying the deviation patterns of the investigated surfaces. The color-coded map was set to display deviations between ±50 µm. On the color-coded map scale, outward displacement is presented towards the red spectrum while inward displacement is presented towards the blue spectrum. The areas with no deviation between the analyzed mesh and the reference one, are being presented in green as there is less than ±1 µm between the compared surfaces in that specific area (Figure 1).
The entire process was executed for all the intraoral scanner meshes in order to obtain the numerical values for trueness (Figure 2). To obtain the precision values, each intraoral scanner mesh was compared with all the other meshes within its group following the same protocol.
The statistical analysis was conducted by uploading the standard deviation data obtained from the metrology software into the MedCalc statistical software. The Kolmogorov–Smirnov test resulted that both the data for trueness and precision were non-parametric. To further analyze the data, the Kruskal–Wallis test was used on the datasets. The level of significance was set to = 0.05.
3. Results
Table 2 presents the trueness values while Table 3 presents the precision values.
Regarding the trueness values, the Kruskal–Wallis test indicated that there is a statistically relevant difference between the groups (p < 0.001). The post-hoc analysis (Conover test) was also executed. The Medit i700 group displayed the best level of trueness with the mean value of 23.25 µm, closely followed by the PrimeScan group which displayed the mean of 25.55 µm. The Omnicam group showed a decrease in trueness with a mean of 32.3 µm while the PlanScan group displayed the lowest level of trueness with a mean value of 75.8 µm. This translates into roughly 2.3 µm difference in trueness between MedIT i700 and Cerec PrimeScan while Cerec Omnicam shows differences of 10 µm and Planmeca Planscan presenting a difference in trueness up to 53 µm (Figure 3).
Regarding the precision values, the Kruskal–Wallis test returned a value of p < 0.001, rejecting the null hypothesis and showing that the groups differ significantly. The post hoc analysis (Conover test) concluded that the PrimeScan group shows the best precision, the Medit i700 group being second in precision performance, the Omnicam group shows a decrease in precision and the PlanScan group being the least precise (Figure 4).
4. Discussion
Accuracy of digital impressions is essential in order to obtain high-quality restorations as well as to the predictability of the treatment [20]. The aim of IOSs is to increase the patient’s acceptance and comfort, reduce the clinical time and also to eliminate the errors that come with the conventional impression procedure [21].
Today, with more and more IOSs being present in the dental offices, there are clinicians who are unsure whether or not there is a real need to replace the existing IOS with a newer one. Thus, the following question arises: “Is the accuracy of the older generation of IOSs an impediment for obtaining a quality intraoral digital impression?”
The results of the present in vitro study were in favor of the null hypothesis that the accuracy of the latest IOSs has improved significantly when compared to older generations of scanners.
The trueness and precision of intraoral scanners has been investigated in numerous scenarios by different authors. The results of this study were consistent with the results of other previous studies [22,23,24]. Mutwalli et al. evaluated the accuracy of three IOS (Trios 3, Trios 3 mono, Itero element) and obtained significant differences between the digital impressions [22]. Diker and Tak evaluated the accuracy of six IOS in the scenario of multiple single crown preparations distributed across a dentate arch. They reported the highest accuracy (trueness, precision) for Primescan (25 and 10 µm), followed by Trios (40.5 and 11 µm), Omnicam (41.5 and 18 µm), Virtuo Vivo (52 and 37 µm), iTero (70 and 12 µm) and Emerald (73.5 and 60 µm) [23]. Ender et al. also evaluated the trueness and precision of eight IOS and concluded that there were significant differences between the digital impressions, the best values being obtained for the Primescan IOS [24].
The clinical accuracy of impressions dictates the marginal and internal fit of the final restorations. However, all the following steps involved in the fabrication of the restoration may also introduce errors [25,26]. Even though CEREC Omnicam and Planmeca PlanScan are available on the market for almost nine years, the Omnicam IOS provides high-quality intraoral scans which are not far behind (difference of approx. 10 µm) of those obtained with the latest IOSs. On the other hand, the differences in trueness between the Planmeca PlanScan IOS when compared to Primescan and Medit i700 were significantly higher (difference of approx. 53 µm).
Regarding the precision of the investigated IOSs, the mean best performance was found for the Primescan IOS, followed by Medit i700, Omnicam and Planmeca IOSs. Primescan and Medit i700 displayed precision values that are very close within the same interval of 5 µm to 15 µm, with Primescan definitely keeping the median lower than 7 µm. Omnicam and Planmeca Planscan displayed values that differ significantly, with more outliers and resulting in a severe decrease of precision.
The accuracy of digital impression is conditioned by a number of external factors including the type of preparation, fluid isolation, limited space and access for the scanner, scanning area as well as lighting conditions in the office [27,28,29,30]. Being an in vitro study, there were a number of limitations. No clinical external factors were simulated (humidity, blood, patient movements). The scanning pattern was the same for all the IOSs, even though previous studies show conflicting reports regarding the influence of the scanning path on the final accuracy of digital impressions [31,32]. As another limitation, a better material could have been used to fill the gap of the edentulous area. Scannable impression materials such as Honigum Pro Scan (DMG) would have been more suitable for the scanning procedure and could have reduced the influence of the reflection of the material. Although the material not being ideal, it must be taken into account that the influence of the reflection of the material was constant throughout the entire experiment as the scans were performed in the same context and under the exact same conditions.
5. Conclusions
Based on the results of the present study, the following conclusion can be drawn:
Both CEREC PrimeScan and Medit i700 IOSs generated digital impressions with the highest accuracy of all the investigated IOSs;
Even if there is a time gap of +7 years between Omnicam and the first two mentioned IOSs, the difference in trueness and precision was low ( approx. 10 µm). The Planscan IOS showed the lowest accuracy, even when compared with Omnicam IOS.
Author Contributions
A.J.: Substantial contributions to the conception or design of the work, interpretation of data, final approval of the version to be published; A.B.: Data collection; A.-B.F.: acquisition, analysis and interpretation of data for the work, critical revision of the article; R.N.R.: acquisition, analysis, and interpretation of data for the work, critical revision of the article; All authors have read and agreed to the published version of the manuscript.
Funding
This research received no external funding.
Institutional Review Board Statement
Not applicable.
Informed Consent Statement
Not applicable.
Data Availability Statement
All data are available upon request.
Conflicts of Interest
The authors declare no conflict of interest.
Abbreviations
The following abbreviations are used in this manuscript:
IOS | Intraoral Scanner |
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Figures and Tables
Figure 1. Color-coded map of the analyzed area. Deviation is displayed in mm on the color scale with outward (red) and inward (blue) displacement.
Figure 4. Boxplot displaying the precision values of the four intraoral scanners.
Table displaying additional information about the selected IOSs.
IOS | Manufacturer | Software | Features |
---|---|---|---|
CEREC Primescan (2019) | Dentsply-Sirona Dental Systems, Bensheim, Germany | Sirona Connect 5.1 (2019) | High-resolution sensors and shortwave light with optical high-frequency contrast analysis for dynamic deep scan (20 mm) |
Medit i700 (2021) | Medit, Seoul, South Korea | Medit Link 2.4.4 (2021) | 3D in Motion Video Technology/3D Full Color Streaming Capture |
CEREC Omnicam (2012) | Dentsply-Sirona Dental Systems, Bensheim, Germany | Sirona Connect 4.6.0 (2018) | Confocal microscopy and active triangulation with a white LED light source |
Planmeca Planscan (2014) | E4D Technologies, LLC, Richardson, TX, USA | Planmeca Romexis 4.6.2 (2018) | Optical coherence tomography and confocal microscopy |
Data obtained when comparing for trueness.
Primescan | i700 | Omnicam | PlanScan | |
---|---|---|---|---|
M1 | 21.7 µm | 25.6 µm | 40.4 µm | 76 µm |
M2 | 23.4 µm | 25.5 µm | 31.7 µm | 68.5 µm |
M3 | 33.3 µm | 26.9 µm | 32.2 µm | 66.6 µm |
M4 | 26.1 µm | 23.9 µm | 43.4 µm | 96.1 µm |
M5 | 21.3 µm | 26.8 µm | 28.5 µm | 86.5 µm |
M6 | 21.6 µm | 25.7 µm | 40.1 µm | 75.6 µm |
M7 | 23.1 µm | 24 µm | 31.5 µm | 67.5 µm |
M8 | 26 µm | 22.8 µm | 32.4 µm | 64.8 µm |
M9 | 24.4 µm | 23.3 µm | 30.1 µm | 89.8 µm |
M10 | 21.1 µm | 25.8 µm | 40.2 µm | 85.2 µm |
Median (IQR) | 23.25 (3.97) µm | 25.55 (1.85) µm | 32.3 (8.62) µm | 75.8 (18.42) µm |
Data obtained when comparing for precision.
Primescan | i700 | Omnicam | PlanScan | |
---|---|---|---|---|
M1 | 4.9 µm | 7.6 µm | 15.4 µm | 28.2 µm |
M2 | 6.6 µm | 11.3 µm | 17.1 µm | 26.7 µm |
M3 | 6.5 µm | 13.8 µm | 37.9 µm | 73.5 µm |
M4 | 6.5 µm | 7.6 µm | 15.9 µm | 60.8 µm |
M5 | 6.8 µm | 11.4 µm | 17.4 µm | 74.2 µm |
M6 | 5.8 µm | 7.7 µm | 15.8 µm | 64.2 µm |
M7 | 4.7 µm | 7.6 µm | 33.4 µm | 78.9 µm |
M8 | 6.4 µm | 13.1 µm | 15.3 µm | 61.4 µm |
M9 | 7.3 µm | 14 µm | 13.1 µm | 29.3 µm |
M10 | 7.1 µm | 14.7 µm | 12.4 µm | 27.8 µm |
M11 | 6 µm | 9.1 µm | 17.8 µm | 90.1 µm |
M12 | 7.1 µm | 8 µm | 13.4 µm | 97 µm |
M13 | 6.1 µm | 7.9 µm | 13.7 µm | 60.8 µm |
M14 | 7.2 µm | 9.1 µm | 15.1 µm | 90.1 µm |
M15 | 7.3 µm | 11.2 µm | 12.1 µm | 64.2 µm |
M16 | 6.8 µm | 14.5 µm | 13.4 µm | 73.5 µm |
M17 | 5.9 µm | 9.4 µm | 12.3 µm | 91.1 µm |
M18 | 6.8 µm | 11.7 µm | 33.7 µm | 73.3 µm |
M19 | 5.7 µm | 9.1 µm | 12.9 µm | 72.5 µm |
M20 | 5.8 µm | 7.3 µm | 13.4 µm | 22.2 µm |
M21 | 7.8 µm | 5.4 µm | 17.4 µm | 74.4 µm |
M22 | 6.4 µm | 7.8 µm | 15.4 µm | 61.7 µm |
M23 | 7.1 µm | 14.7 µm | 17.1 µm | 29.1 µm |
M24 | 6.5 µm | 8.1 µm | 17.3 µm | 74.3 µm |
M25 | 7.3 µm | 5.4 µm | 14.5 µm | 22.2 µm |
M26 | 4.9 µm | 13.1 µm | 15.2 µm | 73.5 µm |
M27 | 5.3 µm | 14.4 µm | 32.1 µm | 90.1 µm |
M28 | 7.3 µm | 11.2 µm | 32.9 µm | 73.5 µm |
M29 | 6 µm | 5.4 µm | 13.3 µm | 77.1 µm |
M30 | 7.2 µm | 5.9 µm | 15.9 µm | 22.9 µm |
M31 | 6.6 µm | 7.4 µm | 13.1 µm | 60.4 µm |
M32 | 4.7 µm | 7.8 µm | 17.2 µm | 97.8 µm |
M33 | 4.8 µm | 11.3 µm | 33.2 µm | 61.5 µm |
M34 | 5.8 µm | 11.1 µm | 13.4 µm | 62.6 µm |
M35 | 7.3 µm | 7.5 µm | 12.3 µm | 75.2 µm |
M36 | 5.9 µm | 7.9 µm | 15.6 µm | 29.3 µm |
M37 | 6.5 µm | 9.2 µm | 17.9 µm | 91.4 µm |
M38 | 6.6 µm | 8 µm | 18.3 µm | 64.7 µm |
M39 | 7.1 µm | 9.1 µm | 15.9 µm | 78.9 µm |
M40 | 4.8 µm | 7.3 µm | 12.3 µm | 22.5 µm |
M41 | 4.9 µm | 13.2 µm | 13.4 µm | 24.3 µm |
M42 | 5.7 µm | 11.2 µm | 14.5 µm | 60.1 µm |
M43 | 6.5 µm | 14.6 µm | 13.6 µm | 68 µm |
M44 | 7.1 µm | 9.3 µm | 33.4 µm | 90.3 µm |
M45 | 6.4 µm | 7.4 µm | 13.4 µm | 74.3 µm |
Median (IQR) | 6.5 (1.3) µm | 9.1 (3.8) µm | 15.4 (4) µm | 68 (15.1) µm |
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2021 by the authors.
Abstract
Is the accuracy of the older generation of IOSs an impediment for obtaining a quality intraoral digital impression? The aim of this study was to evaluate the differences in accuracy between four intraoral scanners (two of the latest IOSs and two older IOSs) in the scenario of a short span fixed partial denture, and to determine if these differences have a major impact on the quality of the final digital impression. Four intraoral scanners: Cerec PrimeScan, Medit i700, Cerec Omnicam and Planmeca PlanScan and one laboratory scanner were used in this study. The scanned data were processed with the help of Geomagic Control X, a metrology software. For the statistical data analysis, the level of significance was set as 0.05 (
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details



1 TADERP Research Center, Department of Prosthodontics, University of Medicine and Pharmacy “Victor Babes”, B-dul Revolutiei 1989, No. 9, 300580 Timisoara, Romania;
2 Department of Prosthodontics, University of Medicine and Pharmacy “Victor Babes”, B-dul Revolutiei 1989, No. 9, 300580 Timisoara, Romania;