1. Introduction
Foodborne pathogens include different infectious biological agents that cause disease and are found in different food or water products. Each year, the United States federal government estimates that there are 128,000 hospitalizations and 3000 deaths due to foodborne illness annually [1]. Foodborne illness outbreak is defined as at least two illnesses caused by the same pathogen and linked to the same infective source [2]. The year 2021 saw 17 reported foodborne illness outbreaks, resulting in 1424 cases of sickness, 379 hospitalizations, and seven deaths [3]. In addition to the health effects, the last published estimated cost of foodborne pathogens was 17.6 billion USD annually [4]. There is a high interest in the early detection of foodborne pathogens that will help minimize the public health and economic burden caused by foodborne illness.
The detection of foodborne pathogens can be carried out using various methods and kits. The food industry conducts periodic testing for microbial contamination at specific control points during manufacturing and delivery as the standard procedure [5,6]. The gold standard for the detection of foodborne pathogens is culture based and includes visual, biochemical, and immunological means before or after enrichment. In this method, a sample is introduced to a nutrient filled medium, incubated, grown, and plated. It is not ideal due to the time-consuming transportation, robustness, is expensive, and requires skilled labor [7,8,9]. It is even more problematic for industries in remote and rural areas that cannot afford the time and monetary cost of such procedures [10,11]. Due to these restraints, alternative detection methods have been developed where the most common are polymerase chain reaction (PCR) and enzyme-linked-immunosorbent assay (ELISA), which detect the pathogen by finding either a specific DNA or RNA sequence or a specific protein. Both PCR and ELISA will detect the pathogen quickly, but in most cases, cannot distinguish between live and dead bacteria in addition to being incompatible with field conditions [9,12]. Another approach is to develop biosensors; these devices combine a biological recognition element (i.e., bacteriophage, antibody, enzyme, protein, etc.) with a transducer that transforms the interaction between the target pathogen and the biorecognition element to an electrical signal [13].
Electrochemical biosensors (ECBS) have an advantage with respect to other biosensors due to their selectivity, sensitivity, and relative simplicity of use. ECBS have several techniques, differentiated according to the electrical signal that is measured. The most common technique for foodborne pathogens is electrochemical impedance spectroscopy (EIS) due to its very high sensitivity but is limited by the alternating current (AC) potentiostat required for signal generation, which reduces its portability and increases the cost with respect to direct current (DC) methods [14,15,16,17,18,19,20]. In contrast, amperometric methods (i.e., cyclic voltammetry, chronoamperometry, linear sweep, etc.) have an advantage with low cost and an equivalent sensitivity to EIS. Their main disadvantages are the need to continuously correct their calibration due to Faraday’s processes, higher sensitivity to mass transfer limitations in the solution, and relative high applied potential [15,21,22,23,24].
Microfluidics is a term describing the movement of fluids in geometrically restricted dimensions in orders of magnitude of 10−4 m or smaller and manipulate them at a µL-nL volume. In the field sensors, the initial use of microfluidics was to improve the sensor’s performance and reduce the consumption of reagents. In addition, microfluidic channels enable the integration of separation, mixing, and monitoring within a single device. Advances in micro and nano fabrication techniques have proven to improve the synergy of new electrochemical biosensors and microfluidic designs to achieve better portability, reduce energy consumption, need of sample pretreatment, and better integrability of the systems into existing production lines. The combination of microfluidics and biosensors have been a major vector in the development of new lab-on-a-chip (LOC) or miniaturized total analysis system (µTAS) platforms [21,25,26,27]. The major reasons for the focus on point of care LOC platforms is their ability to minimize the required pretreatment, automate all fluid handling, and the integration of sample preparation to detection on one easy to use device that is portable and does not require any specially trained personal or additional cost due to expensive equipment [28,29].
Foodborne pathogens ECBS have been the subject of many reviews in recent years. Mei et al. [30] focused mostly on carbon nano-materials on the surface of the electrochemical biosensor and their advantages in detecting foodborne pathogens. Villalonga et al. [31] focused on ECBS for food bioprocess monitoring, and Curulli et al. [32] focused on ECBS for food toxins and contaminants. All three reviews mentioned microfluidic channels, but the focus of those review articles was not specific toward ECBS platforms. Other recent reviews have focused more on the integration of microfluidics to ECBS and other biosensors, but have not placed an emphasis on foodborne pathogens [33,34]. Additionally, none of the foodborne pathogen biosensor reviews discussed and analyzed the established diagnostic standards accepted in the food industry or the health impact of foodborne diseases. Hence, it is clear that the need for a review that combines all three subjects is needed. This review starts with an overview of the established standards regarding foodborne pathogens that set the requirements for current and future detection methods. In addition, recent advances in LOC platforms for the detection of different pathogens will be analyzed, and the design and capabilities of both the microfluidic channels and biosensors will be discussed.
2. Foodborne Pathogen Statistics and Standards
The standard that governs the standard for the allowed concentration of foodborne pathogens is set by governmental organizations. The standard of whether or not a product is fit for human consumption is set for each specific type of food products separately such as dairy, shellfish, ready to eat foods, etc. While some countries like Great Britain set quantitative limits that correspond to consuming population susceptibility and the infective dose, other countries like the United States of America have mostly set the standard on any detectable trace of these pathogens in the products tested [35,36,37,38,39]. These standards are affected by the minimum infective dose (MID) of each foodborne pathogen, and describe the detectable amount in a specific food or drink. Both the standards, MID, products that contain the pathogen, and the fitness of the tested sample for human consumption for the detected concentration are shown in Table 1. As seen in Table 1, not all pathogens appear in both standards, which is due to its prevalence in that country. These standards will be mostly used during testing prior to the product reaching the consumer or during specific steps in the production and distribution chain [35,36,37,38,39,40]. It is important to also analyze the effect of the most common pathogens on the infected individuals. The biggest issue with such analysis is generated from the nature of the illness, which in most cases will be very mild and will not be diagnosed. The analysis shown in Figure 1 displays the percentage of hospitalization and mortality from the total number of confirmed illnesses in the U.S. between 1996 and 2020. With the exception of Listeria monocytogenes, the mortality rate is low, but the hospitalization rate is over 20% for most of the common pathogens. The hospitalization rates can explain the high annual cost of foodborne pathogens and why they have received the focus as targets for different biosensors [4].
3. Microfluidic Channel Material Choice
The characteristics of a microfluidic channel are derived from the material used to fabricate them. The channel’s biocompatibility, reusability, fabrication simplicity, and cost are among the first characteristics that come to mind when considering a microfluidic channel for electrochemical biosensor use. In addition to the qualities that are important, their ability to be used in the field, the use of low sample volume, and compatibility to be used with multiple different pathogens (multiplexability) will make a huge impact when they are considered as a viable solution for detection in industrial settings [51,52,53,54]. Focus on microfluidics for foodborne pathogen electrochemical detection since 2017 has shown that the main materials used are glass [55], polydimethylsiloxane (PDMS) [16,17,19,20,56], thermosets [18], paper [57], and thread-based [58]. In addition, there are also other materials that should be considered as microfluidic channel fabrication materials, although not as common such as polymethylmethacrylate (PMMA), cycloolefin polymer (COP), cyclic olefin copolymer (COC), and silicon. These characteristics were evaluated on a scale of 1–5 and are presented in Figure 2 as per channel material, while the details leading to the evaluation are presented below. Some microfluidic channels can be made by the combination of two or more materials and their characteristics will change accordingly. Figure 2 further explains the popularity of glass and PDMS since both display preferable traits in most categories.
3.1. Glass
The word glass is used to describe various materials among them: borosilicate [51,59,60], Pyrex [61], soda lime [62], quartz [63], and others [64,65]. Glass microfluidic channels are usually fabricated by using photolithography to print complex patterns on them and etch specific areas to form channels of specific height and width. Due to its amorphic structure, glass etching usually results in a round cross-sectional profile, which can help create a more homogenous flow pattern but forms a challenge when a high aspect-ratio is required [51,59,61]. Other fabrication methods include micromachining, where material is removed from the substrate and bonding or adhering it to another substrate [60] or laser patterning, where a beam of high-energy laser is used to pattern the channel [62,63]. These fabrication methods are highly accurate, where their resolution is determined by the etching chemicals or the resolution of the lithography and laser machines. The machinery usually comes with a very high-price tag and the use of harsh chemicals increases the complexity of the fabrication process. Glass is also thermoconductive, which means that the temperature of the work environment will be limited to a range that will not affect either the sample or the sensor. Another limitation comes from the hardness of glass and its brittle nature, which makes it harder to use in field conditions and makes the addition of valves or bonding very challenging. Although glass is very biocompatible, it is not permeable to gases, which limits the time a live pathogenic sample can survive in it. Additionally, glass has some very attractive qualities. Its resistance to most organic solvents significantly improves its ability to be washed and reused for multiple experiments, and it is very compatible with metal deposition, high surface stability, and as a substrate, it is commonly found, which reduces the overall cost of the devices. Another quality of glass is its electro isolation property, which allows for the incorporation of electrophoresis within it [51,52,53,54,55,66].
3.2. Silicon
Silicon microfluidic channel fabrication is very similar to glass. Si microfluidic channels are fabricated by means of micromachining and photolithography and wet etch similarly to the fabrication of glass microfluidics. Another technique that is mostly used in Si is the buried channel technique, where a deep vertical trench is etched into the Si by deep reactive ion etching (DRIE), followed by isotropic etching of the bottom [67]. Another advantage of Si lies in its ability to fabricate thin membranes that can be used to form integrated micropumps [68] and microvalves [69] in the channel. Silicon and glass have very similar characteristics, but one major difference is due to the crystalline structure of silicon, which causes a rectangular cross-sectional shape, while glass has a round one. Another difference between the two is the fact that Si is opaque and will not let light pass through it [51].
3.3. PDMS
PDMS is the most popular material used for microfluidic fabrication, in general, and as a microfluidic channel for foodborne pathogen microfluidic electrochemical detection specifically [16,17,19,20,56]. PDMS microfluidic channels are fabricated using a mold, also known as soft lithography, along with low temperature curing, which makes it very repeatable and highly cost efficient. The advantages of PDMS includes the ease of bonding with other PDMS components to form complexed multi-level channels, and hard substrates such as glass or Si to provide mechanical stability. It is very compatible with a high concentration of valves, it is biocompatible, possesses very low toxicity, has a high permeability to gases, which allows a long biostability time of living pathogens in the channel, and supports a very low resolution and any cross-sectional profile, which depends on the mold used. While these advantages make it a very popular material to be used, it also possesses some significant disadvantages. PDMS is hydrophobic and tends to adsorb or absorb small hydrophobic molecules into its walls and cause swelling. It is very sensitive to most organic solvents, which restricts it to aqueous samples only and reduces its reusability. The rigid surface of PDMS can cause pathogens to be trapped in its surface, and its high permeability to gases can change the concentration of a sample due to water evaporation through its walls [51,52,53,54,66].
3.4. PMMA
PMMA is a transparent and rigid thermoplastic polymer, which makes it ideal for sustainable applications. It also possesses a glass-like quality with its clarity, UV resistance, low-toxicity, and transparency, with half the density and an order of magnitude better impact resistance. PMMA is chemical resistant and is not affected by aqueous solutions, detergents, inorganic acids, alkalis, and aliphatic hydrocarbons. Among its disadvantages are low impact resistance with respect to other polymers, very limited heat resistance, sensitive to some organic solvents, poor wearing resistance, and tends to crack under medium to high load [70,71]. One of the common fabrication methods is hot embossing, where a piece of PMMA is placed on a Si or metal negative master pattern. The system is then heated under continuous pressure [72]. Another technique is room-temperature imprinting, where the PMMA is placed on a silicon template and then pressed together under high pressure [73]. As with most polymers, PMMA microfluidic channels can also be made with injection molding, where PMMA pallets are melted and injected on a master template under high pressure and then cooled down to room temperature [74]. Finally, as in glass and silicon, PMMA can be molded into microfluidic channels by laser ablation or wet etching [75].
3.5. COP/COC
COP and COC are promising materials for microfluidic channels due to their chemical resistance for polar solvents, low water absorption, transparency, ease of fabrication, and bio-inertness [76,77]. COP and COC fabrication methods are laser ablation, injection molding, hot embossing, and nanoimprint lithography. They are also thermoplastic polymers like PMMA and possess a very high electric insulating capability. Their major disadvantage is their low chemical resistance to organic solvents [78,79].
3.6. Thermosets
Thermosets such as VisiJet® M2R-CL [18] are very limited in their microfluidic applications, mostly due to their high cost and high stiffness. Their advantages lie with their high chemical and thermal stability, which improves their field compatibility and reusability. Furthermore, their compatibility with 3D printing allows them to be shaped into highly complexed channels, but reduces their ability to work with a low sample volume. Thermosets can also support very high aspect ratio due to their high strength [51,52].
3.7. Paper
The fabrication of paper-based microfluidics is conducted by creating hydrophobic barriers on selected areas on the paper substrate to force the sample to flow in a specific path. Paper-based microfluidics are very cost effective. Paper-based microfluidics are also very simple to make, have high porosity and physical absorption, which makes them more compatible with the field. They are easily sterilized and modified, which assists in letting only specific biocomponents through, are biocompatible, and do not require a pump or any other supporting equipment. Alternatively, paper-based systems have very low mechanical properties, and in a more complex design, the sample flow might experience some challenges [54,80].
3.8. Thread Based
Thread-based materials for microfluidics have some great advantages with respect to paper-based ones. Thread-based microfluidics are very cost efficient since they do not require a clean room, complex fabrication methods, or expensive machinery. They are therefore also very simple to fabricate, while their hydrophilic and capillary nature make pumps and hydrophobic barriers redundant. Most threads used for bio detection are very biocompatible, easily modified with different biorecognition elements, and can be easily shaped to almost any planar or 3D structure. Their low weight and handling simplicity make them relatively suitable for field work. Additionally, threads cannot be reused, and while they do have high strength compared to paper, overall, they are very sensitive to mechanical strain. Thread-based microfluidics are still only available in laboratory settings and have not yet been used for any commercial applications [58,81,82].
4. Microfluidics for Sample Preparation in Electrochemical Biosensors
Microfluidics for the use of electrochemical pathogenic biosensors can be divided according to their goal in the system [27,60,83]. The goals from recent publications were mapped and found to be focused on separation, concentration, detection, and mixing of reagents before detection, as seen in Figure 3. Examples for each category since 2018 are shown in Table 2, where each example includes a summary of the method of which the goal is achieved, along with the target pathogen, flow characteristics, channel material, and the electrochemical detection technique. A few trends that are emerging from the data are that the majority of microfluidic channels are fabricated with traditional photolithography methods to create a mold out of SU-8 photoresist and use it to create an inversed PDMS channel [16,17,19,20,56,84]; other fabrication methods are micromachining glass [55], cotton thread [58], and 3D printing of polyacrylate [18]. PDMS channels are used since the technology is very established and allows one to form complex 2D designs quickly and accurately when the smallest segment’s dimension is determined by the resolution of the photolithography mask aligner and the type of photoresist used to form the mold. The advantage of 3D printed channels is the ability to create complex 3D shapes without the need for an alignment process. It is also very clear that the majority of microfluidic integrated electrochemical biosensors for foodborne bacteria use EIS as their electrochemical technique [16,17,18,19,20]; differential pulse voltammetry (DPV) was used for the detection of Norovirus [56], potentiometric electromotive force (EMF) was used to detect Salmonella typhimurium [57], and amperometric techniques were used to detect Vibrio parahaemolyticus [58]. It should be noted that most microfluidic channels reported in the literature are meant for multifunctional use, which include two or more of the following functions: separation or isolation, concentration, enrichment, mixing, detection, etc. [83,85].
4.1. Microfluidic Separation Channel
Microfluidic separation can be achieved with different methods. The most common approach is the use of microbeads. These microbeads can act as a filter according to their size and concentration within a specific area in the channel [56]. Another use of microbeads is by creating complexes of the target pathogen and magnetic microbeads and separating them by exposing it to a magnetic field [16,86,87]. Except for nanostructures, another separation approach is to utilize mechanical forces such as centrifugal forces to separate the pathogen from the sample. The channels are designed to separate the target pathogen from the sample according to their size or mass [88,89,90]. In addition, another approach for separation is using external forces such as acoustophoretic separation that can separate large particles (>10 µm) from the target pathogens (≤4 µm) by applying ultrasonic acoustic waves [91]. Electrokinetic separation (electrophoretic and dielectrophoretic), which is quite popular for bacterial separation, has not been reported extensively for foodborne pathogens in recent years [92,93].
4.2. Microfluidic Concentration Channel
The use of microfluidic channels for concentration and enrichment has been of recent interest due to the small volume most sequencing techniques require (~100 µL for PCR) and the low MID of most foodborne pathogens. Using a concentration microfluidic channel reduces the need for pretreatment processes for the tested sample and lowers the required time from sample collection to detection. One approach to increase the concentration of the pathogens, specifically viruses and bacteria, is the use of homobifunctional imidoesters (HIs) that include positively charged chemical solutions followed by isothermal solid-phase nucleic acid amplification to detect pathogens according to their nucleic acids [94], however, although this method will allow significant amplification of the nucleic acids in the sample, it still requires about an hour of pretreatment and laboratory settings for it to work. Another approach is the use of auxiliary forces such as acoustic waves, magnetic fields, or electric fields. Similar to microfluidic separation, the auxiliary forces are used to remove the target pathogen from the main sample volume and force it into a specific area where the same amount of cells are now in a smaller volume of carrier fluid, which in turn effectively increases the concentration of the tested sample [17,89,90]. In a similar fashion, the use of mechanical forces as described for separation channels, again forces the pathogens into a smaller volume and hence increases their concentration [88,89,90]. The major advantages of using auxiliary forces or internal forces are their time saving and compatibility with the field conditions. It is very clear that when using auxiliary or internal mechanical forces, the channel separates the pathogen and increases its concentration all at once.
4.3. Microfluidic Detection Channel
Microfluidic detection channels show the simplest design and are used mostly as a way to bring the sample to the biosensor with the correct flow characteristics, which will allow for the successful detection of the target pathogen. The detection channel will mostly use very low flow rates to avoid kinetic interference from the movement of the particles. They will also be combined with a screen-printed electrode (SPE) platform or a fabricated integrated circuit (IC) to carry out the electrochemical detection [18,19,20,55,84]. A big advantage of these platforms is in their simplicity, which makes their fabrication easily repeatable. Detection channels will also only require a pump of sorts to inject the sample, which will significantly reduce the cost and increase their field compatibility [84].
4.4. Microfluidic Mixing Channel
Microfluidic channels pose the biggest design challenge due to the laminar nature of microfluidic flow and its typically very low Reynolds number, which does not allow for the formation of turbulence. Microfluidic mixing channels in the field of biosensing are used mostly to form complexes of the target pathogens with other micro-particles. These complexes will in turn help the specificity and sensitivity of the detection [16,17]. One approach to allow the mixing within a microfluidic channel is a mechanical one, called the Tesla mixing structure. This design has two inlets that are opposite to one another (top of a “T”) and a single outlet that is perpendicular to both (Tail of a “T”) to form a “T” like shape. To further improve the mixing, the tail is designed so that the flowlines will overlaps by creating a back flow at different regions [17,95]. Another approach is a diffusive one; adding a perpendicular channel with a very low flow rate to the main channel will cause the main flow to carry the secondary flow and mix the reagents or assisted diffusion by placing a magnetic stirrer to a specific area in the channel. In addition, a passive approach for mixing can be achieved by adding different barriers, and holes at strategic locations in the channel will also change the flow profile and force mixing of the different reagents, or storing dry and wet reagents at different locations in the channel to be carried by the main flow when the sample is injected [95]. There are also many other microfluidic mixing techniques including electric or magnetic field-based mixing [96,97], ultrasonic or acoustic mixing [98,99], which could be used, but have not been reported in the literature in recent years.
5. Lab-on-a-Chip Electrochemical Biosensors
The combination of microfluidics and electrochemical detection on one LOC platform offers the benefits of both worlds such as the small fluid volume and quick processing of microfluidics, along with the sensitivity and specificity of electrochemical biosensors. Electrochemical biosensors can be easily implemented into a microfluidic chip by utilizing modern integrated circuit fabrication techniques [55], microfluidic paper-based technology [57], and 3D printing techniques [18], as seen in Table 2. It is also clear that LOC devices present a relatively quick and accurate detection, as portrayed by the quick detection time of a few minutes [100,101,102,103] and up to no more than 3 h and the LOD of down to 4 CFU/mL for whole bacteria, or 60 copies when the biorecognition element is genome-based [104,105], even when the detection is conducted in complex food matrices, as seen in Table 3. LOC devices also have the ability to combine multiple sensors on one platform to detect multiple pathogens at the same time. This technique, known as multiplexing, utilizes more than one biorecognition element on different detection regions of the device, reduces the consumption of resources, and further reduces the sample pretreatment [55].
LOC portability and field compatibility is an important factor when assessing a device. In addition to the material, the microfluidic channel is made from, as shown in Figure 2, auxiliary equipment such as pumps, readers or a potentiostat, and biorecognition element stability. In most cases, the required auxiliary devices are larger than the LOC device and take up significant space [106], and therefore limits the application for on-field use. The miniaturization of pumps [68,69,107,108] and the use of microcapillaries [109,110] have enabled their integration into the device. In some instances, hand held potentiostats [111,112] integrated with a smart phone reader have been shown to provide better field compatibility [107,109], therefore significantly reducing the need for auxiliary equipment. The stability of the biomolecules of the biosensors presents the biggest challenge for widely used LOC biosensors in the field. Most natural biorecognition elements such as peptides, antibodies, bacteriophage, etc. are very sensitive to environmental conditions. To overcome their stability, molecularly imprinted polymers (MIPs) have been rising as an alternative. MIPs are artificially prepared materials the show advantages with respect to natural biorecognition elements such as reversable adsorption/release of the target pathogen. MIPs can also be imprinted with different nanomaterials to improve their magnetic, optic, or electric characteristics [113,114,115].
LOC offers not only the combination of microfluidics and detection, but also the automation of the whole process. For example, a device using microfluidics to bring the sample to the electrochemical biosensor that utilizes amperometric tests to detect Escherichia coli O157:H7 by using horse radish labeled antibody as the biorecognition element and forms an immobilized Ab/bacteria/anti-E. coli antibody structure following exposure to the bacteria [124]. Another interesting approach is the use of paper science to fabricate LOC devices. Paper offers an easier way to transport the fluid and allows for complex 2D structures to be utilized with the use of a printer and specialized ink. They are very portable and can not only deliver the sample to the sensing area, but can also merge multiple reagents, split samples, and delay the delivery by creating hydrophilic or hydrophobic areas on the paper [104,125]. This technology was shown to create a disposable impedimetric biosensor with immobilized antibodies as its biorecognition element. The paper itself acted as the microfluidic channel and carried the sample to the desired location [57]. On the other hand, paper-based LOC platforms tend to lack in sensitivity and their reproducibility is also a big issue that has yet to be solved [126]. LOC platforms can also utilize auxiliary forces such as dielectrophoresis [17,55] or magnetic field [16] to manipulate, focus, and concentrate the sample. They can do so by adding focused electrodes to form the electric field for dielectrophoresis, while an electric coil or a strong magnet can form the magnetic field. Applying such forces have been reported to enhance signal response by up to 18-fold when compared to the reaction without them [55]. Although these methods offer significant advantages and their fabrication is relatively established and common, they also add significant cost and energy requirements to the device.
6. Conclusions
This article began with a review of the impact of foodborne illnesses on public health and the economy. Further in this review, the established standards for foodborne pathogen diagnosis for a range of food matrices and pathogen types were discussed in detail. The types of different foodborne pathogens, along with their minimal infective doses and recommended standards for human consumption, were also discussed. The review comprehensively discusses the various types of microfluidic platforms that have been developed and reported for biosensing applications, with an emphasis on electrochemical-based platforms for foodborne pathogens. The review also discusses the distinction between platforms that can solely focus on detection versus the platforms that combine sample preparation and detection on a single device. The importance of material choices for microfluidic platforms based on the desired sensitivity, selectivity, reusability, portability, and field suitability for end application, was critically reviewed in this article. The use of MIPs instead of the commonly used natural biorecognition elements could significantly improve the shelf life and stability of ECBS platforms. Moreover, the use of hand-held potentiostats, integrated capillaries, micropumps, and valves as part of the LOC instead of using syringes and other large auxiliary equipment will aid in the simplification of these platforms for field use.
Future Directions: The research in the field of microfluidic electrochemical biosensors points to the development of devices that will combine all steps from sample preparation to detection including separation, isolation, enrichment, concentration, mixing, etc. Another trend is to develop LOC with multiplexing capabilities to process large volume samples and to screen multiple pathogens simultaneously. Finally, the integration of LOCs with a smartphone-based user interface could enable the easier adoption of these devices by the food industry for in house testing.
Conceptualization, O.Z.; methodology, O.Z.; validation, O.Z., B.S. and R.P.R.; formal analysis, O.Z. and B.S.; investigation, O.Z.; resources, R.P.R.; data curation, O.Z. and B.S.; writing—original draft preparation, O.Z.; writing—review and editing, O.Z., B.S., and R.P.R.; visualization, O.Z.; supervision, R.P.R.; project administration, R.P.R.; funding acquisition, R.P.R. All authors have read and agreed to the published version of the manuscript.
The data presented in this study are openly available in the FDA website at reference number [
The authors declare no conflict of interest.
Footnotes
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.
Figure 1. Hospitalization and death rates from confirmed infections with common foodborne pathogens [47,48,49,50].
Figure 2. Microfluidic channel material choice according to the main characteristics.
Figure 3. A schematic of the different microfluidic approaches for sample preparation.
Common foodborne pathogens, infection sources, epidemiological data, and standards.
Foodborne Pathogen | Food Products [ |
USDA Minimum Infective Dose (MID) Levels [Cells] [ |
British Standard [CFU/g] [ |
FDA Standard [CFU/g] [ |
Meaning [ |
---|---|---|---|---|---|
Campylobacter spp. | Poultry, beef, dairy products, and untreated drinking water. | xa400–500 | >0.04 | Any detectable presence c | Potential health hazard and unfit for human consumption |
<0.04 | No detectable presence c | Fit for human consumption | |||
Shiga toxin-producing Escherichia coli | Under cooked beef, dairy products made with unpasteurized milk or post pasteurization contaminated milk, vegetables, and untreated drinking water. | 10–100 | >100 | Any detectable presence c | Not fit for human consumption |
20–100 | No detectable presence c | Farther testing is needed mostly still fit for human consumption | |||
<20 | Fit for human consumption | ||||
Salmonella typhimurium | Eggs, poultry, pork, beef, dairy products, seeds, herbs, vegetables, chocolate. | 15–20 | >0.04 | Any detectable presence c | Potential health hazard and unfit for human consumption |
<0.04 | No detectable presence c | Fit for human consumption | |||
Listeria monocytogenes | Poultry, pork, beef, dairy products, bread, fish | Unknown. May vary with the strain and susceptibility of the individual. | >100 | >100 d | Not fit for human consumption |
10–100 | Not fit for vulnerable groups consumption (e.g., hospital food) | ||||
<10 | <100 d | Fit for human consumption | |||
Shigella spp. | Foods that are consumed raw, fruits, vegetables, recreational water, water contaminated with stool. | a 10 | a S. sonnei < 500 CFU | NA | Potentially injurious to |
a S. dysenteriae <200 CFU | |||||
a S. flexneri < 140 CFU | |||||
a Virulent strain < 10 CFU | |||||
Mycobacterium bovis | Contaminated, unpasteurized dairy products, cattle, bison, elk, and deer. Can be acquired through air or wounds of contaminated animal. | b 1 CFU | NA | NA | Potentially injurious to |
Vibrio vulnificus | Raw or undercooked oysters and other seafood | 106 (102 in predisposed persons) | NA | >30 [MPN/g] e | Not fit for human consumption |
Yersinia enterocolitica | Raw or undercooked pork | 109 | NA | Any detectable presence c | Not fit for human consumption |
Norovirus | Contaminated food, drinks, surfaces, or people | 10–100 viral particles | NA | NA | Potentially injurious to |
Rotavirus | Stool particles in food and drinks due to bad hygiene. | 10–100 viral particles | NA | NA | Potentially injurious to |
Cyclospora | Water, fresh produce, food or water contaminated with stool. | Unknown, predicted to be as low as 200 oocysts. | NA | NA | Potentially injurious to |
a Infective dose (ID50) (i.e., the oral dose required to cause disease in 50% of healthy adult volunteers challenged with a virulent strain of the pathogen). b Minimum infective doses in cattle. c In dairy products according to section 402(a)(1) of the Act (21 U.S.C. 342(a)(1)). d In ready to eat foods according to section 402(a)(1) of the Act (21 U.S.C. 342(a)(1)). e In shellfish according to the 2009 NSSP Guide for the Control of Molluscan Shellfish, FDA.
The different uses of microfluidic channels for electrochemical biosensors.
Goal | Methodology | Foodborne Pathogen | Flow Rate [µL/min] | Sample Volume [µL] | Concentration Range | Channel Material | Detection Technique | Ref. |
---|---|---|---|---|---|---|---|---|
Separation | Gradient magnetic field. | Listeria monocytogenes | 1000 | 3500 | 102–105 [CFU/mL] | PDMS | Impedimetric phase shift analysis | [ |
SiO2 microbeads separation. | Norovirus | NA | NA | 100 [pM]–3.5 [nM] | PDMS | Differential pulse voltammetry | [ |
|
Concentration | Dielectrophoresis | Escherichia coli O157:H7 | 2 [µL/min]—sample flow4 [µL/min]—wash flow4 [µL/min]—Ag enhancement | 20 [µL]—sample8 [µL]—wash8 [µL]—Ag enhancement | 103–105 |
PDMS | Impedimetric | [ |
Positive dielectrophoresis | Escherichia coli O157:H7Salmonella typhimurium | 1–2 [µL/min] | 1000 | 10–120 [Cells/mL]—Salmonella typhimurium13–1000 [Cells/mL]—Escherichia coli O157:H7 | Glass | Impedimetric | [ |
|
Detection | Transporting the solutions to the sensor | Escherichia coli Crooks Strain | 180 | 100 | 105–108 [Cells/mL] | Polyacrylate | Impedimetric | [ |
Transporting the solutions to the sensor | Listeria monocytogenes | Sample was dripped into specifically made wells on the microfluidic chip. | 80 | 102–103 [CFU/mL] | PDMS | Impedimetric | [ |
|
Transporting the solutions to the sensor | Salmonella (B and D) | 2 | NA | 290–1000 [Cells/mL] | PDMS | Impedimetric | [ |
|
Polymer coated paper modified with PAMAM(NH2)64-Ab | Salmonella typhimurium | NA | 5000 | 101–108 [Cells/mL] | Paper | Potentiometric | [ |
|
Cotton thread carried the sample to an aptamer with functionalized MoS2 nanosheets | Vibrio parahaemolyticus | NA | 101–106 [CFU/mL] | Cotton thread | CV and DPV | [ |
||
Mixing | Tesla mixing structure | Escherichia coli O157:H7 | 2 | 20 | 103–105 [CFU/mL] | PDMS | Impedimetric | [ |
Magnetic stirring | Listeria monocytogenes | 2000 | 205 | 102–105[CFU/mL] | PDMS | Impedimetric Phase shift analysis | [ |
Different examples for the electrochemical platform for foodborne pathogen detection.
Food Borne Pathogen | Detection Technique | Bio-Recognition Event | Sample Type | Analysis Time | Detection Range/Detection Limit | References |
---|---|---|---|---|---|---|
Listeria monocytogenes | Amperometric | Antigen-antibody | Milk | NR | 102 to 106 [CFU/mL] | [ |
Impedimetric | Antigen-antibody | Filtered tomato extract | NR | 4 [CFU/mL] | [ |
|
Impedimetric | Magnetic nanoparticles-antibody-urease | Spiked lettuce | NR | 3 × 102 [cells] | [ |
|
Impedimetric | Modified magnetic nanoparticles—antibody-urease | Spiked lettuce | 1 h | 1.6 x 102 [CFU/mL] | [ |
|
Impedimetric | Micro-electrodes functionalized with antibodies—miniaturized, portable EIS biochip | Milk | NR | 55 [CFU/mL] | [ |
|
Impedimetric | Immunomagnetic nanoparticles-urease -screen-printed electrode | Spiked lettuce | <3 h | 1.6 x 103 [CFU/mL] | [ |
|
Campylobacter jejuni | Amperometric | Antibody, phosphatase | Turkey carcass wash | 2.5 h | 102–107 [CFU/mL]LOD = 2 × 104 [CFU/mL] | [ |
Amperometric | Antibody | Milk | <1.5 h | 1 × 103–5 × 105 [CFU/ mL]LOD = 4 × 102 [CFU/mL] | [ |
|
E. coli O157:H7 | Impedimetric | Antibody | Ground beef | NR | 2.05 × 103 [CFU/gr] | [ |
Salmonella typhimurium | Amperometric | Antibody | Milk | 125 min | 10 [CFU/mL] | [ |
Impedimetric | Aptamer | Apple juice | 45 min | NR | [ |
|
Impedimetric | Aptamer—Diazonium base | Apple juice | 30 min | NR | [ |
|
Impedimetric | Interdigitated electrode array coated with Salmonella antibody | Ready to eat turkey | 1 h | 300 cells/mL | [ |
|
Staphylococcus aureus | Impedimetric | Antibody | Spiked milk | ~30 min | 13 [CFU/mL] | [ |
Potentiometric | Aptamers | Pig skin | NR | 2.4 × 103–2.0 × 104 [CFU/mL] | [ |
|
Norovirus | Cyclic voltammetry | Selective capture agent concanavalin A | Lettuce extract | 60 copies/mL | [ |
|
Immuno-based electrochemical biosensor | Monoclonal antibody | Clinical fecal sample | 1 h | 104 copies/mL | [ |
References
1. FDA. Foodborne Pathogens. Available online: https://www.fda.gov/food/outbreaks-foodborne-illness/foodborne-pathogens (accessed on 10 January 2022).
2. Jahan, S. Epidemiology of foodborne illness. Scientific, Health and Social Aspects of the Food Industry; Valdez, B.; Schor, M.; Zlatev, R. InTech: Rijeka, Croatia, 2012; pp. 321-342. ISBN 978-953-307-916-5
3. CDC. Foodborne Outbreaks. Available online: https://www.cdc.gov/foodsafety/outbreaks/lists/outbreaks-list.html (accessed on 10 December 2022).
4. Hoffmann, S.; Batz, M.B.; Morris, J.G. Annual Cost of Illness and Quality-Adjusted Life Year Losses in the United States Due to 14 Foodborne Pathogens†. J. Food Prot.; 2012; 75, pp. 1292-1302. [DOI: https://dx.doi.org/10.4315/0362-028X.JFP-11-417] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/22980013]
5. Hedberg, C.W. Foodborne illness acquired in the United States. Emerg. Infect. Dis.; 2011; 17, pp. 1338-1340. [DOI: https://dx.doi.org/10.3201/eid1707.110019] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/21762617]
6. LeDuc, P.; Agaba, M.; Cheng, C.M.; Gracio, J.; Guzman, A.; Middelberg, A. Beyond disease, how biomedical engineering can improve global health. Sci. Transl. Med.; 2014; 6, 266fs48. [DOI: https://dx.doi.org/10.1126/scitranslmed.3009067]
7. Priyanka, B.; Patil, R.K.; Dwarakanath, S. A review on detection methods used for foodborne pathogens. Indian J. Med. Res.; 2016; 144, pp. 327-338. [DOI: https://dx.doi.org/10.4103/0971-5916.198677]
8. Feng, P. Impact of Molecular Biology on the Detection of Foodborne Pathogens. Appl. Biochem. Biotechnol.—Part B Mol. Biotechnol.; 1997; 7, pp. 267-278. [DOI: https://dx.doi.org/10.1007/BF02740817]
9. Foddai, A.C.G.; Grant, I.R. Methods for detection of viable foodborne pathogens: Current state-of-art and future prospects. Appl. Microbiol. Biotechnol.; 2020; 104, pp. 4281-4288. [DOI: https://dx.doi.org/10.1007/s00253-020-10542-x] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/32215710]
10. Lam, H.M.; Remais, J.; Fung, M.C.; Xu, L.; Sun, S.S.M. Food supply and food safety issues in China. Lancet; 2013; 381, pp. 2044-2053. [DOI: https://dx.doi.org/10.1016/S0140-6736(13)60776-X] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/23746904]
11. Chin, C.D.; Linder, V.; Sia, S.K. Lab-on-a-chip devices for global health: Past studies and future opportunities. Lab Chip; 2007; 7, pp. 41-57. [DOI: https://dx.doi.org/10.1039/B611455E]
12. Castle, L.M.; Schuh, D.A.; Reynolds, E.E.; Furst, A.L. Electrochemical Sensors to Detect Bacterial Foodborne Pathogens. ACS Sensors; 2021; 6, pp. 1717-1730. [DOI: https://dx.doi.org/10.1021/acssensors.1c00481] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/33955227]
13. Kumar, H.; Kuča, K.; Bhatia, S.K.; Saini, K.; Kaushal, A.; Verma, R.; Bhalla, T.C.; Kumar, D. Applications of nanotechnology in biosensor-based detection of foodborne pathogens. Sensors; 2020; 20, 1966. [DOI: https://dx.doi.org/10.3390/s20071966]
14. Lin, C.-Y.; Nhat Nguyen, U.T.; Hsieh, H.-Y.; Tahara, H.; Chang, Y.-S.; Wang, B.-Y.; Gu, B.-C.; Dai, Y.-H.; Wu, C.-C.; Tsai, I.-J. et al. Peptide-based electrochemical sensor with nanogold enhancement for detecting rheumatoid arthritis. Talanta; 2022; 236, 122886. [DOI: https://dx.doi.org/10.1016/j.talanta.2021.122886]
15. Colburn, A.W.; Levey, K.J.; O’Hare, D.; Macpherson, J.V. Lifting the lid on the potentiostat: A beginner’s guide to understanding electrochemical circuitry and practical operation. Phys. Chem. Chem. Phys.; 2021; 23, pp. 8100-8117. [DOI: https://dx.doi.org/10.1039/D1CP00661D]
16. Chen, Q.; Wang, D.; Cai, G.; Xiong, Y.; Li, Y.; Wang, M.; Huo, H.; Lin, J. Fast and sensitive detection of foodborne pathogen using electrochemical impedance analysis, urease catalysis and micro fl uidics. Biosens. Bioelectron.; 2016; 86, pp. 770-776. [DOI: https://dx.doi.org/10.1016/j.bios.2016.07.071] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/27476059]
17. Wang, R.; Xu, Y.; Sors, T.; Irudayaraj, J.; Ren, W.; Wang, R. Impedimetric detection of bacteria by using a microfluidic chip and silver nanoparticle based signal enhancement. Microchim. Acta; 2018; 185, 184. [DOI: https://dx.doi.org/10.1007/s00604-017-2645-x]
18. Siller, I.G.; Preuss, J.; Urmann, K.; Ho, M.R.; Scheper, T.; Bahnemann, J. 3D-Printed Flow Cells for Aptamer-Based. Sensors; 2020; 20, 4421. [DOI: https://dx.doi.org/10.3390/s20164421]
19. Chiriacò, M.S.; Parlangeli, I.; Sirsi, F.; Poltronieri, P.; Primiceri, E. Impedance sensing platform for detection of the food pathogen listeria monocytogenes. Electronics; 2018; 7, 347. [DOI: https://dx.doi.org/10.3390/electronics7120347]
20. Liu, J.; Jasim, I.; Shen, Z.; Zhao, L.; Dweik, M.; Zhang, S.; Almasri, M. A microfluidic based biosensor for rapid detection of Salmonella in food products. PLoS ONE; 2019; 14, e0216873. [DOI: https://dx.doi.org/10.1371/journal.pone.0216873]
21. Dzyadevych, S.V.; Arkhypova, V.N.; Soldatkin, A.P.; El’skaya, A.V.; Martelet, C.; Jaffrezic-Renault, N. Amperometric enzyme biosensors: Past, present and future. Itbm-Rbm; 2008; 29, pp. 171-180. [DOI: https://dx.doi.org/10.1016/j.rbmret.2007.11.007]
22. Lopez-Tellez, J.; Ramirez-Montes, S.; Ferreira, T.A.; Santos, E.M.; Rodriguez, J.A. Application of Voltammetric Sensors for Pathogen Bacteria Detection: A Review. Chemosensors; 2022; 10, 424. [DOI: https://dx.doi.org/10.3390/chemosensors10100424]
23. Dutta, P.; Lu, Y.-J.; Hsieh, H.-Y.; Lee, T.-Y.; Lee, Y.-T.; Cheng, C.-M.; Fan, Y.-J. Detection of Candida albicans Using a Manufactured Electrochemical Sensor. Micromachines; 2021; 12, 166. [DOI: https://dx.doi.org/10.3390/mi12020166]
24. Fan, Y.-J.; Hsu, Y.-C.; Gu, B.-C.; Wu, C.-C. Voltammetric measurement of Escherichia coli concentration through p-APG hydrolysis by endogenous β-galactosidase. Microchem. J.; 2020; 154, 104641. [DOI: https://dx.doi.org/10.1016/j.microc.2020.104641]
25. Mairhofer, J.; Roppert, K.; Ertl, P. Microfluidic systems for pathogen sensing: A review. Sensors; 2009; 9, pp. 4804-4823. [DOI: https://dx.doi.org/10.3390/s90604804] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/22408555]
26. Fernández-la-Villa, A.; Pozo-Ayuso, D.F.; Castaño-Álvarez, M. Microfluidics and electrochemistry: An emerging tandem for next-generation analytical microsystems. Curr. Opin. Electrochem.; 2019; 15, pp. 175-185. [DOI: https://dx.doi.org/10.1016/j.coelec.2019.05.014]
27. Sheen, H.-J.; Panigrahi, B.; Kuo, T.-R.; Hsu, W.-C.; Chung, P.-S.; Xie, Q.-Z.; Lin, C.-Y.; Chang, Y.-S.; Lin, C.-T.; Fan, Y.-J. Electrochemical biosensor with electrokinetics-assisted molecular trapping for enhancing C-reactive protein detection. Biosens. Bioelectron.; 2022; 210, 114338. [DOI: https://dx.doi.org/10.1016/j.bios.2022.114338] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/35550939]
28. Zhang, C.; Xing, D.; Li, Y. Micropumps, microvalves, and micromixers within PCR microfluidic chips: Advances and trends. Biotechnol. Adv.; 2007; 25, pp. 483-514. [DOI: https://dx.doi.org/10.1016/j.biotechadv.2007.05.003] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/17601695]
29. Yoon, J.-Y.; Kim, B. Lab-on-a-Chip Pathogen Sensors for Food Safety. Sensors; 2012; 12, pp. 10713-10741. [DOI: https://dx.doi.org/10.3390/s120810713]
30. Mei, Y.; He, C.; Zeng, W.; Luo, Y.; Liu, C.; Yang, M.; Kuang, Y.; Lin, X.; Huang, Q. Electrochemical Biosensors for Foodborne Pathogens Detection Based on Carbon Nanomaterials: Recent Advances and Challenges. Food Bioprocess Technol.; 2022; 15, pp. 498-513. [DOI: https://dx.doi.org/10.1007/s11947-022-02759-7]
31. Villalonga, A.; Sánchez, A.; Mayol, B.; Reviejo, J.; Villalonga, R. Electrochemical biosensors for food bioprocess monitoring. Curr. Opin. Food Sci.; 2022; 43, pp. 18-26. [DOI: https://dx.doi.org/10.1016/j.cofs.2021.09.006]
32. Curulli, A. Electrochemical Biosensors in Food Safety: Challenges and Perspectives. Molecules; 2021; 26, 2940. [DOI: https://dx.doi.org/10.3390/molecules26102940]
33. Schmidt-Speicher, L.M.; Länge, K. Microfluidic integration for electrochemical biosensor applications. Curr. Opin. Electrochem.; 2021; 29, 100755. [DOI: https://dx.doi.org/10.1016/j.coelec.2021.100755]
34. Lonchamps, P.-L.; He, Y.; Wang, K.; Lu, X. Detection of pathogens in foods using microfluidic “lab-on-chip”: A mini review. J. Agric. Food Res.; 2022; 10, 100430. [DOI: https://dx.doi.org/10.1016/j.jafr.2022.100430]
35. Health Protection Agency. Guidelines for Assessing the Microbiological Safety of Ready-to-Eat Foods Placed on the Market; Health Protection Agency: London, UK, 2009; 33.
36. Faherty, C.S.; Lampel, K.A. Shigella. Food Microbiology; ASM Press: Washington, DC, USA, 2019; pp. 317-345.
37. Applied Nutrition Center for Food Safety and of Regulatory Affairs Office CPG Sec 555.320 Listeria Monocytogenes. Available online: https://www.fda.gov/regulatory-information/search-fda-guidance-documents/cpg-sec-555320-listeria-monocytogenes#policy (accessed on 15 November 2022).
38. Applied Nutrition Center for Food Safety and of Regulatory Affairs Office CPG Sec. 527.300 Dairy Products—Microbial Contaminants and Alkaline Phosphatase Activity (CPG 7106.08). Available online: https://www.federalregister.gov/documents/2010/12/23/2010-32232/compliance-policy-guide-sec-527300-dairy-products-microbial-contaminants-and-alkaline-phosphatase (accessed on 15 November 2022).
39. Sanitation Program National Shellfish Nssp Guide for the Control of Molluscan Shellfish. Available online: https://www.issc.org/Data/Sites/1/media/2009 nssp guide/2009 nssp guide updated 08-08-2011.pdf (accessed on 15 November 2022).
40. Zolti, O.; Suganthan, B.; Maynard, R.; Asadi, H.; Locklin, J.; Ramasamy, R.P. Electrochemical Biosensor for Rapid Detection of Listeria monocytogenes. J. Electrochem. Soc.; 2022; 169, 067510. [DOI: https://dx.doi.org/10.1149/1945-7111/ac7a63]
41. United States Department of Agriculture. Introduction to the Microbiology of Food Processing: Small Plant News Guidebook Series; United States Department of Agriculture: Washington DC, USA, 2012; pp. 1-64.
42. Centers for Disease Control and Prevention. Mycobacterium Bovis (bovine tuberculosis) in Humans; Centers for Disease Control and Prevention: Atlanta, Georgia, 2011; 2.
43. Ohio Department of Health. IDCM Section 3: Vibriosis. Infectious Disease Control Manual (IDCM); Ohio Department of Health: Columbus, OH, USA, 2019; pp. 1-8.
44. Ohio Department of Health. IDCM Section 3: Yersiniosis. Infectious Disease Control Manual (IDCM); Ohio Department of Health: Columbus, OH, USA, 2019; pp. 1-5.
45. Robilotti, E.; Deresinski, S.; Pinsky, B.A. Norovirus. Clin. Microbiol. Rev.; 2015; 28, pp. 134-164. [DOI: https://dx.doi.org/10.1128/CMR.00075-14] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/25567225]
46. Smith, H. V Cyclospora. Infectious Disease: Foodborne Disease; Simjee, S. Humana Press Inc.: Totowa, NJ, USA, 2007; pp. 277-301. ISBN 9781588295187
47. Rangel, J.M.; Sparling, P.H.; Crowe, C.; Griffin, P.M.; Swerdlow, D.L. Epidemiology of Escherichia coli O157:H7 outbreaks, United States, 1982–2002. Emerg. Infect. Dis.; 2005; 11, pp. 603-609. [DOI: https://dx.doi.org/10.3201/eid1104.040739]
48. Cartwright, E.J.; Jackson, K.A.; Johnson, S.D.; Graves, L.M.; Silk, B.J.; Mahon, B.E. Listeriosis outbreaks and associated food vehicles, United States, 1998–2008. Emerg. Infect. Dis.; 2013; 19, pp. 1-9. [DOI: https://dx.doi.org/10.3201/eid1901.120393]
49. McCrickard, L.S.; Crim, S.M.; Kim, S.; Bowen, A. Disparities in severe shigellosis among adults—Foodborne diseases active surveillance network, 2002–2014. BMC Public Health; 2018; 18, 221. [DOI: https://dx.doi.org/10.1186/s12889-018-5115-4]
50. Foodborne Diseases Active Surveillance Network. Foodborne Diseases Active Surveillance Network (FoodNet): FoodNet Surveillance Reports for 1996–2020 (Final Reports); U.S. Department of Health and Human Services, CDC: Atlanta, GA, USA, 2021.
51. Ren, K.; Zhou, J.; Wu, H. Materials for Microfluidic Chip Fabrication. Acc. Chem. Res.; 2013; 46, pp. 2396-2406. [DOI: https://dx.doi.org/10.1021/ar300314s]
52. Hou, X.; Zhang, Y.S.; Santiago, G.T.; Alvarez, M.M.; Ribas, J.; Jonas, S.J.; Weiss, P.S.; Andrews, A.M.; Aizenberg, J.; Khademhosseini, A. Interplay between materials and microfluidics. Nat. Rev. Mater.; 2017; 2, 17016. [DOI: https://dx.doi.org/10.1038/natrevmats.2017.16]
53. Nielsen, J.B.; Hanson, R.L.; Almughamsi, H.M.; Pang, C.; Fish, T.R.; Woolley, A.T. Microfluidics: Innovations in Materials and Their Fabrication and Functionalization. Anal. Chem.; 2020; 92, pp. 150-168. [DOI: https://dx.doi.org/10.1021/acs.analchem.9b04986]
54. Niculescu, A.-G.; Chircov, C.; Bîrcă, A.C.; Grumezescu, A.M. Fabrication and Applications of Microfluidic Devices: A Review. Int. J. Mol. Sci.; 2021; 22, 2011. [DOI: https://dx.doi.org/10.3390/ijms22042011]
55. Abdullah, A.; Dastider, S.G.; Jasim, I.; Shen, Z.; Yuksek, N.; Zhang, S.; Dweik, M.; Almasri, M. Microfluidic based impedance biosensor for pathogens detection in food products. Electrophoresis; 2019; 40, pp. 508-520. [DOI: https://dx.doi.org/10.1002/elps.201800405] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/30556147]
56. Chand, R.; Neethirajan, S. Microfluidic platform integrated with graphene-gold nano-composite aptasensor for one-step detection of norovirus. Biosens. Bioelectron.; 2017; 98, pp. 47-53. [DOI: https://dx.doi.org/10.1016/j.bios.2017.06.026] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/28649024]
57. Silva, N.F.D.; Almeida, C.M.R.; Magalhães, J.M.C.S.; Gonçalves, M.P.; Freire, C.; Delerue-Matos, C. Development of a disposable paper-based potentiometric immunosensor for real-time detection of a foodborne pathogen. Biosens. Bioelectron.; 2019; 141, 111317. [DOI: https://dx.doi.org/10.1016/j.bios.2019.111317]
58. Jiang, H.; Sun, Z.; Guo, Q.; Weng, X. Microfluidic thread-based electrochemical aptasensor for rapid detection of Vibrio parahaemolyticus. Biosens. Bioelectron.; 2021; 182, 113191. [DOI: https://dx.doi.org/10.1016/j.bios.2021.113191] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/33780852]
59. Pinti, M.; Kambham, T.; Wang, B.; Prakash, S. Fabrication of Centimeter Long, Ultra-Low Aspect Ratio Nanochannel Networks in Borosilicate Glass Substrates. J. Nanotechnol. Eng. Med.; 2013; 4, 020905. [DOI: https://dx.doi.org/10.1115/1.4025366]
60. Pawar, P.; Ballav, R.; Kumar, A. Micromachining of Borosilicate Glass: A State of Art Review. Mater. Today Proc.; 2017; 4, pp. 2813-2821. [DOI: https://dx.doi.org/10.1016/j.matpr.2017.02.161]
61. Jung, H.C.; Lu, W.; Wang, S.; Lee, L.J.; Hu, X. Etching of Pyrex glass substrates by inductively coupled plasma reactive ion etching for micro/nanofluidic devices. J. Vac. Sci. Technol. B Microelectron. Nanom. Struct.; 2006; 24, 3162. [DOI: https://dx.doi.org/10.1116/1.2388959]
62. Chang, T.-L.; Chen, Z.-C.; Lee, Y.-W.; Li, Y.-H.; Wang, C.-P. Ultrafast laser ablation of soda-lime glass for fabricating microfluidic pillar array channels. Microelectron. Eng.; 2016; 158, pp. 95-101. [DOI: https://dx.doi.org/10.1016/j.mee.2016.03.034]
63. Perrone, E.; Cesaria, M.; Zizzari, A.; Bianco, M.; Ferrara, F.; Raia, L.; Guarino, V.; Cuscunà, M.; Mazzeo, M.; Gigli, G. et al. Potential of CO2-laser processing of quartz for fast prototyping of microfluidic reactors and templates for 3D cell assembly over large scale. Mater. Today Bio; 2021; 12, 100163. [DOI: https://dx.doi.org/10.1016/j.mtbio.2021.100163]
64. Tang, T.; Yuan, Y.; Yalikun, Y.; Hosokawa, Y.; Li, M.; Tanaka, Y. Glass based micro total analysis systems: Materials, fabrication methods, and applications. Sens. Actuators B Chem.; 2021; 339, 129859. [DOI: https://dx.doi.org/10.1016/j.snb.2021.129859]
65. Aralekallu, S.; Boddula, R.; Singh, V. Development of glass-based microfluidic devices: A review on its fabrication and biologic applications. Mater. Des.; 2023; 225, 111517. [DOI: https://dx.doi.org/10.1016/j.matdes.2022.111517]
66. Fan, Y.-J.; Huang, M.-Z.; Hsiao, Y.-C.; Huang, Y.-W.; Deng, C.-Z.; Yeh, C.; Husain, R.A.; Lin, Z.-H. Enhancing the sensitivity of portable biosensors based on self-powered ion concentration polarization and electrical kinetic trapping. Nano Energy; 2020; 69, 104407. [DOI: https://dx.doi.org/10.1016/j.nanoen.2019.104407]
67. Iliescu, C.; Taylor, H.; Avram, M.; Miao, J.; Franssila, S. A practical guide for the fabrication of microfluidic devices using glass and silicon. Biomicrofluidics; 2012; 6, 016505. [DOI: https://dx.doi.org/10.1063/1.3689939] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/22662101]
68. Laser, D.J.; Santiago, J.G. A review of micropumps. J. Micromechanics Microengineering; 2004; 14, pp. R35-R64. [DOI: https://dx.doi.org/10.1088/0960-1317/14/6/R01]
69. Oh, K.W.; Ahn, C.H. A review of microvalves. J. Micromechanics Microengineering; 2006; 16, pp. R13-R39. [DOI: https://dx.doi.org/10.1088/0960-1317/16/5/R01]
70. Ali, U.; Karim, K.J.B.A.; Buang, N.A. A Review of the Properties and Applications of Poly (Methyl Methacrylate) (PMMA). Polym. Rev.; 2015; 55, pp. 678-705. [DOI: https://dx.doi.org/10.1080/15583724.2015.1031377]
71. Frazer, R.Q.; Byron, R.T.; Osborne, P.B.; West, K.P. PMMA: An Essential Material in Medicine and Dentistry. J. Long. Term. Eff. Med. Implants; 2005; 15, pp. 629-639. [DOI: https://dx.doi.org/10.1615/JLongTermEffMedImplants.v15.i6.60]
72. Martynova, L.; Locascio, L.E.; Gaitan, M.; Kramer, G.W.; Christensen, R.G.; MacCrehan, W.A. Fabrication of Plastic Microfluid Channels by Imprinting Methods. Anal. Chem.; 1997; 69, pp. 4783-4789. [DOI: https://dx.doi.org/10.1021/ac970558y]
73. Xu, J.; Locascio, L.; Gaitan, M.; Lee, C.S. Room-Temperature Imprinting Method for Plastic Microchannel Fabrication. Anal. Chem.; 2000; 72, pp. 1930-1933. [DOI: https://dx.doi.org/10.1021/ac991216q]
74. McCormick, R.M.; Nelson, R.J.; Alonso-Amigo, M.G.; Benvegnu, D.J.; Hooper, H.H. Microchannel Electrophoretic Separations of DNA in Injection-Molded Plastic Substrates. Anal. Chem.; 1997; 69, pp. 2626-2630. [DOI: https://dx.doi.org/10.1021/ac9701997]
75. Chen, Y.; Zhang, L.; Chen, G. Fabrication, modification, and application of poly(methyl methacrylate) microfluidic chips. Electrophoresis; 2008; 29, pp. 1801-1814. [DOI: https://dx.doi.org/10.1002/elps.200700552] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/18384069]
76. Jena, R.K.; Yue, C.Y.; Lam, Y.C. Micro fabrication of cyclic olefin copolymer (COC) based microfluidic devices. Microsyst. Technol.; 2012; 18, pp. 159-166. [DOI: https://dx.doi.org/10.1007/s00542-011-1366-z]
77. Nunes, P.S.; Ohlsson, P.D.; Ordeig, O.; Kutter, J.P. Cyclic olefin polymers: Emerging materials for lab-on-a-chip applications. Microfluid. Nanofluidics; 2010; 9, pp. 145-161. [DOI: https://dx.doi.org/10.1007/s10404-010-0605-4]
78. Pemg, B.-Y.; Wu, C.-W.; Shen, Y.-K.; Lin, Y. Microfluidic chip fabrication using hot embossing and thermal bonding of COP. Polym. Adv. Technol.; 2010; 21, pp. 457-466.
79. Azouz, A.B.; Murphy, S.; Karazi, S.; Vázquez, M.; Brabazon, D. Fast Fabrication Process of Microfluidic Devices Based on Cyclic Olefin Copolymer. Mater. Manuf. Process.; 2014; 29, pp. 93-99. [DOI: https://dx.doi.org/10.1080/10426914.2013.811739]
80. Gao, H.; Yan, C.; Wu, W.; Li, J. Application of Microfluidic Chip Technology in Food Safety Sensing. Sensors; 2020; 20, 1792. [DOI: https://dx.doi.org/10.3390/s20061792]
81. Weng, X.; Kang, Y.; Guo, Q.; Peng, B.; Jiang, H. Recent advances in thread-based microfluidics for diagnostic applications. Biosens. Bioelectron.; 2019; 132, pp. 171-185. [DOI: https://dx.doi.org/10.1016/j.bios.2019.03.009]
82. Tan, W.; Powles, E.; Zhang, L.; Shen, W. Go with the capillary flow. Simple thread-based microfluidics. Sensors Actuators B Chem.; 2021; 334, 129670. [DOI: https://dx.doi.org/10.1016/j.snb.2021.129670]
83. Berzina, B.; Anand, R.K. Tutorial review: Enrichment and separation of neutral and charged species by ion concentration polarization focusing. Anal. Chim. Acta; 2020; 1128, pp. 149-173. [DOI: https://dx.doi.org/10.1016/j.aca.2020.06.021]
84. Singh, C.; Ali, M.A.; Kumar, V.; Ahmad, R.; Sumana, G. Functionalized MoS2 nanosheets assembled microfluidic immunosensor for highly sensitive detection of food pathogen. Sens. Actuators B Chem.; 2018; 259, pp. 1090-1098. [DOI: https://dx.doi.org/10.1016/j.snb.2017.12.094]
85. Park, J.; Han, D.H.; Park, J.-K. Towards practical sample preparation in point-of-care testing: User-friendly microfluidic devices. Lab Chip; 2020; 20, pp. 1191-1203. [DOI: https://dx.doi.org/10.1039/D0LC00047G] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/32119024]
86. Ghuman, A. Bacteriophage-Assisted Microfluidic Separation of Pathogens from Food Matrices. Ph.D. Thesis; University of Georgia: Athens, Georgia, 2016.
87. Zhou, Y.; Ramasamy, R.P. Isolation and separation of Listeria monocytogenes using bacteriophage P100- modi fi ed magnetic particles. Colloids Surf. B Biointerfaces; 2019; 175, pp. 421-427. [DOI: https://dx.doi.org/10.1016/j.colsurfb.2018.12.007] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/30562716]
88. Tang, M.; Wang, G.; Kong, S.K.; Ho, H.P. A review of biomedical centrifugal microfluidic platforms. Micromachines; 2016; 7, 26. [DOI: https://dx.doi.org/10.3390/mi7020026]
89. Sivaramakrishnan, M.; Kothandan, R.; Govindarajan, D.K.; Meganathan, Y.; Kandaswamy, K. Active microfluidic systems for cell sorting and separation. Curr. Opin. Biomed. Eng.; 2020; 13, pp. 60-68. [DOI: https://dx.doi.org/10.1016/j.cobme.2019.09.014]
90. Wang, B.; Park, B. Microfluidic Sampling and Biosensing Systems for Foodborne Escherichia coli and Salmonella. Foodborne Pathog. Dis.; 2022; 19, pp. 359-375. [DOI: https://dx.doi.org/10.1089/fpd.2021.0087]
91. Ngamsom, B.; Lopez-Martinez, M.J.; Raymond, J.-C.; Broyer, P.; Patel, P.; Pamme, N. On-chip acoustophoretic isolation of microflora including S. typhimurium from raw chicken, beef and blood samples. J. Microbiol. Methods; 2016; 123, pp. 79-86. [DOI: https://dx.doi.org/10.1016/j.mimet.2016.01.016] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/26835844]
92. Song, Y.; Li, Z.; Feng, A.; Zhang, J.; Liu, Z.; Li, D. Electrokinetic detection and separation of living algae in a microfluidic chip: Implication for ship’s ballast water analysis. Environ. Sci. Pollut. Res.; 2021; 28, pp. 22853-22863. [DOI: https://dx.doi.org/10.1007/s11356-020-12315-5]
93. Morani, M.; Mai, T.D.; Krupova, Z.; van Niel, G.; Defrenaix, P.; Taverna, M. Recent electrokinetic strategies for isolation, enrichment and separation of extracellular vesicles. TrAC Trends Anal. Chem.; 2021; 135, 116179. [DOI: https://dx.doi.org/10.1016/j.trac.2021.116179]
94. Eun, C.; Koo, B.; Yeong, E.; Yeun, J.; Kim, S.; Shin, Y. Simple and label-free pathogen enrichment via homobifunctional imidoesters using a micro fl uidic (SLIM) system for ultrasensitive pathogen detection in various clinical specimens. Biosens. Bioelectron.; 2018; 111, pp. 66-73.
95. Castillo-león, J.; Svendsen, W.E.; Dimaki, M.; Arima, V.; Akram, M.S.; Miserere, S.; Neuman, C.; Kipling, G.D. Lab-on-a-Chip Devices and Micro-Total Analysis Systems A Practical Guide; 1st ed. Castillo-león, J.; Svendsen, W.E. Springer: Cham, Switzerland, Heidelberg, Germany, New York, NY, USA, Dordrecht, The Netherlands, London, UK, 2015; ISBN 9783319086866
96. Shanko, E.-S.; van de Burgt, Y.; Anderson, P.D.; den Toonder, J.M.J. Microfluidic Magnetic Mixing at Low Reynolds Numbers and in Stagnant Fluids. Micromachines; 2019; 10, 731. [DOI: https://dx.doi.org/10.3390/mi10110731]
97. Gao, X.; Li, Y. Simultaneous microfluidic pumping and mixing using an array of asymmetric 3D ring electrode pairs in a cylindrical microchannel by the AC electroosmosis effect. Eur. J. Mech.—B/Fluids; 2019; 75, pp. 361-371. [DOI: https://dx.doi.org/10.1016/j.euromechflu.2018.10.008]
98. Li, Y.; Cai, S.; Shen, H.; Chen, Y.; Ge, Z.; Yang, W. Recent advances in acoustic microfluidics and its exemplary applications. Biomicrofluidics; 2022; 16, 031502. [DOI: https://dx.doi.org/10.1063/5.0089051] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/35712527]
99. Demcenko, A.; Witte, C.; Reboud, J.; Cooper, J.M. Ultrasonic Wave Mixing for Nonlinear Ultrasonics in a Microfluidic Capillary. Proceedings of the 2019 IEEE International Ultrasonics Symposium (IUS); Glasgow, UK, 6–9 October 2019; pp. 284-286.
100. Alexandre, D.L.; Melo, A.M.A.; Furtado, R.F.; Borges, M.F.; Figueiredo, E.A.T.; Biswas, A.; Cheng, H.N.; Alves, C.R. A Rapid and Specific Biosensor for Salmonella Typhimurium Detection in Milk. Food Bioprocess Technol.; 2018; 11, pp. 748-756. [DOI: https://dx.doi.org/10.1007/s11947-017-2051-8]
101. Sheikhzadeh, E.; Chamsaz, M.; Turner, A.P.F.; Jager, E.W.H.; Beni, V. Label-free impedimetric biosensor for Salmonella Typhimurium detection based on poly [pyrrole-co-3-carboxyl-pyrrole] copolymer supported aptamer. Biosens. Bioelectron.; 2016; 80, pp. 194-200. [DOI: https://dx.doi.org/10.1016/j.bios.2016.01.057] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/26836649]
102. Bagheryan, Z.; Raoof, J.-B.; Golabi, M.; Turner, A.P.F.; Beni, V. Diazonium-based impedimetric aptasensor for the rapid label-free detection of Salmonella typhimurium in food sample. Biosens. Bioelectron.; 2016; 80, pp. 566-573. [DOI: https://dx.doi.org/10.1016/j.bios.2016.02.024]
103. Bhardwaj, J.; Devarakonda, S.; Kumar, S.; Jang, J. Development of a paper-based electrochemical immunosensor using an antibody-single walled carbon nanotubes bio-conjugate modified electrode for label-free detection of foodborne pathogens. Sensors Actuators B Chem.; 2017; 253, pp. 115-123. [DOI: https://dx.doi.org/10.1016/j.snb.2017.06.108]
104. Hong, S.A.; Kwon, J.; Kim, D.; Yang, S. A rapid, sensitive and selective electrochemical biosensor with concanavalin A for the preemptive detection of norovirus. Biosens. Bioelectron.; 2015; 64, pp. 338-344. [DOI: https://dx.doi.org/10.1016/j.bios.2014.09.025]
105. Wang, N.; Pan, G.; Guan, S.; Rong, S.; Wang, D.; Gao, Z.; Tian, P.; Li, Q. A Broad-Range Disposable Electrochemical Biosensor Based on Screen-Printed Carbon Electrodes for Detection of Human Noroviruses. Front. Bioeng. Biotechnol.; 2022; 10, 845660. [DOI: https://dx.doi.org/10.3389/fbioe.2022.845660] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/35402404]
106. Mohammed, M.I.; Haswell, S.; Gibson, I. Lab-on-a-chip or Chip-in-a-lab: Challenges of Commercialization Lost in Translation. Procedia Technol.; 2015; 20, pp. 54-59. [DOI: https://dx.doi.org/10.1016/j.protcy.2015.07.010]
107. Cai, G.; Wang, Y.; Zhang, Y.; Zheng, L.; Lin, J. Magnetorheological elastomer and smartphone enable microfluidic biosensing of foodborne pathogen. Chinese Chem. Lett.; 2022; 2022, 108059. [DOI: https://dx.doi.org/10.1016/j.cclet.2022.108059]
108. Qi, W.; Zheng, L.; Hou, Y.; Duan, H.; Wang, L.; Wang, S.; Liu, Y.; Li, Y.; Liao, M.; Lin, J. A finger-actuated microfluidic biosensor for colorimetric detection of foodborne pathogens. Food Chem.; 2022; 381, 131801. [DOI: https://dx.doi.org/10.1016/j.foodchem.2021.131801]
109. Hassan, S.; Tariq, A.; Noreen, Z.; Donia, A.; Zaidi, S.Z.J.; Bokhari, H.; Zhang, X. Capillary-Driven Flow Microfluidics Combined with Smartphone Detection: An Emerging Tool for Point-of-Care Diagnostics. Diagnostics; 2020; 10, 509. [DOI: https://dx.doi.org/10.3390/diagnostics10080509] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/32708045]
110. Lakhera, P.; Chaudhary, V.; Bhardwaj, B.; Kumar, P.; Kumar, S. Development and recent advancement in microfluidics for point of care biosensor applications: A review. Biosens. Bioelectron. X; 2022; 11, 100218. [DOI: https://dx.doi.org/10.1016/j.biosx.2022.100218]
111. Joe, C.; Lee, B.H.; Kim, S.H.; Ko, Y.; Gu, M.B. Aptamer duo-based portable electrochemical biosensors for early diagnosis of periodontal disease. Biosens. Bioelectron.; 2022; 199, 113884. [DOI: https://dx.doi.org/10.1016/j.bios.2021.113884] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/34942544]
112. Cruz, A.F.D.; Norena, N.; Kaushik, A.; Bhansali, S. A low-cost miniaturized potentiostat for point-of-care diagnosis. Biosens. Bioelectron.; 2014; 62, pp. 249-254. [DOI: https://dx.doi.org/10.1016/j.bios.2014.06.053] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/25016332]
113. Wang, L.; Hao, L.; Qi, W.; Huo, X.; Xue, L.; Liu, Y.; Zhang, Q.; Lin, J. A sensitive Salmonella biosensor using platinum nanoparticle loaded manganese dioxide nanoflowers and thin-film pressure detector. Sensors Actuators B Chem.; 2020; 321, 128616. [DOI: https://dx.doi.org/10.1016/j.snb.2020.128616]
114. Zhang, F.; Luo, L.; Gong, H.; Chen, C.; Cai, C. A magnetic molecularly imprinted optical chemical sensor for specific recognition of trace quantities of virus. RSC Adv.; 2018; 8, pp. 32262-32268. [DOI: https://dx.doi.org/10.1039/C8RA06204H]
115. Yang, S.; Ouyang, H.; Su, X.; Gao, H.; Kong, W.; Wang, M.; Shu, Q.; Fu, Z. Dual-recognition detection of Staphylococcus aureus using vancomycin-functionalized magnetic beads as concentration carriers. Biosens. Bioelectron.; 2016; 78, pp. 174-180. [DOI: https://dx.doi.org/10.1016/j.bios.2015.11.041]
116. Cheng, C.; Peng, Y.; Bai, J.; Zhang, X.; Liu, Y.; Fan, X.; Ning, B.; Gao, Z. Rapid detection of Listeria monocytogenes in milk by self-assembled electrochemical immunosensor. Sens. Actuators B Chem.; 2014; 190, pp. 900-906. [DOI: https://dx.doi.org/10.1016/j.snb.2013.09.041]
117. Radhakrishnan, R.; Jahne, M.; Rogers, S.; Suni, I.I. Detection of Listeria Monocytogenes by Electrochemical Impedance Spectroscopy. Electroanalysis; 2013; 25, pp. 2231-2237. [DOI: https://dx.doi.org/10.1002/elan.201300140]
118. Chen, Q.; Lin, J.; Gan, C.; Wang, Y.; Wang, D.; Xiong, Y.; Lai, W.; Li, Y.; Wang, M. A sensitive impedance biosensor based on immunomagnetic separation and urease catalysis for rapid detection of Listeria monocytogenes using an immobilization-free interdigitated array microelectrode. Biosens. Bioelectron.; 2015; 74, pp. 504-511. [DOI: https://dx.doi.org/10.1016/j.bios.2015.06.007]
119. Wang, D.; Chen, Q.; Huo, H.; Bai, S.; Cai, G.; Lai, W.; Lin, J. Efficient separation and quantitative detection of Listeria monocytogenes based on screen-printed interdigitated electrode, urease and magnetic nanoparticles. Food Control.; 2017; 73, pp. 555-561. [DOI: https://dx.doi.org/10.1016/j.foodcont.2016.09.003]
120. Che, Y.; Li, Y.; Slavik, M. Detection of Campylobacter jejuni in poultry samples using an enzyme-linked immunoassay coupled with an enzyme electrode. Biosens. Bioelectron.; 2001; 16, pp. 791-797. [DOI: https://dx.doi.org/10.1016/S0956-5663(01)00222-6] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/11679257]
121. Viswanathan, S.; Rani, C.; Ho, J.A. Electrochemical immunosensor for multiplexed detection of food-borne pathogens using nanocrystal bioconjugates and MWCNT screen-printed electrode. Talanta; 2012; 94, pp. 315-319. [DOI: https://dx.doi.org/10.1016/j.talanta.2012.03.049] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/22608454]
122. Xu, M.; Wang, R.; Li, Y. Rapid detection of Escherichia coli O157:H7 and Salmonella Typhimurium in foods using an electrochemical immunosensor based on screen-printed interdigitated microelectrode and immunomagnetic separation. Talanta; 2016; 148, pp. 200-208. [DOI: https://dx.doi.org/10.1016/j.talanta.2015.10.082]
123. Zelada-Guillén, G.A.; Sebastián-Avila, J.L.; Blondeau, P.; Riu, J.; Rius, F.X. Label-free detection of Staphylococcus aureus in skin using real-time potentiometric biosensors based on carbon nanotubes and aptamers. Biosens. Bioelectron.; 2012; 31, pp. 226-232. [DOI: https://dx.doi.org/10.1016/j.bios.2011.10.021] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/22154169]
124. Altintas, Z.; Akgun, M.; Kokturk, G.; Uludag, Y. A fully automated microfluidic-based electrochemical sensor for real-time bacteria detection. Biosens. Bioelectron.; 2018; 100, pp. 541-548. [DOI: https://dx.doi.org/10.1016/j.bios.2017.09.046]
125. Puiu, M.; Bala, C. Microfluidics-integrated biosensing platforms as emergency tools for on-site field detection of foodborne pathogens. TrAC Trends Anal. Chem.; 2020; 125, 115831. [DOI: https://dx.doi.org/10.1016/j.trac.2020.115831]
126. Carrell, C.; Kava, A.; Nguyen, M.; Menger, R.; Munshi, Z.; Call, Z.; Nussbaum, M.; Henry, C. Beyond the lateral flow assay: A review of paper-based microfluidics. Microelectron. Eng.; 2019; 206, pp. 45-54. [DOI: https://dx.doi.org/10.1016/j.mee.2018.12.002]
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2023 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
Foodborne pathogens are an important diagnostic target for the food, beverage, and health care industries due to their prevalence and the adverse effects they can cause to public health, food safety, and the economy. The standards that determine whether a given type of food is fit for consumption are set by governments and must be taken into account when designing a new diagnostic tool such as a biosensor platform. In order to meet these stringent detection limits, cost, and reliability standards, recent research has been focused on developing lab-on-a-chip-based approaches for detection devices that use microfluidic channels and platforms. The microfluidics-based devices are designed, developed, and used in different ways to achieve the established common standards for food pathogen testing that enable high throughput, rapid detection, low sample volume, and minimal pretreatment procedures. Combining microfluidic approaches with electrochemical biosensing could offer affordable, portable, and easy to use devices for food pathogen diagnostics. This review presents an analysis of the established common standards and the recent progress made in electrochemical sensors toward the development of future lab-on-a-chip devices that will aid ‘collection-to-detection’ using a single method and platform.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer