Content area
Full text
BACKGROUND: Methods from 7 manufacturers and 1 distributor for directly measuring HDL cholesterol (C) and LDL-C were evaluated for imprecision, trueness, total error, and specificity in nonfrozen serum samples.
METHODS: We performed each direct method according to the manufacturer's instructions, using a Roche/Hitachi 917 analyzer, and compared the results with those obtained with reference measurement procedures for HDL-C and LDL-C. Imprecision was estimated for 35 runs performed with frozen pooled serum specimens and triplicate measurements on each individual sample. Sera from 37 individuals without disease and 138 with disease (primarily dyslipidemic and cardiovascular) were measured by each method. Trueness and total error were evaluated from the difference between the direct methods and reference measurement procedures. Specificity was evaluated from the dispersion in differences observed.
RESULTS: Imprecision data based on 4 frozen serum pools showed total CVs <3.7% for HDL-C and <4.4% for LDL-C. Bias for the nondiseased group ranged from -5.4% to 4.8% for HDL-C and from -6.8% to 1.1% for LDL-C, and for the diseased group from -8.6% to 8.8% for HDL-C and from -11.8% to 4.1% for LDL-C. Total error for the nondiseased group ranged from -13.4% to 13.6% for HDL-C and from -13.3% to 13.5% for LDL-C, and for the diseased group from -19.8% to 36.3% for HDL-C and from -26.6% to 31.9% for LDL-C.
CONCLUSIONS: Six of 8 HDL-C and 5 of 8 LDL-C direct methods met the National Cholesterol Education Program total error goals for nondiseased individuals. All the methods failed to meet these goals for diseased individuals, however, because of lack of specificity toward abnormal lipoproteins.
© 2010 American Association for Clinical Chemistry
The current national guidelines on the use of LDL cholesterol (LDL-C)10 and HDL-C for cardiovascular risk assessment (1, 2 ) were largely based on early epidemiologic studies that established the link between lipoproteins and cardiovascular disease. Most of these studies used chemical precipitation methods for HDL-C, and β-quantification or the Friedewald calculation for LDL-C. These older procedures have largely been supplanted by newer direct measurement methods for lipoprotein that do not depend on the physical separation of the different lipoprotein classes. The development of direct measurement methods was prompted, in part, by a recommendation from a National Cholesterol Education Program (NCEP) expert laboratory panel that...