© 2016 Coetzee et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Background

The BD-FACSPresto CD4 is a new, point-of-care (POC) instrument utilising finger-stick capillary blood sampling. This study evaluated its performance against predicate CD4 testing in South Africa.

Methods

Phase-I testing: HIV+ patient samples (n = 214) were analysed on the Presto under ideal laboratory conditions using venous blood. During Phase-II, 135 patients were capillary-bled for CD4 testing on FACSPresto, performed according to manufacturer instruction. Comparative statistical analyses against predicate PLG/CD4 method and industry standards were done using GraphPad Prism 6. It included Bland-Altman with 95% limits of agreement (LOA) and percentage similarity with coefficient of variation (%CV) analyses for absolute CD4 count (cells/μl) and CD4 percentage of lymphocytes (CD4%).

Results

In Phase-I, 179/217 samples yielded reportable results with Presto using venous blood filled cartridges. Compared to predicate, a mean bias of 40.4±45.8 (LOA of -49.2 to 130.2) and %similarity (%CV) of 106.1%±7.75 (7.3%) was noted for CD4 absolute counts. In Phase-2 field study, 118/135 capillary-bled Presto samples resulted CD4 parameters. Compared to predicate, a mean bias of 50.2±92.8 (LOA of -131.7 to 232) with %similarity (%CV) 105%±10.8 (10.3%), and 2.87±2.7 (LOA of -8.2 to 2.5) with similarity of 94.7±6.5% (6.83%) noted for absolute CD4 and CD4% respectively. No significant clinical differences were indicated for either parameter using two sampling methods.

Conclusion

The Presto produced remarkable precision to predicate methods, irrespective of venous or capillary blood sampling. A consistent, clinically insignificant over-estimation (5–7%) of counts against PLG/CD4 and equivalency to FACSCount was noted. Further field studies are awaited to confirm longer-term use.

Details

Title
Performance Evaluation of the Becton Dickinson FACSPresto™ Near-Patient CD4 Instrument in a Laboratory and Typical Field Clinic Setting in South Africa
Author
Coetzee, Lindi-Marie; Moodley, Keshendree; Deborah Kim Glencross
First page
e0156266
Section
Research Article
Publication year
2016
Publication date
May 2016
Publisher
Public Library of Science
e-ISSN
19326203
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
1791324810
Copyright
© 2016 Coetzee et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.